An energy-optimized collimator design for a CZT-based SPECT camera.

Fenghua Weng1, Srijeeta Bagchi2, Yunlong Zan1, Qiu Huang1, Youngho Seo3.   

Abstract

In single photon emission computed tomography, it is a challenging task to maintain reasonable performance using only one specific collimator for radio-tracers over a broad spectrum of diagnostic photon energies, since photon scatter and penetration in a collimator differ with the photon energy. Frequent collimator exchanges are inevitable in daily clinical SPECT imaging, which hinders throughput while subjecting the camera to operational errors and damage. Our objective is to design a collimator, which independent of the photon energy performs reasonably well for commonly used radiotracers with low- to medium-energy levels of gamma emissions. Using the Geant4 simulation toolkit, we simulated and evaluated a parallel-hole collimator mounted to a CZT detector. With the pixel-geometry-matching collimation, the pitch of the collimator hole was fixed to match the pixel size of the CZT detector throughout this work. Four variables, hole shape, hole length, hole radius/width and the source-to-collimator distance were carefully studied. Scatter and penetration of the collimator, sensitivity and spatial resolution of the system were assessed for four radionuclides including 57Co, 99m Tc, 123I and 111In, with respect to the aforementioned four variables. An optimal collimator was then decided upon such that it maximized the total relative sensitivity (TRS) for the four considered radionuclides while other performance parameters, such as scatter, penetration and spatial resolution, were benchmarked to prevalent commercial scanners and collimators. Digital phantom studies were also performed to validate the system with the optimal square-hole collimator (23 mm hole length, 1.28 mm hole width, 0.32 mm septal thickness) in terms of contrast, contrast-to-noise ratio and recovery ratio. This study demonstrates promise of our proposed energy-optimized collimator to be used in a CZT-based gamma camera, with comparable or even better imaging performance versus commercial collimators such as low-energy high resolution (LEHR) and medium energy general purpose (MEGP) collimators.

Entities:  

Keywords:  CZT detector; Monte Carlo simulation; SPECT; collimator design

Year:  2016        PMID: 26640308      PMCID: PMC4665111          DOI: 10.1016/j.nima.2015.09.115

Source DB:  PubMed          Journal:  Nucl Instrum Methods Phys Res A        ISSN: 0168-9002            Impact factor:   1.455


  15 in total

1.  Simulation study comparing high-purity germanium and cadmium zinc telluride detectors for breast imaging.

Authors:  D L Campbell; T E Peterson
Journal:  Phys Med Biol       Date:  2014-10-31       Impact factor: 3.609

2.  Simulation-based evaluation and optimization of a new CdZnTe gamma-camera architecture (HiSens).

Authors:  Charlotte Robert; Guillaume Montémont; Véronique Rebuffel; Irène Buvat; Lucie Guérin; Loïck Verger
Journal:  Phys Med Biol       Date:  2010-04-19       Impact factor: 3.609

Review 3.  Collimator design for single photon emission tomography.

Authors:  S C Moore; K Kouris; I Cullum
Journal:  Eur J Nucl Med       Date:  1992

4.  A Monte Carlo and physical phantom evaluation of quantitative In-111 SPECT.

Authors:  Bin He; Yong Du; Xiyun Song; W Paul Segars; Eric C Frey
Journal:  Phys Med Biol       Date:  2005-08-24       Impact factor: 3.609

5.  Recent advances in SPECT imaging.

Authors:  Mark T Madsen
Journal:  J Nucl Med       Date:  2007-04       Impact factor: 10.057

6.  Calculation and validation of the use of effective attenuation coefficient for attenuation correction in In-111 SPECT.

Authors:  Youngho Seo; Kenneth H Wong; Bruce H Hasegawa
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

7.  A four-pixel matched collimator for high-sensitivity SPECT imaging.

Authors:  Atsuro Suzuki; Wataru Takeuchi; Takafumi Ishitsu; Katsutoshi Tsuchiya; Yuichiro Ueno; Keiji Kobashi
Journal:  Phys Med Biol       Date:  2013-03-11       Impact factor: 3.609

8.  Collimator performance evaluation for In-111 SPECT using a detection/localization task.

Authors:  Yihuan Lu; Lin Chen; Gene Gindi
Journal:  Phys Med Biol       Date:  2014-02-07       Impact factor: 3.609

9.  Optimal collimator choice for sequential iodine-123 and technetium-99m imaging.

Authors:  F D Geeter; P R Franken; M Defrise; H Andries; E Saelens; A Bossuyt
Journal:  Eur J Nucl Med       Date:  1996-07

10.  High-sensitivity brain SPECT system using cadmium telluride (CdTe) semiconductor detector and 4-pixel matched collimator.

Authors:  Atsuro Suzuki; Wataru Takeuchi; Takafumi Ishitsu; Katsutoshi Tsuchiya; Yuichi Morimoto; Yuichiro Ueno; Keiji Kobashi; Naoki Kubo; Tohru Shiga; Nagara Tamaki
Journal:  Phys Med Biol       Date:  2013-10-18       Impact factor: 3.609

View more
  1 in total

1.  Evaluation of a variable-aperture full-ring SPECT system using large-area pixelated CZT modules: A simulation study for brain SPECT applications.

Authors:  Yoonsuk Huh; Jaewon Yang; Odera U Dim; Yonggang Cui; Weijie Tao; Qiu Huang; Grant T Gullberg; Youngho Seo
Journal:  Med Phys       Date:  2021-03-30       Impact factor: 4.071

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.