Measurement of bow tie profiles in CT scanners using a real-time dosimeter.

Bruce R Whiting1, Joshua D Evans2, Andreea C Dohatcu3, Jeffrey F Williamson2, David G Politte4.   

Abstract

PURPOSE: Several areas of computed tomography (CT) research require knowledge about the intensity profile of the x-ray fan beam that is introduced by a bow tie filter. This information is considered proprietary by CT manufacturers, so noninvasive measurement methods are required. One method using real-time dosimeters has been proposed in the literature. A commercially available dosimeter was used to apply that method, and analysis techniques were developed to extract fan beam profiles from measurements.
METHODS: A real-time ion chamber was placed near the periphery of an empty CT gantry and the dose rate versus time waveform was recorded as the x-ray source rotated about the isocenter. In contrast to previously proposed analysis methods that assumed a pointlike detector, the finite-size ion chamber received varying amounts of coverage by the collimated x-ray beam during rotation, precluding a simple relationship between the source intensity as a function of fan beam angle and measured intensity. A two-parameter model for measurement intensity was developed that included both effective collimation width and source-to-detector distance, which then was iteratively solved to minimize the error between duplicate measurements at corresponding fan beam angles, allowing determination of the fan beam profile from measured dose-rate waveforms. Measurements were performed on five different scanner systems while varying parameters such as collimation, kVp, and bow tie filters. On one system, direct measurements of the bow tie profile were collected for comparison with the real-time dosimeter technique.
RESULTS: The data analysis method for a finite-size detector was found to produce a fan beam profile estimate with a relative error between duplicate measurement intensities of <5%. It was robust over a wide range of collimation widths (e.g., 1-40 mm), producing fan beam profiles that agreed with a relative error of 1%-5%. Comparison with a direct measurement technique on one system produced agreement with a relative error of 2%-6%. Fan beam profiles were found to differ for different filter types on a given system and between different vendors.
CONCLUSIONS: A commercially available real-time dosimeter probe was found to be a convenient and accurate instrument for measuring fan beam profiles. An analysis method was developed that could handle a wide range of collimation widths by explicitly considering the finite width of the ion chamber. Relative errors in the profiles were found to be less than 5%. Measurements of five different clinical scanners demonstrate the variation in bow tie designs, indicating that generic bow tie models will not be adequate for CT system research.

Mesh:

Year:  2014        PMID: 25281967      PMCID: PMC4281104          DOI: 10.1118/1.4896196

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  The influence of the heel effect in cone-beam computed tomography: artifacts in standard and novel geometries and their correction.

Authors:  H Braun; Y Kyriakou; M Kachelriess; W A Kalender
Journal:  Phys Med Biol       Date:  2010-09-21       Impact factor: 3.609

2.  Prototype heel effect compensation filter for cone-beam CT.

Authors:  Shinichiro Mori; Masahiro Endo; Kanae Nishizawa; Mari Ohno; Hiroaki Miyazaki; Kazuhiko Tsujita; Yasuo Saito
Journal:  Phys Med Biol       Date:  2005-11-01       Impact factor: 3.609

3.  Validation of CT dose-reduction simulation.

Authors:  Parinaz Massoumzadeh; Steven Don; Charles F Hildebolt; Kyongtae T Bae; Bruce R Whiting
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

4.  Method for evaluating bow tie filter angle-dependent attenuation in CT: theory and simulation results.

Authors:  John M Boone
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

5.  Experimental validation of a method characterizing bow tie filters in CT scanners using a real-time dose probe.

Authors:  Sarah E McKenney; Anita Nosratieh; Dale Gelskey; Kai Yang; Shin-Ying Huang; Lin Chen; John M Boone
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

6.  Prospects for quantitative computed tomography imaging in the presence of foreign metal bodies using statistical image reconstruction.

Authors:  Jeffrey F Williamson; Bruce R Whiting; Jasenka Benac; Ryan J Murphy; G James Blaine; Joseph A O'Sullivan; David G Politte; Donald L Snyder
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

7.  Real-time dosimeter employed to evaluate the half-value layer in CT.

Authors:  Sarah E McKenney; J Anthony Seibert; George W Burkett; Dale Gelskey; Paul B Sunde; James D Newman; John M Boone
Journal:  Phys Med Biol       Date:  2013-12-19       Impact factor: 3.609

8.  Dose spread functions in computed tomography: a Monte Carlo study.

Authors:  John M Boone
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

  8 in total
  7 in total

1.  A new technique to characterize CT scanner bow-tie filter attenuation and applications in human cadaver dosimetry simulations.

Authors:  Xinhua Li; Jim Q Shi; Da Zhang; Sarabjeet Singh; Atul Padole; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  Technical Note: Measurement of bow tie profiles in CT scanners using radiochromic film.

Authors:  Bruce R Whiting; Andreea C Dohatcu; Joshua D Evans; David G Politte; Jeffrey F Williamson
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

3.  A measurement-based generalized source model for Monte Carlo dose simulations of CT scans.

Authors:  Xin Ming; Yuanming Feng; Ransheng Liu; Chengwen Yang; Li Zhou; Hezheng Zhai; Jun Deng
Journal:  Phys Med Biol       Date:  2017-01-12       Impact factor: 3.609

4.  Neutrosophic segmentation of breast lesions for dedicated breast computed tomography.

Authors:  Juhun Lee; Robert M Nishikawa; Ingrid Reiser; John M Boone
Journal:  J Med Imaging (Bellingham)       Date:  2018-03-06

5.  Determination and verification of the x-ray spectrum of a CT scanner.

Authors:  Ahmad Ibrahim Hassan; Martin Skalej; Helmut Schlattl; Christoph Hoeschen
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-07

6.  Direct and fast measurement of CT beam filter profiles with simultaneous geometrical calibration.

Authors:  Kai Yang; Xinhua Li; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2017-01       Impact factor: 4.071

7.  Estimation of primary radiation output for wide-beam computed tomography scanner.

Authors:  Atsushi Fukuda; Pei-Jan P Lin; Nao Ichikawa; Kosuke Matsubara
Journal:  J Appl Clin Med Phys       Date:  2019-05-02       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.