Dislodgment of polyethylene liners in first and second-generation Harris-Galante acetabular components. A report of eighteen cases.

A González della Valle1, P S Ruzo, S Li, P Pellicci, T P Sculco, E A Salvati.   

Abstract

BACKGROUND: Dislodgment of the polyethylene liner is an increasingly common complication following total hip arthroplasty. The purposes of this study are to present the results in a series of patients with this complication and to analyze the mode of failure.
METHODS: Between November 1995 and January 2001, eighteen patients who had had a total hip arthroplasty presented with dislodgment of the polyethylene liner from a Harris-Galante metal acetabular shell. The medical records, radiographs, operative notes, and retrieved components were reviewed. In addition, scanning electron microscopy was used to study the fractured surfaces in a shell that had four broken tines.
RESULTS: The components had been in situ for an average of seven years (range, three to eleven years). Seventeen components were second generation, and one was first generation. Symptoms developed spontaneously in sixteen patients, during sexual intercourse in one, and following a fall on the hip in one. Radiographs showed eccentric positioning of the head in all of the hips and broken tines in six. All of the shells were well fixed. Treatment consisted of revision of the shell in four patients, exchange of the liner in four, cementation of a new liner into the shell in seven, and cementation of an all-polyethylene cup in three. The liners had severe damage of the rim. Scanning-electron microscopy of the fractured surfaces of four tines revealed a fatigue pattern.
CONCLUSIONS: We believe that, as the liner wears and becomes loose because of an inadequate locking mechanism, progressive micromotion occurs and the load increases on the polyethylene rim until it deforms and/or fractures. Subsequently, nothing prevents the liner from rotating out of the shell. As this mechanism of failure appears to include fatigue failure of the locking tines and wear of the liner, this complication is likely to increase as the components age in situ.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11315784     DOI: 10.2106/00004623-200104000-00010

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  18 in total

1.  Early follow-up for a hybrid total hip arthroplasty using a metal-backed acetabular component designed to reduce "backside" polyethylene wear.

Authors:  Colin E Poole; Shantanu S Patil; Darryl D D'Lima; Clifford W Colwell
Journal:  HSS J       Date:  2005-09

2.  Using nonlinear finite element models to analyse stress distribution during subluxation and torque required for dislocation of newly developed total hip structure after prosthetic impingement.

Authors:  Wei-Min Chi; Chien-Chung Lin; Ying-Jui Ho; Hsiao-Che Lin; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2017-07-01       Impact factor: 2.602

3.  Effects of design parameters of total hip components on the impingement angle and determination of the preferred liner skirt shape with an adequate oscillation angle.

Authors:  Hsiao-Che Lin; Wei-Min Chi; Ying-Jui Ho; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2012-11-29       Impact factor: 2.602

4.  Polyethylene Liner Dissociation Is a Complication of the DePuy Pinnacle Cup: A Report of 23 Cases.

Authors:  Andrew Yun; Emmanuel N Koli; John Moreland; Richard Iorio; John F Tilzey; J Wesley Mesko; Gwo-Chin Lee; Mark Froimson
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

5.  Second-generation modular acetabular components provide fixation at 10 to 16 years.

Authors:  Paul F Lachiewicz; Elizabeth S Soileau
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

6.  Clinical and radiographic results and wear performance in different generations of a cementless porous-coated acetabular cup.

Authors:  E García-Rey; E García-Cimbrelo
Journal:  Int Orthop       Date:  2007-01-31       Impact factor: 3.075

7.  Surface Roughness of CoCr and ZrO(2) Femoral Heads with Metal Transfer: A Retrieval and Wear Simulator Study.

Authors:  Alan W Eberhardt; R Travis McKee; John M Cuckler; Donald W Peterson; Preston R Beck; Jack E Lemons
Journal:  Int J Biomater       Date:  2009-07-01

8.  Retrieval analysis of new-generation yttria-stabilized zirconia femoral heads after total hip arthroplasty.

Authors:  Kiyokazu Fukui; Ayumi Kaneuji; Tanzo Sugimori; Toru Ichiseki; Tadami Matsumoto
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-12-31

9.  Cementless total hip replacement: past, present, and future.

Authors:  Harumoto Yamada; Yasuo Yoshihara; Osamu Henmi; Mitsuhiro Morita; Yuichiro Shiromoto; Tomoki Kawano; Arihiko Kanaji; Kennichi Ando; Masato Nakagawa; Naoto Kosaki; Eiichi Fukaya
Journal:  J Orthop Sci       Date:  2009-04-01       Impact factor: 1.601

10.  Retrieval analysis of Harris-Galante I and II acetabular liners in situ for more than 10 years.

Authors:  Keisha French; Rebecca Moore; Heather Gawel; Steven M Kurtz; Matthew J Kraay; Ke Xie; Victor M Goldberg; Clare M Rimnac
Journal:  Acta Orthop       Date:  2012-08-10       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.