| Literature DB >> 33720527 |
Takahiro Kato1,2, Tatsuhiko Sagara2, Shinya Komori2, Ryohei Kato2, Akihiko Takeuchi2, Yuki Narita2.
Abstract
PURPOSE: This study aimed to evaluate the dosimetric properties of a newly developed thermoluminescent sheet-type dosimeter (TLD-sheet) for clinical proton beams.Entities:
Keywords: in vivo skin dosimetry; proton therapy; quality assurance; thermoluminescent dosimeter
Year: 2021 PMID: 33720527 PMCID: PMC8035564 DOI: 10.1002/acm2.13222
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1The appearances of various size of the TLD‐sheets (left) and the TLD reader (TLDR‐1) (right). The TLD‐sheet is white and semi‐transparent. Data can be acquired using the TLDR‐1 within 15 min.
Fig. 2Schematic diagram of the TLD reader. The reading cassette is shown in the measuring position (left) or cooling‐down position (right). After the cassette is inserted via the left entrance, the heating block adheres to the flexible cover of the cassette and increases the temperature of the TLD‐sheet. The thermoluminescence light is detected using a charge‐coupled device camera during the time determined by the user. After the measurement, the reading cassette is transferred to the cooling‐down room until the temperature becomes low enough for safe manual removal.
Fig. 3Fading characteristics of the TLD‐sheet. The proton energy was 150 MeV, and the sheets were placed at the center of the spread‐out Bragg peak. The error bar indicates the standard deviation.
Fig. 4Dose linearity of the TLD‐sheet. The two different symbols represent measurements obtained in the plateau region of a 210‐MeV unmodulated beam and at the center of the 150‐MeV beam with a spread‐out Bragg peak (SOBP) size of 20 mm. The error bar indicates the standard deviation.
Fig. 5Sample 2‐dimensional image of the TLD‐sheet at the plateau region of the 150‐MeV beam with a spread‐out Bragg peak size of 60 mm (left). Normalized horizontal profile (middle) and vertical profile (right) through the center of the TLD‐sheet. The directions are shown as yellow arrows on the left image and comparison with those measured with the PinPoint 3D Ion Chamber is shown.
Fig. 6Comparison of (a) the unmodulated and (b) modulated percentage depth doses obtained using the TLD‐sheet and the Advanced Markus (A. Markus) chamber.