| Literature DB >> 26151914 |
Hiroaki Hasegawa1, Masanori Sato2, Hiroshi Tanaka3.
Abstract
The purpose of this study was to evaluate scatter radiation dose to the subject surface during X-ray computed tomography (CT) fluoroscopy using the integrated dose ratio (IDR) of an X-ray dose profile derived from an optically stimulated luminescent (OSL) dosimeter. We aimed to obtain quantitative evidence supporting the radiation protection methods used during previous CT fluoroscopy. A multislice CT scanner was used to perform this study. OSL dosimeters were placed on the top and the lateral side of the chest phantom so that the longitudinal direction of dosimeters was parallel to the orthogonal axis-to-slice plane for measurement of dose profiles in CT fluoroscopy. Measurement of fluoroscopic conditions was performed at 120 kVp and 80 kVp. Scatter radiation dose was evaluated by calculating the integrated dose determined by OSL dosimetry. The overall percent difference of the integrated doses between OSL dosimeters and ionization chamber was 5.92%. The ratio of the integrated dose of a 100-mm length area to its tails (-50 to -6 mm, 50 to 6 mm) was the lowest on the lateral side at 80 kVp and the highest on the top at 120 kVp. The IDRs for different measurement positions were larger at 120 kVp than at 80 kVp. Similarly, the IDRs for the tube voltage between the primary X-ray beam and scatter radiation was larger on the lateral side than on the top of the phantom. IDR evaluation suggested that the scatter radiation dose has a high dependence on the position and a low dependence on tube voltage relative to the primary X-ray beam for constant dose rate fluoroscopic conditions. These results provided quantitative evidence supporting the radiation protection methods used during CT fluoroscopy in previous studies.Entities:
Mesh:
Year: 2015 PMID: 26151914 PMCID: PMC4494858 DOI: 10.1371/journal.pone.0132154
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Measuring positions used for optically stimulated luminescence dosimeters.
A: Computed tomography image of the chest phantom at the center position during fluoroscopy; T, the position of optically stimulated luminescence (OSL) dosimeters on the top of the chest phantom; L, the position of OSL dosimeters on the lateral side of the chest phantom. B: Photograph of a measurement on the top of the chest phantom.
Integrated dose measurements performed using optically stimulated luminescence dosimeters and an ionization chamber.
| Measurement conditions | Integrated dose | | ε | | |||||
|---|---|---|---|---|---|---|---|
| Ionization chamber | OSL dosimeters | ||||||
| Tube voltage | Position | 15 seconds | 30 seconds | 15 seconds | 30 seconds | 15 seconds | 30 seconds |
|
|
| 134.04 ± 0.29 | 266.53 ± 0.36 | 134.06 ± 1.12 | 269.71 ± 5.78 | 0.01 | 1.19 |
|
| 79.07 ± 021 | 158.72 ± 0.39 | 82.31 ± 0.61 | 165.88 ± 2.39 | 4.10 | 4.51 | |
|
|
| 127.68 ± 0.09 | 255.54 ± 0.08 | 113.59 ± 3.17 | 227.31 ± 3.00 | 11.03 | 11.05 |
|
| 80.69 ± 0.37 | 160.51 ± 0.30 | 67.60 ± 1.11 | 130.18 ± 0.08 | 16.22 | 18.90 | |
aValues are presented as the mean ± standard error.
bMean value of five repeated measurements.
cMean value of triplicate measurements.
dThe absolute value of the percent difference was defined as (DOSL − DIC)/DIC × 100.
Fig 2Dose profiles of CT fluoroscopy performed using optically stimulated luminescence dosimeters.
A: Top. B: Lateral side. Each data point corresponds to the mean value of triplicate measurements.
Fig 3Integrated doses for the different measurement positions and tube voltage.
There were statistically significant differences (* p < 0.0001) as determined by two-way ANOVA between the main effects (measurement positions and tube voltage) for each fluoroscopic time. Error bars depict standard error of triplicates. A: Primary X-ray beam. B: Scatter radiation.
Integrated dose ratio (IDRS/T) of the 100-mm length integrated dose (DOSL) and scatter radiation dose (DS) determined by optically stimulated luminescence dosimetry.
| Measurement conditions | IDRS/T
| ||
|---|---|---|---|
| Tube voltage | Position | 15 seconds | 30 seconds |
|
|
| 0.34 | 0.35 |
|
| 0.31 | 0.31 | |
|
|
| 0.37 | 0.37 |
|
| 0.34 | 0.34 | |
aThe integrated dose ratio of the dose caused by scatter radiation distribution (outside the primary X-ray beam width) in the dose profile tails to that of the 100-mm length along the z-axis of the chest phantom was measured.
Integrated dose ratio for the measurement positions (IDRP) with optically stimulated luminescence dosimetry.
| Measurement conditions | IDRPP
| IDRPS
| ||
|---|---|---|---|---|
| Tube voltage | 15 seconds | 30 seconds | 15 seconds | 30 seconds |
|
| 1.56 | 1.54 | 1.79 | 1.82 |
|
| 1.59 | 1.67 | 1.85 | 1.90 |
The IDRP is the ratio of integrated dose that measured each tube voltage on the top to that on the lateral side of the chest phantom.
aIDRP in primary X-ray beam width along the z-axis at the chest phantom.
bIDRP of the dose profile tails caused by scatter radiation distribution (outside of the primary X-ray beam width) along the z-axis of the chest phantom.
Integrated dose ratio for the tube voltage (IDRV) determined by optically stimulated luminescence (OSL) dosimetry.
| Measurement conditions | IDRVP
| IDRVS
| ||
|---|---|---|---|---|
| Position | 15 seconds | 30 seconds | 15 seconds | 30 seconds |
|
| 1.24 | 1.23 | 1.09 | 1.11 |
|
| 1.27 | 1.34 | 1.12 | 1.15 |
The IDRV is the ratio of the integrated dose measured at each position of the chest phantom at 80 kVp to that at 120 kVp.
aIDRV of the primary X-ray beam width along the z-axis of the chest phantom.
bIDRV of the dose profile tails caused by scatter radiation distribution (outside of the primary X-ray beam width) along the z-axis of the chest phantom.