Literature DB >> 20879587

Dosimetric characteristics of a single use MOSFET dosimeter for in vivo dosimetry in proton therapy.

Chee-Wai Cheng1, Mark Wolanski, Q Zhao, Leia Fanelli, Archana Gautam, Daniel Pack, Indra J Das.   

Abstract

PURPOSE: Entrance dose (or skin dose) is an important part of patient quality assurance in external beam radiation therapy. However, entrance dose verification in proton beam is not routinely performed. In this study, the OneDose single use MOSFET detector system for in vivo dosimetry measurement in proton therapy is investigated.
METHODS: Using a solid water phantom, several fundamental dosimetric characteristics of the OneDose system are studied with a proton beam: The reproducibility (consistency) of the dosimeter, the linearity with dose and dose rate, energy dependence, directional dependence, LET dependence, and fading (delay readout with time) is studied.
RESULTS: OneDose detectors show dose and dose rate linearity but exhibit pronounced energy dependence at depth and a large variation in dose response with LET. On the other hand, the detector response remain relatively constant (within 3%) at surface over a wide range of energies. There is also a slight angular dependence (about 2%) up to 60 degrees angle of incidence. However, detector orientation such that incidence along the long axis of the detector should be avoided as the proton beam will have to traverse a large amount of the copper backing. Since most in vivo dosimetry involves entrance dose measurement, the OneDose at surface appears to be well suited for such application. OneDose exhibits small intrabatch variation (< or = 2% at one SD) indicating that it is only necessary to calibration a few detectors from each batch. The interbatch variation is generally within 3%.
CONCLUSIONS: The small detector size and its relatively flexible design of OneDose allow dose measurement to be performed on a curved surface or in small cavities that is otherwise difficult with the conventional diode detectors. The slight drawback in its angular dependence can be easily handled by angular dependence table. However, since OneDose is a single use detector, the intra-batch consistency must be verified before the remaining detectors from the same batch could be used for in vivo dosimetry. It is advisable that the detectors from the same batch be taken for the same application to reduce the dosimetric uncertainty. For detectors from different batches, inter-batch consistency should also be verified to obtain reliable results. OneDose provides an opportunity to measure in vivo dose with proton beam within acceptable clinical criterion of +/- (5.0%-6.5%).

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20879587     DOI: 10.1118/1.3467753

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Passively scattered proton beam entrance dosimetry with a plastic scintillation detector.

Authors:  Landon Wootton; Charles Holmes; Narayan Sahoo; Sam Beddar
Journal:  Phys Med Biol       Date:  2015-01-15       Impact factor: 3.609

2.  Is the in vivo dosimetry with the OneDosePlusTM system able to detect intra-fraction motion? A retrospective analysis of in vivo data from breast and prostate patients.

Authors:  Maria Daniela Falco; Marco D'Andrea; Alessia Lo Bosco; Mauro Rebuzzi; Elisabetta Ponti; Barbara Tolu; Grazia Tortorelli; Rosaria Barbarino; Luana Di Murro; Riccardo Santoni
Journal:  Radiat Oncol       Date:  2012-06-20       Impact factor: 3.481

3.  In vivo proton dosimetry using a MOSFET detector in an anthropomorphic phantom with tissue inhomogeneity.

Authors:  Ryosuke Kohno; Kenji Hotta; Kana Matsubara; Shie Nishioka; Taeko Matsuura; Mitsuhiko Kawashima
Journal:  J Appl Clin Med Phys       Date:  2012-03-08       Impact factor: 2.102

4.  Proton dose distribution measurements using a MOSFET detector with a simple dose-weighted correction method for LET effects.

Authors:  Ryosuke Kohno; Kenji Hotta; Taeko Matsuura; Kana Matsubara; Shie Nishioka; Teiji Nishio; Mitsuhiko Kawashima; Takashi Ogino
Journal:  J Appl Clin Med Phys       Date:  2011-04-04       Impact factor: 2.102

5.  Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams.

Authors:  A Sathish Kumar; S D Sharma; B Paul Ravindran
Journal:  J Med Phys       Date:  2014-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.