| Literature DB >> 32514421 |
Omid Jalali1, Travis Scudday2, Marie-Claire Fickenscher2, Steven Barnett2, Robert Gorab2.
Abstract
BACKGROUND: Cross-linked polyethylene (XLPE) liners have shown lower wear rates than conventional polyethylene liners in total hip arthroplasty. The primary aim of our study was to report our most recent analysis of wear rates and clinical outcomes of a third-generation XLPE liner. Secondary aims were to investigate the rate of adverse events related to mechanical failure or oxidation of this liner.Entities:
Keywords: Cross-linked polyethylene; Implant design; Outcome; Total hip arthroplasty; Wear
Year: 2020 PMID: 32514421 PMCID: PMC7267711 DOI: 10.1016/j.artd.2020.04.006
Source DB: PubMed Journal: Arthroplast Today ISSN: 2352-3441
Component factors used.
| Head Size (mean ± SD) | 34.6 ± 3.5 mms |
| Femoral head material (count, %) | Ceramic: 116 (43.6%) |
| Metal: 150 (56.4%) | |
| Head size distribution (count, %) | 28 mm: 23 (8.7%) |
| 32 mm: 83 (31.2%) | |
| 36 mm: 128 (48.1%) | |
| 40 mm: 25 (9.4%) | |
| 44 mm: 7 (2.6%) |
Figure 1Anteroposterior pelvis radiographs of a left THA at 6-wk follow-up (a) and 4.9-y follow-up (b). Using the Roman Software, 2 best-fit circles were generated at the edge of the acetabular component (red circles) and the femoral head (blue circles) and the distance between their centers determined (distance between red and blue dots) at each time point. THA, total hip arthroplasty.
Patient characteristics.
| Age (mean ± SD years) | 65.8 ± 9.9 |
| Follow-up time (mean ± SD years) | 5.5 ± 0.8 |
| BMI (mean ± SD kg/m2) | 26.7 ± 4.6 |
| Gender (count, %) | Male: 115 (43%) |
| Female: 151 (57%) | |
| Laterality (count, %) | Right: 145 (55%) |
| Left: 121 (45%) | |
| Bilateral (count, %) | 23 (9%) |
| Diagnoses (count, %) | Avascular necrosis: 9 (3%) |
| Hip dysplasia: 26 (10%) | |
| Osteoarthritis: 228 (86%) | |
| Post-traumatic: 3 (1%) |
BMI, body mass index.
Linear and volumetric wear results.
| Best case (total, mm3) | Best case (rate, mm/y) | Worst case (total, mm3) | Worst case (rate, mm/y) | |
|---|---|---|---|---|
| Linear wear (mean, SD) | 0.017 ± 0.26 | 0.003 ± 0.05 | 0.11 ± 0.16 | 0.02 ± 0.03 |
| Volumetric wear (mean, SD) | 2.31 ± 50.5 | 0.419 ± 9.8 | 19.8 ± 30.7 | 3.75 ± 6.03 |
SD, standard deviation.
Preoperative and postoperative Harris hip scores.
| Harris hip score measures | Preoperative | Latest follow-up |
|---|---|---|
| Functional (mean ± SD) | 20.6 ± 2.4 | 31.0 ± 4.5 |
| Range of motion (mean ± SD) | 4.3 ± 0.7 | 4.8 ± 0.7 |
| Activity (mean ± SD) | 9.0 ± 2.4 | 13.1 ± 1.5 |
| Total (mean ± SD) | 50.9 ± 12.2 | 96.0 ± 7.9 |
SD, standard deviation.
Multivariate analysis of risk factors for accelerated polyethylene wear.
| Patient factors | |
|---|---|
| Age | .840 |
| BMI | .612 |
| Sex | .821 |
| Laterality | .915 |
| Component factors | |
| Femoral head material | .989 |
| Femoral head size | .316 |
| Wear over time | |
| Time since implantation | .936 |
BMI, body mass index.