Simulation study comparing high-purity germanium and cadmium zinc telluride detectors for breast imaging.

D L Campbell1, T E Peterson.   

Abstract

We conducted simulations to compare the potential imaging performance for breast cancer detection with High-Purity Germanium (HPGe) and Cadmium Zinc Telluride (CZT) systems with 1% and 3.8% energy resolution at 140 keV, respectively. Using the Monte Carlo N-Particle (MCNP5) simulation package, we modelled both 5 mm-thick CZT and 10 mm-thick HPGe detectors with the same parallel-hole collimator for the imaging of a breast/torso phantom. Simulated energy spectra were generated, and planar images were created for various energy windows around the 140 keV photopeak. Relative sensitivity and scatter and the torso fractions were calculated along with tumour contrast and signal-to-noise ratios (SNR). Simulations showed that utilizing a ±1.25% energy window with an HPGe system better suppressed torso background and small-angle scattered photons than a comparable CZT system using a -5%/+10% energy window. Both systems provided statistically similar contrast and SNR, with HPGe providing higher relative sensitivity. Lowering the counts of HPGe images to match CZT count density still yielded equivalent contrast between HPGe and CZT. Thus, an HPGe system may provide equivalent breast imaging capability at lower injected radioactivity levels when acquiring for equal imaging time.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25360792      PMCID: PMC4237407          DOI: 10.1088/0031-9155/59/22/7059

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  11 in total

1.  Performance of a PSPMT based detector for scintimammography.

Authors:  M B Williams; M B Williams; A R Goode; V Galbis-Reig; S Majewski; A G Weisenberger; R Wojcik
Journal:  Phys Med Biol       Date:  2000-03       Impact factor: 3.609

2.  Charge transport in arrays of semiconductor gamma-ray detectors.

Authors: 
Journal:  Phys Rev Lett       Date:  1995-07-03       Impact factor: 9.161

Review 3.  Radiation doses and cancer risks from breast imaging studies.

Authors:  R Edward Hendrick
Journal:  Radiology       Date:  2010-08-24       Impact factor: 11.105

4.  Optimization of a parallel hole collimator/CdZnTe gamma-camera architecture for scintimammography.

Authors:  Charlotte Robert; Guillaume Montémont; Véronique Rebuffel; Loïck Verger; Irène Buvat
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

5.  Uptake of the cation hexakis(2-methoxyisobutylisonitrile)-technetium-99m by human carcinoma cell lines in vitro.

Authors:  L I Delmon-Moingeon; D Piwnica-Worms; A D Van den Abbeele; B L Holman; A Davison; A G Jones
Journal:  Cancer Res       Date:  1990-04-01       Impact factor: 12.701

6.  Evaluation of a small cadmium zinc telluride detector for scintimammography.

Authors:  Bryon Mueller; Michael K O'Connor; Ira Blevis; Deborah J Rhodes; Robin Smith; Douglas A Collins; Stephen W Phillips
Journal:  J Nucl Med       Date:  2003-04       Impact factor: 10.057

7.  Investigation of basic imaging properties in digital radiography. 6. MTFs of II-TV digital imaging systems.

Authors:  H Fujita; K Doi; M L Giger
Journal:  Med Phys       Date:  1985 Nov-Dec       Impact factor: 4.071

8.  Characterization of a high-purity germanium detector for small-animal SPECT.

Authors:  Lindsay C Johnson; Desmond L Campbell; Ethan L Hull; Todd E Peterson
Journal:  Phys Med Biol       Date:  2011-08-18       Impact factor: 3.609

9.  Effects of age, breast density, ethnicity, and estrogen replacement therapy on screening mammographic sensitivity and cancer stage at diagnosis: review of 183,134 screening mammograms in Albuquerque, New Mexico.

Authors:  R D Rosenberg; W C Hunt; M R Williamson; F D Gilliland; P W Wiest; C A Kelsey; C R Key; M N Linver
Journal:  Radiology       Date:  1998-11       Impact factor: 11.105

10.  CZT detectors: How important is energy resolution for nuclear breast imaging?

Authors:  Carrie B Hruska; Michael K O'connor
Journal:  Phys Med       Date:  2006       Impact factor: 2.685

View more
  3 in total

1.  An energy-optimized collimator design for a CZT-based SPECT camera.

Authors:  Fenghua Weng; Srijeeta Bagchi; Yunlong Zan; Qiu Huang; Youngho Seo
Journal:  Nucl Instrum Methods Phys Res A       Date:  2016-01-11       Impact factor: 1.455

2.  Molecular Breast Imaging using Synthetic Projections from High-Purity Germanium Detectors: A Simulation Study.

Authors:  Desmond Campbell; Todd Peterson
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2017-07-11

3.  Pinhole X-ray fluorescence imaging of gadolinium and gold nanoparticles using polychromatic X-rays: a Monte Carlo study.

Authors:  Seongmoon Jung; Wonmo Sung; Sung-Joon Ye
Journal:  Int J Nanomedicine       Date:  2017-08-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.