Effect of orthodontic brackets and different wires on radiofrequency heating and magnetic field interactions during 3-T MRI.

S Görgülü1, S Ayyildiz, K Kamburoglu, S Gökçe, T Ozen.   

Abstract

OBJECTIVES: To evaluate the heating and magnetic field interactions of fixed orthodontic appliances with different wires and ligaments in a 3-T MRI environment and to estimate the safety of these orthodontic materials.
METHODS: 40 non-carious extracted human maxillary teeth were embedded in polyvinyl chloride boxes, and orthodontic brackets were bonded. Nickel-titanium and stainless steel arch wires, and elastic and stainless steel ligaments were used to obtain four experimental groups in total. Specimens were evaluated at 3 T for radiofrequency heating and magnetic field interactions. Radiofrequency heating was evaluated by placing specimens in a cylindrical plastic container filled with isotonic solution and measuring changes in temperature after T1 weighted axial sequencing and after completion of all sequences. Translational attraction and torque values of specimens were also evaluated. One-way ANOVA test was used to compare continuous variables of temperature change. Significance was set at p < 0.05.
RESULTS: None of the groups exhibited excessive heating (highest temperature change: <3.04 °C), with the maximum increase in temperature observed at the end of the T1 weighted axial sequence. Magnetic field interactions changed depending on the material used. Although the brackets presented minor interactions that would not cause movement in situ, nickel-titanium and stainless steel wires presented great interactions that may pose a risk for the patient.
CONCLUSIONS: The temperature changes of the specimens were considered to be within acceptable ranges. With regard to magnetic field interactions, brackets can be considered "MR safe"; however, it would be safe to replace the wires before MRI.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24257741      PMCID: PMC4064615          DOI: 10.1259/dmfr.20130356

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  20 in total

1.  Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system.

Authors:  Frank G Shellock
Journal:  J Magn Reson Imaging       Date:  2002-12       Impact factor: 4.813

Review 2.  Magnetic resonance safety update 2002: implants and devices.

Authors:  Frank G Shellock
Journal:  J Magn Reson Imaging       Date:  2002-11       Impact factor: 4.813

3.  The benefits of aesthetic orthodontic brackets in patients requiring multiple MRI scanning.

Authors:  T M J Harris; M R Faridrad; J A S Dickson
Journal:  J Orthod       Date:  2006-06

4.  Programmable CSF shunt valve: in vitro assessment of MR imaging safety at 3T.

Authors:  F G Shellock; R Habibi; J Knebel
Journal:  AJNR Am J Neuroradiol       Date:  2006-03       Impact factor: 3.825

5.  Effect of MRI on the mechanical properties of stainless steel orthodontic wires.

Authors:  Morteza Oshagh; Mohammad Rahim Hematian; Shoaleh Shahidi; Negar Feizi; Farnaz Bayani; Lida Pishbin
Journal:  World J Orthod       Date:  2010

6.  Radiofrequency heating of metallic dental devices during 3.0 T MRI.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishigami
Journal:  Dentomaxillofac Radiol       Date:  2013-03-21       Impact factor: 2.419

7.  Radiofrequency heating and magnetic field interactions of fixed partial dentures during 3-tesla magnetic resonance imaging.

Authors:  Simel Ayyıldız; Kıvanç Kamburoğlu; Cumhur Sipahi; Sema Murat; Serkan Görgülü; Bülent Pişkin
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2013-09-07

8.  Temporomandibular joint remodeling in adolescents and young adults during Herbst treatment: A prospective longitudinal magnetic resonance imaging and cephalometric radiographic investigation.

Authors:  S Ruf; H Pancherz
Journal:  Am J Orthod Dentofacial Orthop       Date:  1999-06       Impact factor: 2.650

9.  Dental materials and magnetic resonance imaging.

Authors:  Hana Hubálková; Karel Hora; Zdenek Seidl; Jan Krásenský
Journal:  Eur J Prosthodont Restor Dent       Date:  2002-09

Review 10.  MR procedures: biologic effects, safety, and patient care.

Authors:  Frank G Shellock; John V Crues
Journal:  Radiology       Date:  2004-07-29       Impact factor: 11.105

View more
  18 in total

1.  Effect of orthodontic brackets and different wires on radiofrequency heating and magnetic field interactions during 3.0-T MRI.

Authors:  Seyed Mj Mortazavi; Maryam Paknahad
Journal:  Dentomaxillofac Radiol       Date:  2015-10-15       Impact factor: 2.419

2.  Effect of orthodontic brackets and different wires on radiofrequency heating and magnetic field interactions during 3.0-T MRI: Author response.

Authors:  Serkan Görgülü; Simel Ayyıldız; Kıvanç Kamburoğlu; Sıla Gökçe; Tuncer Ozen
Journal:  Dentomaxillofac Radiol       Date:  2015-10-15       Impact factor: 2.419

3.  Induced magnetic moment in stainless steel components of orthodontic appliances in 1.5 T MRI scanners.

Authors:  Zhiyue J Wang; Nancy K Rollins; Hui Liang; Yong Jong Park
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

4.  Effect of magnetic resonance imaging on microleakage of amalgam restorations: an in vitro study.

Authors:  S M J Mortazavi; Maryam Paknahad
Journal:  Dentomaxillofac Radiol       Date:  2015-08-17       Impact factor: 2.419

Review 5.  Magnetic resonance imaging in children with implants.

Authors:  Camilo Jaimes; Diane Biaggotti; Gayathri Sreedher; Apeksha Chaturvedi; Michael M Moore; Amy R Danehy
Journal:  Pediatr Radiol       Date:  2021-04-19

6.  MRI with intraoral orthodontic appliance-a comparative in vitro and in vivo study of image artefacts at 1.5 T.

Authors:  C Zachriat; P Asbach; K I Blankenstein; I Peroz; F H Blankenstein
Journal:  Dentomaxillofac Radiol       Date:  2015-03-03       Impact factor: 2.419

7.  [Metal objects of the head and neck region in magnetic resonance imaging : Survey among radiologists].

Authors:  Ulrike Kielburg; Felix H Blankenstein
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

Review 8.  MRI in Dentistry- A Future Towards Radiation Free Imaging - Systematic Review.

Authors:  Lav Kumar Niraj; Basavaraj Patthi; Ashish Singla; Ritu Gupta; Irfan Ali; Kuldeep Dhama; Jishnu Krishna Kumar; Monika Prasad
Journal:  J Clin Diagn Res       Date:  2016-10-01

9.  Magnetic permeability as a predictor of the artefact size caused by orthodontic appliances at 1.5 T magnetic resonance imaging.

Authors:  Felix H Blankenstein; Patrick Asbach; Florian Beuer; Johannes Glienke; Stefan Mayer; Christine Zachriat
Journal:  Clin Oral Investig       Date:  2016-03-17       Impact factor: 3.573

Review 10.  Should the orthodontic brackets always be removed prior to magnetic resonance imaging (MRI)?

Authors:  Arash Poorsattar-Bejeh Mir; Manouchehr Rahmati-Kamel
Journal:  J Oral Biol Craniofac Res       Date:  2015-09-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.