Dosimetric verification of lung cancer treatment using the CBCTs estimated from limited-angle on-board projections.

You Zhang1, Fang-Fang Yin2, Lei Ren2.   

Abstract

PURPOSE: Lung cancer treatment is susceptible to treatment errors caused by interfractional anatomical and respirational variations of the patient. On-board treatment dose verification is especially critical for the lung stereotactic body radiation therapy due to its high fractional dose. This study investigates the feasibility of using cone-beam (CB)CT images estimated by a motion modeling and free-form deformation (MM-FD) technique for on-board dose verification.
METHODS: Both digital and physical phantom studies were performed. Various interfractional variations featuring patient motion pattern change, tumor size change, and tumor average position change were simulated from planning CT to on-board images. The doses calculated on the planning CT (planned doses), the on-board CBCT estimated by MM-FD (MM-FD doses), and the on-board CBCT reconstructed by the conventional Feldkamp-Davis-Kress (FDK) algorithm (FDK doses) were compared to the on-board dose calculated on the "gold-standard" on-board images (gold-standard doses). The absolute deviations of minimum dose (ΔDmin), maximum dose (ΔDmax), and mean dose (ΔDmean), and the absolute deviations of prescription dose coverage (ΔV100%) were evaluated for the planning target volume (PTV). In addition, 4D on-board treatment dose accumulations were performed using 4D-CBCT images estimated by MM-FD in the physical phantom study. The accumulated doses were compared to those measured using optically stimulated luminescence (OSL) detectors and radiochromic films.
RESULTS: Compared with the planned doses and the FDK doses, the MM-FD doses matched much better with the gold-standard doses. For the digital phantom study, the average (± standard deviation) ΔDmin, ΔDmax, ΔDmean, and ΔV100% (values normalized by the prescription dose or the total PTV) between the planned and the gold-standard PTV doses were 32.9% (±28.6%), 3.0% (±2.9%), 3.8% (±4.0%), and 15.4% (±12.4%), respectively. The corresponding values of FDK PTV doses were 1.6% (±1.9%), 1.2% (±0.6%), 2.2% (±0.8%), and 17.4% (±15.3%), respectively. In contrast, the corresponding values of MM-FD PTV doses were 0.3% (±0.2%), 0.9% (±0.6%), 0.6% (±0.4%), and 1.0% (±0.8%), respectively. Similarly, for the physical phantom study, the average ΔDmin, ΔDmax, ΔDmean, and ΔV100% of planned PTV doses were 38.1% (±30.8%), 3.5% (±5.1%), 3.0% (±2.6%), and 8.8% (±8.0%), respectively. The corresponding values of FDK PTV doses were 5.8% (±4.5%), 1.6% (±1.6%), 2.0% (±0.9%), and 9.3% (±10.5%), respectively. In contrast, the corresponding values of MM-FD PTV doses were 0.4% (±0.8%), 0.8% (±1.0%), 0.5% (±0.4%), and 0.8% (±0.8%), respectively. For the 4D dose accumulation study, the average (± standard deviation) absolute dose deviation (normalized by local doses) between the accumulated doses and the OSL measured doses was 3.3% (±2.7%). The average gamma index (3%/3 mm) between the accumulated doses and the radiochromic film measured doses was 94.5% (±2.5%).
CONCLUSIONS: MM-FD estimated 4D-CBCT enables accurate on-board dose calculation and accumulation for lung radiation therapy. It can potentially be valuable for treatment quality assessment and adaptive radiation therapy.

Entities:  

Mesh:

Year:  2015        PMID: 26233206      PMCID: PMC4514726          DOI: 10.1118/1.4926559

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  47 in total

1.  4D Cone-beam CT reconstruction using a motion model based on principal component analysis.

Authors:  David Staub; Alen Docef; Robert S Brock; Constantin Vaman; Martin J Murphy
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Audio-visual biofeedback for respiratory-gated radiotherapy: impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy.

Authors:  Rohini George; Theodore D Chung; Sastry S Vedam; Viswanathan Ramakrishnan; Radhe Mohan; Elisabeth Weiss; Paul J Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-07-01       Impact factor: 7.038

3.  Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation.

Authors:  Yong Yang; Eduard Schreibmann; Tianfang Li; Chuang Wang; Lei Xing
Journal:  Phys Med Biol       Date:  2007-01-12       Impact factor: 3.609

Review 4.  Advances in image-guided radiation therapy.

Authors:  Laura A Dawson; David A Jaffray
Journal:  J Clin Oncol       Date:  2007-03-10       Impact factor: 44.544

5.  Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapy.

Authors:  Joan Hatton; Boyd McCurdy; Peter B Greer
Journal:  Phys Med Biol       Date:  2009-07-10       Impact factor: 3.609

6.  Accumulated dose in liver stereotactic body radiotherapy: positioning, breathing, and deformation effects.

Authors:  Michael Velec; Joanne L Moseley; Tim Craig; Laura A Dawson; Kristy K Brock
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-12-28       Impact factor: 7.038

7.  Synchronized moving aperture radiation therapy (SMART): improvement of breathing pattern reproducibility using respiratory coaching.

Authors:  Toni Neicu; Ross Berbeco; John Wolfgang; Steve B Jiang
Journal:  Phys Med Biol       Date:  2006-01-19       Impact factor: 3.609

8.  Comparing digital tomosynthesis to cone-beam CT for position verification in patients undergoing partial breast irradiation.

Authors:  Junan Zhang; Q Jackie Wu; Devon J Godfrey; Toyosi Fatunase; Lawrence B Marks; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-01-08       Impact factor: 7.038

9.  A study on adaptive IMRT treatment planning using kV cone-beam CT.

Authors:  George X Ding; Dennis M Duggan; Charles W Coffey; Matthew Deeley; Dennis E Hallahan; Anthony Cmelak; Arnold Malcolm
Journal:  Radiother Oncol       Date:  2007-08-20       Impact factor: 6.280

10.  Realistic CT simulation using the 4D XCAT phantom.

Authors:  W P Segars; M Mahesh; T J Beck; E C Frey; B M W Tsui
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

View more
  15 in total

1.  Reducing scan angle using adaptive prior knowledge for a limited-angle intrafraction verification (LIVE) system for conformal arc radiotherapy.

Authors:  Yawei Zhang; Fang-Fang Yin; You Zhang; Lei Ren
Journal:  Phys Med Biol       Date:  2017-03-24       Impact factor: 3.609

2.  A biomechanical modeling-guided simultaneous motion estimation and image reconstruction technique (SMEIR-Bio) for 4D-CBCT reconstruction.

Authors:  Xiaokun Huang; You Zhang; Jing Wang
Journal:  Phys Med Biol       Date:  2018-02-08       Impact factor: 3.609

3.  Low dose CBCT reconstruction via prior contour based total variation (PCTV) regularization: a feasibility study.

Authors:  Yingxuan Chen; Fang-Fang Yin; Yawei Zhang; You Zhang; Lei Ren
Journal:  Phys Med Biol       Date:  2018-04-19       Impact factor: 3.609

4.  A Biomechanical Modeling Guided CBCT Estimation Technique.

Authors:  You Zhang; Joubin Nasehi Tehrani; Jing Wang
Journal:  IEEE Trans Med Imaging       Date:  2016-11-01       Impact factor: 10.048

5.  A new CT reconstruction technique using adaptive deformation recovery and intensity correction (ADRIC).

Authors:  You Zhang; Jianhua Ma; Puneeth Iyengar; Yuncheng Zhong; Jing Wang
Journal:  Med Phys       Date:  2017-05-12       Impact factor: 4.071

6.  Automatic liver tumor localization using deep learning-based liver boundary motion estimation and biomechanical modeling (DL-Bio).

Authors:  Hua-Chieh Shao; Xiaokun Huang; Michael R Folkert; Jing Wang; You Zhang
Journal:  Med Phys       Date:  2021-11-19       Impact factor: 4.071

7.  4D liver tumor localization using cone-beam projections and a biomechanical model.

Authors:  You Zhang; Michael R Folkert; Bin Li; Xiaokun Huang; Jeffrey J Meyer; Tsuicheng Chiu; Pam Lee; Joubin Nasehi Tehrani; Jing Cai; David Parsons; Xun Jia; Jing Wang
Journal:  Radiother Oncol       Date:  2018-11-14       Impact factor: 6.280

8.  Low dose cone-beam computed tomography reconstruction via hybrid prior contour based total variation regularization (hybrid-PCTV).

Authors:  Yingxuan Chen; Fang-Fang Yin; Yawei Zhang; You Zhang; Lei Ren
Journal:  Quant Imaging Med Surg       Date:  2019-07

9.  Daily edge deformation prediction using an unsupervised convolutional neural network model for low dose prior contour based total variation CBCT reconstruction (PCTV-CNN).

Authors:  Yingxuan Chen; Fang-Fang Yin; Zhuoran Jiang; Lei Ren
Journal:  Biomed Phys Eng Express       Date:  2019-10-07

10.  An unsupervised 2D-3D deformable registration network (2D3D-RegNet) for cone-beam CT estimation.

Authors:  You Zhang
Journal:  Phys Med Biol       Date:  2021-03-24       Impact factor: 4.174

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.