Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study.

Senay Asik Nacaroglu1, Emine Seker Un1, Mehmet Giray Ersoz1, Yelda Tasci2.   

Abstract

AIM: To evaluate the intraocular pressure (IOP) measurements by Icare rebound tonometer over a contact lens in comparison with Goldmann applanation tonometry (GAT).
METHODS: Fifty patients using contact lens were included in this study. One of the eyes of the patients was selected randomly and their IOP were measured by rebound tonometer with and without contact lens (RTCL, RT respectively) and by GAT, as well as their central corneal thickness (CCT) by optical pachymeter. The results of both methods were compared by correlation analysis, general linear method repeated measure and Bland-Altman analysis.
RESULTS: Mean IOP values measured by RTCL, RT and GAT were 15.68±3.7, 14.50±3.4 and 14.16±2.8 (P<0.001), respectively. Mean IOP by RTCL was significantly higher than the measurements implemented by RT and GAT (P<0.001), while there was no difference between the measurements by GAT and RT (P=0.629). There was a good level of positive correlation between GAT and RTCL as well as RT (r=0.786 P<0.001, r=0.833 P<0.001, respectively). We have observed that CCT increase did not show any correlation with the differences of the measurements between RTCL and RT (P=0.329), RTCL and GAT (P=0.07) as well as RT and GAT (P=0.189) in linear regression model.
CONCLUSION: The average of the measurements over contact lens by rebound tonometer was found to be higher than what was measured by GAT. Although this difference is statistically significant, it may be clinically negligible in the normal population.

Entities:  

Keywords:  contact lens; intraocular pressure; rebound tonometer

Year:  2015        PMID: 26086004      PMCID: PMC4458659          DOI: 10.3980/j.issn.2222-3959.2015.03.18

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  20 in total

1.  Effect of contact lens removal or displacement on intraocular pressure.

Authors:  J A Khan; C E Graham
Journal:  Arch Ophthalmol       Date:  1991-06

2.  Comparing applanation tonometry and rebound tonometry in glaucomatous and ocular hypertensive eyes.

Authors:  Marino Marini; Stefano Da Pozzo; Agostino Accardo; Tiziana Canziani
Journal:  Eur J Ophthalmol       Date:  2011 May-Jun       Impact factor: 2.597

3.  Pascal, ICare and Goldmann applanation tonometry--a comparative study.

Authors:  Gauti Jóhannesson; Per Hallberg; Anders Eklund; Christina Lindén
Journal:  Acta Ophthalmol       Date:  2008-09       Impact factor: 3.761

4.  Rebound tonometer compared with goldmann tonometer in normal and pathologic corneas.

Authors:  Javier Moreno-Montañés; Noelia García; Ana Fernández-Hortelano; Alfredo García-Layana
Journal:  Cornea       Date:  2007-05       Impact factor: 2.651

5.  Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice.

Authors:  Jose M Martinez-de-la-Casa; Maria Jimenez-Santos; Federico Saenz-Frances; Maria Matilla-Rodero; Carmen Mendez-Hernandez; Rocio Herrero-Vanrell; Julian Garcia-Feijoo
Journal:  Acta Ophthalmol       Date:  2009-11-09       Impact factor: 3.761

6.  The effect of hydrogel and silicone hydrogel contact lenses on the measurement of intraocular pressure with rebound tonometry.

Authors:  Fabrizio Zeri; Paolo Calcatelli; Bernardo Donini; Luigi Lupelli; Luciana Zarrilli; Peter G Swann
Journal:  Cont Lens Anterior Eye       Date:  2011-06-01       Impact factor: 3.077

Review 7.  Sources of error with use of Goldmann-type tonometers.

Authors:  M M Whitacre; R Stein
Journal:  Surv Ophthalmol       Date:  1993 Jul-Aug       Impact factor: 6.048

8.  Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry.

Authors:  Jose M Martinez-de-la-Casa; Julian Garcia-Feijoo; Eva Vico; Ana Fernandez-Vidal; Jose M Benitez del Castillo; Mohamed Wasfi; J Garcia-Sanchez
Journal:  Ophthalmology       Date:  2006-09-25       Impact factor: 12.079

9.  Relationship of central corneal thickness and intraocular pressure by iCare rebound tonometer.

Authors:  Aparna Rao; Mukesh Kumar; B Prakash; Gopal Varshney
Journal:  J Glaucoma       Date:  2014-08       Impact factor: 2.503

10.  Clinical comparison of the Icare tonometer and Goldmann applanation tonometry.

Authors:  Nima Pakrou; Tim Gray; Richard Mills; John Landers; Jamie Craig
Journal:  J Glaucoma       Date:  2008 Jan-Feb       Impact factor: 2.503

View more
  1 in total

1.  Rebound Tonometry over Soft Contact Lenses.

Authors:  Iva Krolo; Boze Mihaljevic; Aida Kasumovic; Freja Bagatin; Maja Malenica Ravlic; Jelena Skunca Herman
Journal:  Acta Inform Med       Date:  2020-09
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.