Xuefeng Zhang1, Chengcheng Zhu2, Wenjia Peng1, Bing Tian1, Luguang Chen1, Zhongzhao Teng3, Jianping Lu1, Umar Sadat4, David Saloner2, Qi Liu1.
Abstract
PURPOSE: To evaluate the scan-rescan reproducibility of high-resolution magnetic resonance imaging (MRI) of middle cerebral artery (MCA) plaque, and calculate the number of subjects needed for future longitudinal clinical studies.Entities:
Mesh:
Year: 2015 PMID: 26247869 PMCID: PMC4527831 DOI: 10.1371/journal.pone.0134913
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Image acquisition and segmentation of middle cerebral artery (MCA) plaque.
(A) Prescription of slices perpendicular to MCA base on the maximal intensity projection (MIP) of 3D TOF images. Red arrow shows the stenosis site. (B) Sample images (slice number 6) shows MCA plaque (yellow arrow). (C) Segmentation of the MCA lumen and outer wall boundaries.
Comparison of area and volume measurement between two repeated scans.
Data are summarized as median [inter quartile range].
| Scan 1 | Scan 2 | p | |
|---|---|---|---|
| Lumen Area (mm2) | 3.00 [1.71, 4.69] | 2.87 [1.63, 4.44] | 0.49 |
| Plaque Area (mm2) | 9.42 [6.98, 11.08] | 9.66 [6.99, 11.65] | 0.08 |
| Vessel Area (mm2) | 12.40 [10.09, 14.42] | 12.64 [10.00, 14.71] | 0.67 |
| Lumen Volume (mm3) | 9.58 [5.22, 16.66] | 9.64 [4.84, 16.56] | 0.60 |
| Plaque volume (mm3) | 30.08 [18.09, 45.80] | 32.38 [18.17, 47.16] | 0.65 |
| Vessel Volume (mm3) | 39.62 [20.67, 62.26] | 42.85 [20.86, 63.47] | 0.89 |
Fig 2Sample scan and re-scan images of middle cerebral artery plaques from two patients.
Arrows show the plaques.
Reproducibility analysis between two scans.
| Mean | SD(between scan) | CV | bias | LOA | ICC | |
|---|---|---|---|---|---|---|
| Lumen Area (mm2) | 3.14 | 0.37 | 11.8% | 0.05 | (-0.67, 0.78) | 0.98 |
| Plaque Area (mm2) | 9.37 | 0.70 | 7.5% | -0.15 | (-1.52, 1.22) | 0.97 |
| Vessel Area (mm2) | 12.56 | 0.77 | 6.1% | -0.10 | (-1.60, 1.41) | 0.97 |
| Lumen Volume (mm3) | 10.68 | 0.85 | 8.0% | 0.18 | (-1.50, 1.85) | 0.99 |
| Plaque volume (mm3) | 31.84 | 1.88 | 5.9% | -0.52 | (-4.21, 3.17) | 0.99 |
| Vessel Volume (mm3) | 42.70 | 2.08 | 4.9% | -0.34 | (-4.42, 3.74) | 0.99 |
SD: standard deviation; CV: coefficient of variance; LOA: limit of agreement; ICC: intra-class coefficient.
Fig 3Bland-Altman plot of area measurements between two repeated scans.
Solid lines define bias, dashed lines define limits of agreement (LOA).
Fig 4Bland-Altman plot of volume measurements between two repeated scans.
Solid lines define bias, dashed lines define limits of agreement (LOA).
Sample size calculation for detecting 5% to 20% change in area and volume.
(Power: 80%, p = 0.05).
| CV | 5% | 10% | 15% | 20% | |
|---|---|---|---|---|---|
| Lumen Area | 11.8% | 176 | 44 | 20 | 11 |
| Vessel Area | 6.1% | 47 | 12 | 6 | 3 |
| Plaque Area | 7.5% | 72 | 18 | 8 | 5 |
| Lumen Volume | 8.0% | 81 | 21 | 9 | 6 |
| Vessel Volume | 4.9% | 31 | 8 | 4 | 2 |
| Plaque volume | 5.9% | 44 | 11 | 5 | 3 |
CV: coefficient of variance.