J Eduardo Villarreal-Barajas1, Rao F H Khan.
Abstract
The objective of this study is to evaluate the suitability of recently introduced radiochromic film EBT3 for clinical dosimetry in the kilovoltage (kV) range. For this purpose, a kV X-ray irradiator, X RAD 320ix in the range 70 to 300 kVp, a clinical 60Co source, a 6 MV and an 18 MV X-ray clinical beam from a Varian linear accelerator were calibrated following AAPM dosimetry protocols. EBT3 films from two different EBT3 batches were placed side-by-side on the surface of a water phantom; doses from 0.5 to 4 Gy were delivered. Similarly, irradiations were performed for 60Co and 6 and 18 MV beams in a water equivalent phantom. Films were reproducibly placed at the center of a flatbed scanner and 48-bit RGB scans were obtained both pre- and postirradiations. Net optical density (netOD) and response for a given radiation quality relative to 60Co was determined for each EBT3 film. The netOD of the red color showed reproducible response (within 1%) for both batches when irradiated using the 60Co source. For a given dose of 1 Gy of kVp X-ray, the response relative to 60Co using the three color channels (red, green, and blue) decreases with decrease in kVp, reaching a maximum underresponse of ~ 20% for the 70 kVp. A significant underresponse of ~ 5% was observed at 300kVp. Responses of MV X-ray beams with respect to 60Co at the 1 Gy dose level showed no statistically significant difference. A relatively small difference in the response was observed between the two EBT3 batches used in this study in the kV X-ray range.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24423839 PMCID: PMC5711253 DOI: 10.1120/jacmp.v15i1.4439
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Radiation beam characteristics employed in experiments. Effective energies were derived using the measured beam qualities and the mass attenuation coefficient tables published by NIST.
|
|
|
|
|
|
|---|---|---|---|---|
| 70 | 2 mm Al | 3.0 mm Al | 32 | 0.316 |
| 100 | 2 mm Al | 4.2 mm Al | 38 | 0.682 |
| 150 | 2 mm Al | 6.3 mm Al | 47 | 1.538 |
| 200 |
| 2.6 mm Cu | 132 | 0.392 |
| 300 |
| 4.0 mm Cu | 168 | 0.824 |
| Co‐60 | 1250 | 1.250 |
Ratio of net optical densities for two batches of EBT3 for irradiation
|
| |||
|---|---|---|---|
|
|
|
|
|
| 50 |
|
|
|
| 75 |
|
|
|
| 100 |
|
|
|
| 200 |
|
|
|
| 400 |
|
|
|
| Mean | 0.997 | 1.025 | 1.059 |
Ratio of netOD for two batches of EBT3 film for kVp energies for 1 Gy of absorbed dose
|
| |||
|---|---|---|---|
|
|
|
|
|
| 300 |
|
|
|
| 200 |
|
|
|
| 150 |
|
|
|
| 100 |
|
|
|
| 70 |
|
|
|
| Mean | 1.04 | 1.06 | 1.10 |
Response ratio, Rx, of various kVp beams relative to for B1 and B2 batches for 1 Gy of absorbed dose
|
| |||
|---|---|---|---|
|
|
|
|
|
| 300 |
|
|
|
| 200 |
|
|
|
| 150 |
|
|
|
| 100 |
|
|
|
| 70 |
|
|
|