Literature DB >> 23556881

Microionization chamber air-kerma calibration coefficients as a function of photon energy for x-ray spectra in the range of 20-250 kVp relative to 60Co.

J R Snow1, J A Micka, L A DeWerd.   

Abstract

PURPOSE: To investigate the applicability of a wide range of microionization chambers for reference dosimetry measurements in low- and medium-energy x-ray beams.
METHODS: Measurements were performed with six cylindrical microchamber models, as well as one scanning chamber and two Farmer-type chambers for comparison purposes. Air-kerma calibration coefficients were determined at the University of Wisconsin Accredited Dosimetry Calibration Laboratory for each chamber for a range of low- and medium-energy x-ray beams (20-250 kVp), with effective energies ranging from 11.5 keV to 145 keV, and a (60)Co beam. A low-Z proof-of-concept microchamber was developed and calibrated with and without a high-Z silver epoxy on the collecting electrode.
RESULTS: All chambers composed of low-Z materials (Z ≤ 13), including the Farmer-type chambers, the scanning chamber, and the PTW TN31014 and the proof-of-concept microchambers, exhibited air-kerma calibration coefficients with little dependence on the quality of the beam. These chambers typically exhibited variations in calibration coefficients of less than 3% with the beam quality, for medium energy beams. However, variations in air-kerma calibration coefficients of greater than 50% were measured over the range of medium-energy x-ray beams for each of the microchambers containing high-Z collecting electrodes (Z > 13). For these high-Z chambers, which include the Exradin A14SL and A16 chambers, the PTW TN31006 chamber, the IBA CC01 chamber, and the proof-of-concept chamber containing silver, the average variation in air-kerma calibration coefficients between any two calibration beams was nearly 25% over the entire range of beam qualities investigated.
CONCLUSIONS: Due to the strong energy dependence observed with microchambers containing high-Z components, these chambers may not be suitable dosimeters for kilovoltage x-ray applications, as they do not meet the TG-61 requirements. It is recommended that only microchambers containing low-Z materials (Z ≤ 13) be considered for air-kerma calibrations for reference dosimetry in low- and medium-energy x-ray beams.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23556881     DOI: 10.1118/1.4794491

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Dosimetric characterization and output verification for conical brachytherapy surface applicators. Part I. Electronic brachytherapy source.

Authors:  Regina K Fulkerson; John A Micka; Larry A DeWerd
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

3.  Influence of Specific Treatment Parameters on Nontarget and Out-of-Field Doses in a Phantom Model of Prostate SBRT with CyberKnife and TrueBeam.

Authors:  Marta Kruszyna-Mochalska; Agnieszka Skrobala; Piotr Romanski; Adam Ryczkowski; Wiktoria Suchorska; Katarzyna Kulcenty; Igor Piotrowski; Dorota Borowicz; Kinga Graczyk; Natalia Matuszak; Julian Malicki
Journal:  Life (Basel)       Date:  2022-04-23

4.  Computed tomography dosimetry with high-resolution detectors commonly used in radiotherapy - an energy dependence study.

Authors:  Mario Liebmann; Bjoern Poppe; Heiner von Boetticher
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

5.  Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs.

Authors:  Lori A Young; Fei Yang; Ning Cao; Juergen Meyer
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.