Performance characteristics of the Pantak DXT-300 kilovoltage X-ray treatment machine.

R J Aukett1, D W Thomas, A W Seaby, J T Gittins.   

Abstract

The beam and performance characteristics of a new orthovoltage X-ray unit, the Pantak DXT-300 have been evaluated. Data were collated for four qualities: 3.27 mmAl, 7.15 mmAl, 1.65 mmCu and 3.51 mmCu half value layer (HVL) (SE = 0.04 mm). Parameters which were investigated included beam quality, central axis depth dose, relative output, backscatter factors, field uniformity, peripheral dose and head leakage. The calibration procedure and the performance of the dosimetry system have also been described.

Mesh:

Year:  1996        PMID: 8949675     DOI: 10.1259/0007-1285-69-824-726

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  3 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Characterization of an orthovoltage biological irradiator used for radiobiological research.

Authors:  Rezvan Azimi; Parham Alaei; Emiliano Spezi; Susanta K Hui
Journal:  J Radiat Res       Date:  2015-02-17       Impact factor: 2.724

3.  MicroRNA Profiling in Oesophageal Adenocarcinoma Cell Lines and Patient Serum Samples Reveals a Role for miR-451a in Radiation Resistance.

Authors:  Frederike Butz; Ann-Kathrin Eichelmann; George C Mayne; Tingting Wang; Isabell Bastian; Karen Chiam; Shashikanth Marri; Pamela J Sykes; Bas P Wijnhoven; Eelke Toxopeus; Michael Z Michael; Christos S Karapetis; Richard Hummel; David I Watson; Damian J Hussey
Journal:  Int J Mol Sci       Date:  2020-11-24       Impact factor: 5.923

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.