Optimization of a parallel hole collimator/CdZnTe gamma-camera architecture for scintimammography.

Charlotte Robert1, Guillaume Montémont, Véronique Rebuffel, Loïck Verger, Irène Buvat.   

Abstract

PURPOSE: Small field-of-view CdZnTe (CZT) gamma cameras are increasingly studied for breast lesion detection to complement mammography or ultrasonographic findings. However, in classical collimation configurations, they remain limited by the trade-off between spatial resolution and sensitivity. The HiSens architecture was proposed to overcome these limitations. Using an accurate 3D localization of the interactions inside the detector, this architecture leads to a gain in sensitivity without loss in spatial resolution. In this article, the relevance of the HiSens architecture for planar scintimammography is studied.
METHODS: A detective quantum efficiency (DQE) computation method is developed and used to optimize the dimensioning of a parallel hole collimator dedicated to scintimammography. Based on the DQE curves, the impact of the collimator-to-detector distance is studied. Two algorithms are proposed to combine data acquired with different collimator-to-detector distances.
RESULTS: It is shown that CZT detector virtual pixelization increases system sensitivity by 3.3 while preserving a standard LEHR spatial resolution. The introduction of a gap between the CZT detector and the collimator is useful to modulate the DQE curve shape. The combination of data acquired using different gaps in the image formation process leads to enhanced restoration of the frequency content of the images, resulting in image contrast and spatial resolution improvements.
CONCLUSIONS: Acquisition duration or injected activity could be markedly reduced if the HiSens architecture with an appropriate collimator-detector gap were used.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21626915     DOI: 10.1118/1.3560423

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Simulation study comparing high-purity germanium and cadmium zinc telluride detectors for breast imaging.

Authors:  D L Campbell; T E Peterson
Journal:  Phys Med Biol       Date:  2014-10-31       Impact factor: 3.609

Review 2.  Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging.

Authors:  Karen Van Audenhaege; Roel Van Holen; Stefaan Vandenberghe; Christian Vanhove; Scott D Metzler; Stephen C Moore
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

3.  Molecular Breast Imaging using Synthetic Projections from High-Purity Germanium Detectors: A Simulation Study.

Authors:  Desmond Campbell; Todd Peterson
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2017-07-11

4.  Technical Note: Nuclear imaging with an x-ray flat panel detector: A proof-of-concept study.

Authors:  Martijn M A Dietze; Wilco J C Koppert; Rob van Rooij; Hugo W A M de Jong
Journal:  Med Phys       Date:  2020-05-08       Impact factor: 4.071

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.