Reena Sharma1, Sunil Dutt Sharma1, Shramika Pawar1, Ajay Chaubey2, S Kantharia2, D A R Babu1.
Abstract
The screen-film system is replaced by computed radiography system for recording the images of the patients during X-ray radiography examinations. The change in imaging system requires the re-establishment of the institutional diagnostic reference levels (DRLs) for different types of X-ray examinations conducted at the hospital. For this purpose, patient specific parameters [age, height, weight, body mass index (BMI), object to image distance (OID)] and machine specific parameters (kVp, mAs, distance and field sizes) of 1875 patients during 21 different types of X-ray examinations were recorded for estimating the entrance skin dose (ESD). The ESD for each of these patients were estimated using measured X-ray beam output and the standard value of the back scatter factor. Five number summary was calculated for all the data for their presentation in the Box-Whisker plot, which provides the statistical distribution of the data. The data collected indicates that majorly performed examinations are cervical spine AP, Chest PA and Knee Lat with percentage contributions of 16.05, 16 and 8.27% respectively. The lowest contribution comes from Hip Lat which is about 1.01%. The ratio of measured ESD (maximum to minimum) for these examinations is found to be highest for the cervical spine AP with a value of 50 followed by Thoracic spine AP of 32.36. The ESD ratio for Chest PA, Knee Lat and Lumbar Spine AP are 30.75, 30.4 and 30.2 respectively. The lowest ESD ratio is for Hip Lat which is 2.68. The third quartile values of ESDs are established as the institutional DRLs. The ESD values obtained for 21 different X-ray projections are either comparable or lesser than the reported national/international values.Entities:
Keywords: CR imaging system; Patient dose; X-ray imaging; X-ray machine; computed radiography; diagnostic reference level; entrance skin dose
Year: 2015 PMID: 26150685 PMCID: PMC4471642 DOI: 10.4103/0971-6203.152244
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
QA test parameters measured for the X-ray machine used during the course of the present study
Figure 1Output (mGy/mAs) of the X-ray machine measured at different kVp stations
Range of values of the patient specific and machine specific parameters of the 21 different types of radiography projections
Figure 2Percentage contribution for 21 X-ray examinations performed in the hospital
Statistical distribution of ESD values measured for 21 different types of radiography projections
Figure 3The Box and Whisker plot for patient specific parameters measured for 21 X-ray projections
Figure 4The Box and Whisker plot for X-ray machine parameters employed for 21 X-ray projections
Figure 5The Box and Whisker plot for measured ESDs of 21 different types of X-ray projections
Measured third quartile value (75th percentile) related to patient, machine parameters and ESD for 21 different types of radiographic examinations
Comparison of measured ESD (mGy) values (75th percentile) with established DRLs at national and international levels