Passively scattered proton beam entrance dosimetry with a plastic scintillation detector.

Landon Wootton1, Charles Holmes, Narayan Sahoo, Sam Beddar.   

Abstract

We tested the feasibility of using plastic scintillation detectors (PSDs) for proton entrance dosimetry. A PSD built with BCF-12 scintillating fiber was used to measure the absolute entrance dose of a passively scattered proton beam for energies ranging from 140 to 250 MeV, and for a range of spread out Bragg peak (SOBP) widths at two energies, to quantify the effect of ionization quenching on the response of the detector and to determine the necessity of Cerenkov radiation correction in proton beams. The overall accuracy and precision of the PSD was evaluated by measuring lateral beam profiles and comparing the results with profiles measured using film. The PSD under-responded owing to ionization quenching, exhibiting approximately a 7% loss of signal at the highest energy studied (250 MeV) and a 10% loss of signal at the lowest energy studied (140 MeV). For a given nominal energy, varying the SOBP width did not significantly alter the response of the PSD. Cerenkov radiation contributed negligibly to the PSD signal and can be safely ignored without introducing more than 1% error in the measured dose. Profiles measured with the PSD and film agreed to within the uncertainty of the detector, demonstrating good relative accuracy. Although correction factors were necessary to account for ionization quenching, the magnitude of the correction varied minimally over a broad range of energies; PSDs therefore represent a practical detector for proton entrance dosimetry.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25591037      PMCID: PMC4326057          DOI: 10.1088/0031-9155/60/3/1185

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  17 in total

1.  Gafchromic EBT film dosimetry in proton beams.

Authors:  Li Zhao; Indra J Das
Journal:  Phys Med Biol       Date:  2010-04-29       Impact factor: 3.609

2.  Dosimetric characteristics of a single use MOSFET dosimeter for in vivo dosimetry in proton therapy.

Authors:  Chee-Wai Cheng; Mark Wolanski; Q Zhao; Leia Fanelli; Archana Gautam; Daniel Pack; Indra J Das
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

3.  Radiochromic film dosimetry for clinical proton beams.

Authors:  S M Vatnitsky
Journal:  Appl Radiat Isot       Date:  1997-05       Impact factor: 1.513

4.  Early findings on toxicity of proton beam therapy with concurrent chemotherapy for nonsmall cell lung cancer.

Authors:  Samir Sejpal; Ritsuko Komaki; Anne Tsao; Joe Y Chang; Zhongxing Liao; Xiong Wei; Pamela K Allen; Charles Lu; Michael Gillin; James D Cox
Journal:  Cancer       Date:  2011-01-24       Impact factor: 6.860

5.  Toxicity and patterns of failure of adaptive/ablative proton therapy for early-stage, medically inoperable non-small cell lung cancer.

Authors:  Joe Y Chang; Ritsuko Komaki; Hong Y Wen; Beth De Gracia; Jaques B Bluett; Mary F McAleer; Stephen G Swisher; Michael Gillin; Radhe Mohan; James D Cox
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-01-20       Impact factor: 7.038

6.  Spectral discrimination of Cerenkov radiation in scintillating dosimeters.

Authors:  A M Frelin; J M Fontbonne; G Ban; J Colin; M Labalme; A Batalla; A Isambert; A Vela; T Leroux
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

7.  Proton beam therapy as a nonsurgical approach to mucosal melanoma of the head and neck: a pilot study.

Authors:  Sadamoto Zenda; Mitsuhiko Kawashima; Teiji Nishio; Ryosuke Kohno; Keiji Nihei; Masakatsu Onozawa; Satoko Arahira; Takashi Ogino
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-13       Impact factor: 7.038

8.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

9.  Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration.

Authors:  A S Beddar; T R Mackie; F H Attix
Journal:  Phys Med Biol       Date:  1992-10       Impact factor: 3.609

10.  Protective effect of transparent film dressing on proton therapy induced skin reactions.

Authors:  Jonathan T Whaley; Maura Kirk; Keith Cengel; James McDonough; Justin Bekelman; John P Christodouleas
Journal:  Radiat Oncol       Date:  2013-01-24       Impact factor: 3.481

View more
  3 in total

1.  Exradin W1 plastic scintillation detector for in vivo skin dosimetry in passive scattering proton therapy.

Authors:  Fahed Alsanea; Landon Wootton; Narayan Sahoo; Rajat Kudchadker; Usama Mahmood; Sam Beddar
Journal:  Phys Med       Date:  2018-02-27       Impact factor: 2.685

2.  Novel Gd3+-doped silica-based optical fiber material for dosimetry in proton therapy.

Authors:  C Hoehr; A Morana; O Duhamel; B Capoen; M Trinczek; P Paillet; C Duzenli; M Bouazaoui; G Bouwmans; A Cassez; Y Ouerdane; A Boukenter; H El Hamzaoui; S Girard
Journal:  Sci Rep       Date:  2019-11-08       Impact factor: 4.379

3.  Dosimetric properties of a newly developed thermoluminescent sheet-type dosimeter for clinical proton beams.

Authors:  Takahiro Kato; Tatsuhiko Sagara; Shinya Komori; Ryohei Kato; Akihiko Takeuchi; Yuki Narita
Journal:  J Appl Clin Med Phys       Date:  2021-03-15       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.