Polyethylene Liner Dissociation Is a Complication of the DePuy Pinnacle Cup: A Report of 23 Cases.

Andrew Yun1, Emmanuel N Koli2, John Moreland1, Richard Iorio3, John F Tilzey4, J Wesley Mesko5, Gwo-Chin Lee6, Mark Froimson7.   

Abstract

BACKGROUND: Polyethylene liner dissociation is a rare but catastrophic event in total hip arthroplasty (THA), and certain implant designs are known to be at greater risk. Although the DePuy Pinnacle (Warsaw, IN, USA) modular acetabular construct has an excellent record of fixation and wear, an unexpectedly high number of liner dissociations has been noted. QUESTIONS/PURPOSES: The purposes of this study were (1) to characterize the clinical parameters observed in a large group of patients who have experienced liner dissociations with the DePuy Pinnacle acetabular component; (2) to describe the radiographic findings in this group of patients; and (3) to calculate a minimum frequency of this complication.
METHODS: Since 2001, 23 patients with previously well-functioning THAs presented with sudden atraumatic polyethylene liner dissociation at four separate institutions. These THAs were performed between 2001 and 2013. Eight different arthroplasty specialists had performed the index hip arthroplasties using the DePuy Pinnacle acetabular component with a polyethylene liner. Polyethylene failures were evaluated for liner type and radiographic cup position. For three of the surgeons who contributed cases, institutional registries allowed the calculation of the number of components of this type that they used during the period in question, which provided a conservative estimate of the frequency of this type of failure.
RESULTS: All 23 liner failures occurred atraumatically in previously asymptomatic THAs at a mean of 48 months (range, 3-138 months). Patients characteristically reported a new and sudden onset of discomfort with audible, reproducible squeaking. Surgical inspection of dissociated liners demonstrated displacement of polyethylene with shearing of the peripheral locking tabs. Radiographic evaluation demonstrated that 14 cups were well positioned and nine cups were malpositioned outside the so-called safe zone. Conservative estimates of the frequency of this complication from the three surgeons' practices whose institutional registries allowed calculation of the lowest possible frequency were 0.32% (six of 1888), 0.77% (three of 391), and 0.82% (three of 367).
CONCLUSIONS: With this report of 23 additional liner dissociations, we suggest that surgeons should be aware of the problem and take extra precautions when using this implant to ensure locking mechanism integrity at the time of surgery. We caution that the frequency of liner dissociation may be higher than previously reported. LEVEL OF EVIDENCE: Level IV, therapeutic study.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26105151      PMCID: PMC4709308          DOI: 10.1007/s11999-015-4396-5

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  17 in total

1.  Late liner disassociation of a Pinnacle system acetabular component.

Authors:  Stephanie W Mayer; Samuel S Wellman; Michael P Bolognesi; David E Attarian
Journal:  Orthopedics       Date:  2012-04       Impact factor: 1.390

2.  Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components.

Authors:  C A Engh; P Massin; K E Suthers
Journal:  Clin Orthop Relat Res       Date:  1990-08       Impact factor: 4.176

3.  Dislodgment of polyethylene liners in first and second-generation Harris-Galante acetabular components. A report of eighteen cases.

Authors:  A González della Valle; P S Ruzo; S Li; P Pellicci; T P Sculco; E A Salvati
Journal:  J Bone Joint Surg Am       Date:  2001-04       Impact factor: 5.284

4.  A comparison of the disassociation strength of modular acetabular components.

Authors:  S Tradonsky; P D Postak; A I Froimson; A S Greenwald
Journal:  Clin Orthop Relat Res       Date:  1993-11       Impact factor: 4.176

5.  Late disassembly of modular acetabular components. A report of two cases.

Authors:  J B Retpen; S Solgaard
Journal:  Acta Orthop Scand       Date:  1993-04

6.  Dislocations after total hip-replacement arthroplasties.

Authors:  G E Lewinnek; J L Lewis; R Tarr; C L Compere; J R Zimmerman
Journal:  J Bone Joint Surg Am       Date:  1978-03       Impact factor: 5.284

7.  Primary total hip arthroplasty with a porous-coated acetabular component. Seven-to-ten-year results.

Authors:  G S Tompkins; J J Jacobs; L R Kull; A G Rosenberg; J O Galante
Journal:  J Bone Joint Surg Am       Date:  1997-02       Impact factor: 5.284

8.  Spontaneous dissociation of offset, face-changing polyethylene liners from the acetabular shell: a report of four cases.

Authors:  Chancellor F Gray; Ryan E Moore; Gwo-Chin Lee
Journal:  J Bone Joint Surg Am       Date:  2012-05-02       Impact factor: 5.284

9.  Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide.

Authors:  M D Ries; K Weaver; R M Rose; J Gunther; W Sauer; N Beals
Journal:  Clin Orthop Relat Res       Date:  1996-12       Impact factor: 4.176

10.  Independent predictors of failure up to 7.5 years after 35 386 single-brand cementless total hip replacements: a retrospective cohort study using National Joint Registry data.

Authors:  S S Jameson; P N Baker; J Mason; M Rymaszewska; P J Gregg; D J Deehan; M R Reed
Journal:  Bone Joint J       Date:  2013-06       Impact factor: 5.082

View more
  11 in total

1.  Letter to the Editor: Polyethylene Liner Dissociation is a Complication of the DePuy Pinnacle Cup: A Report of 23 Cases.

Authors:  Guillermo Bonilla; Maria Bautista
Journal:  Clin Orthop Relat Res       Date:  2016-03-14       Impact factor: 4.176

2.  Gait instability may indicate liner failure in patients with total hip arthroplasty. A report of three cases.

Authors:  F Díaz-Dilernia; M Lattore; G Zanotti; F Comba; F Piccaluga; M Buttaro
Journal:  Ann R Coll Surg Engl       Date:  2021-08-20       Impact factor: 1.951

3.  Lewinnek Safe Zone References are Frequently Misquoted.

Authors:  Aonnicha Burapachaisri; Ameer Elbuluk; Edem Abotsi; Jim Pierrepont; Seth A Jerabek; Aaron J Buckland; Jonathan M Vigdorchik
Journal:  Arthroplast Today       Date:  2020-11-26

4.  Acetabular Liner Dissociation: A Comparative Study of Two Contemporary Uncemented Acetabular Components.

Authors:  David P Gwynne-Jones; Adeel Memon
Journal:  Arthroplast Today       Date:  2020-06-12

5.  Liner dissociation in a large-diameter ceramic-bearing acetabular component: a report of five cases.

Authors:  Lazaros Kostretzis; Sagi Martinov; Martin Lavigne; Vincent Massé; Pascal-André Vendittoli
Journal:  BMC Musculoskelet Disord       Date:  2022-02-09       Impact factor: 2.362

6.  Acetabular Liner Dissociation following Total Hip Arthroplasty: A Rare but Serious Complication That May Be Easily Misinterpreted in the Emergency Department.

Authors:  Christopher K J O'Neill; Richard J Napier; Owen J Diamond; Seamus O'Brien; David E Beverland
Journal:  Case Rep Emerg Med       Date:  2015-10-12

7.  Third-Generation Medium Cross-Linked Polyethylene Demonstrates Very Low Wear in Total Hip Arthroplasty.

Authors:  Omid Jalali; Travis Scudday; Marie-Claire Fickenscher; Steven Barnett; Robert Gorab
Journal:  Arthroplast Today       Date:  2020-06-02

8.  Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?

Authors:  Adeel R Memon; David Gwynne-Jones
Journal:  Arthroplast Today       Date:  2020-01-09

9.  Titanium Acetabular Component Deformation under Cyclic Loading.

Authors:  Nicholas A Beckmann; Rudi G Bitsch; Theresa Bormann; Steffen Braun; Sebastian Jaeger
Journal:  Materials (Basel)       Date:  2019-12-20       Impact factor: 3.623

10.  Protrusion of a ceramic femoral head through the acetabular metallic cup in total-hip arthroplasty: A case report.

Authors:  Ding Zhao; De-Bao Zhang; Dong-Feng Han; Gui-Shan Gu
Journal:  Medicine (Baltimore)       Date:  2020-06-12       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.