Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beams.

X A Li1, C M Ma, D Salhani.   

Abstract

In this work, nine commonly used dosimetry detectors have been investigated to determine suitable relative dosimeters for kilovoltage x-ray beams. By comparison with the Monte Carlo calculated data, it was determined that for the detectors studied the PTW N23342, Markus and NACP parallel-plate chambers are more suitable for the measurement of percentage depth dose (PDD) data for this beam quality range with an uncertainty of about 3%. A diode detector may be used to measure the PDD for the 100 kVp beam, but it is not suitable for higher energies (300 kVp). The Capintec parallel-plate chamber may be adequate for medium-energy photons, but it has a slightly higher uncertainty for low-energy x-rays (100 kVp). For the measurement of beam profiles, diode and film yield incorrect profile tails, which can be corrected using the RK ionization chamber.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9434308     DOI: 10.1088/0031-9155/42/12/019

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Fast and accurate Monte Carlo modeling of a kilovoltage X-ray therapy unit using a photon-source approximation for treatment planning in complex media.

Authors:  B Zeinali-Rafsanjani; M A Mosleh-Shirazi; R Faghihi; S Karbasi; A Mosalaei
Journal:  J Med Phys       Date:  2015 Apr-Jun

3.  Role of in vivo dosimetry with radiochromic films for dose verification during cutaneous radiation therapy.

Authors:  Hong-Wei Liu; James Gräfe; Rao Khan; Ivo Olivotto; J Eduardo Villarreal Barajas
Journal:  Radiat Oncol       Date:  2015-01-10       Impact factor: 3.481

4.  Computed tomography dosimetry with high-resolution detectors commonly used in radiotherapy - an energy dependence study.

Authors:  Mario Liebmann; Bjoern Poppe; Heiner von Boetticher
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.