Matthew D Lew1, W E Moerner. 1. Departments of Chemistry and ‡Electrical Engineering, Stanford University , Stanford, California 94305, United States.
Abstract
Many single nanoemitters such as fluorescent molecules produce dipole radiation that leads to systematic position errors in both particle tracking and super-resolution microscopy. Via vectorial diffraction equations and simulations, we show that imaging only azimuthally polarized light in the microscope naturally avoids emission from the z-component of the transition dipole moment, resulting in negligible localization errors for all emitter orientations and degrees of objective lens misfocus. Furthermore, localization accuracy is maintained even in the presence of aberrations resulting from imaging in mismatched media.
Many single nanoemitters such as fluorescent molecules produce dipole radiation that leads to systematic position errors in both particle tracking and super-resolution microscopy. Via vectorial diffraction equations and simulations, we show that imaging only azimuthally polarized light in the microscope naturally avoids emission from the z-component of the transition dipole moment, resulting in negligible localization errors for all emitter orientations and degrees of objective lens misfocus. Furthermore, localization accuracy is maintained even in the presence of aberrations resulting from imaging in mismatched media.
Single-molecule localization
microscopy has made possible the nanoscale imaging of biological and
material specimens with visible light, elucidating previously unobserved
cellular structures[1−7] and polymer structural detail[8,9] with resolution beyond
the diffraction limit (∼λ/2 ≈ 250 nm). While many
different mechanisms have been utilized since 2006 to achieve super-resolution,
such as (F)PALM,[10,11] STORM,[12] and PAINT,[13] this family of methods relies
upon the active control of single molecules in order to stochastically
switch molecular labels between active states (molecules fluoresce
in response to excitation light) and dark states (molecules do not
fluoresce). Once the concentration of active molecules has been lowered
such that each molecule is resolvable by a diffraction-limited microscope,
the location of each molecule is estimated via image fitting or other
statistical methods.[14,15] Then, another stochastic collection
of molecules can be activated and their positions measured. This cycle
continues until the structure of interest has been sampled with sufficient
density, and finally, a reconstructed image of the specimen can be
computed with resolution beyond the diffraction limit. A related optical
measurement technique, two-dimensional (2D) and three-dimensional
(3D) single particle tracking, images an individual emitter or scatterer
and follows its position with nanometric precision as a function of
time, providing scientists with unprecedented ability to directly
observe molecular behaviors and interactions in living cells.[7,16] These methods can use fluorescent objects such as a single molecule,
a nitrogen-vacancy (NV) center in diamond, or a semiconductor nanocrystal
(e.g., quantum dots and rods), or can rely on scattering from a nanoscale
metal particle such as a gold nanorod, etc. For simplicity, we refer
to these single emitters or scatterers as molecules below.The
statistical precision with which a molecule’s position
can be estimated from an image is termed the localization precision
and scales primarily as the inverse square root of the number of photons N detected from the molecule.[17−19] It is commonly measured
by computing the standard deviation σ of repeated measurements x of the location of a molecule, given in one dimension bywhere x̅ is the mean
of the measurements. For the brightest of emitters, localization precisions
of 1 nm are possible.[20−22] However, a key performance metric that is often overlooked
in single-molecule imaging is that of localization accuracy, defined as the difference between the mean x̅ of repeated localizations and the molecule’s true position.
In particular, accurate estimation of a molecule’s position
from its image requires an accurate model of image formation by the
microscope. Many super-resolution analysis methods assume that the
image of a molecule generated by an optical system, called its point
spread function (PSF), is equivalent to that of a point source and
thus circularly symmetric, and images are often fit to an Airy disk
or 2D Gaussian function.[23] However, the
emission pattern of an orientationally fixed single molecule resembles
that of an oscillating electric dipole in the far field.[24,25] Thus, the PSF of a rotationally fixed or moderately constrained
single molecule is asymmetric, and the center of mass of its image
shifts position as a function of molecular orientation and microscope
defocus, leading to ∼100 nm localization errors when a microscope
is misfocused by 200 nm.[23,26−28] Such errors affect both imaging applications designed to elucidate
a structure below the diffraction limit, as well as tracking of spatially
separated emitters in space and time. These errors also affect other
orientationally constrained single-particle labels that emit or scatter
light in a dipole-like radiation pattern, such as gold nanorods,[29,30] NV centers in diamond,[31,32] and semiconductor quantum
rods.[33,34]Fortunately, several solutions to
these systematic errors exist
for single-molecule localization microscopy. The first is to simply
ensure that the fluorescent probe is rotationally mobile, such that
its transition dipole moment explores a large range of angles during
a camera’s acquisition time; thus, the molecule effectively
looks like an isotropic point source. For example, if the molecule’s
transition dipole moment explores a cone of half angle α >
60°
in rotational space, the localization error is bounded to ≤10
nm.[28] A second solution is to incorporate
molecular dipole emission effects into a sophisticated model that
is used for single-molecule image analysis; some of these methods
specifically engineer the microscope PSF for additional orientation
information. The molecular PSF is no longer assumed to be circularly
symmetric, and the image fitting algorithm can now extract accurate
molecular positions even if the image’s center of mass is shifted
away from the molecule’s true position. Such approaches have
been proposed and demonstrated for both 2D[18,35,36] and 3D[37,38] single-molecule
imaging.In this Letter, we propose a third solution to the
dipole shift
problem, namely, the use of an azimuthal polarization filter or azimuthal
polarizer. This all-optical approach for correcting dipole mislocalization
errors is compatible with any widefield epifluorescence microscope
capable of imaging single molecules; can work with any nanoscale dipole
emitter or scatterer regardless of wavelength, label attachment chemistry,
sample geometry, or label rotational mobility; and does not require
modeling of the exact imaging system in order to localize single emitters
with nanoscale accuracy. We demonstrate, via vectorial diffraction
equations and simulations, how the azimuthal polarizer enforces symmetry
in the back focal and image planes of the microscope, thereby leading
to images whose centers of mass always reflect the true position of
the molecules in the sample of interest. We also characterize how
reduced collection efficiency due to polarization filtering affects
localization precision. Finally, we show that this scheme is robust
against aberrations arising from refractive index mismatch of the
sample, an aberration common while imaging biological specimens.Simulation
of image formation of a single dipole emitter by a high
numerical aperture microscope. (A) Schematic of the optical imaging
system (focal lengths not to scale), where OL denotes the objective
lens with focal length fobj, BFP denotes
the back focal plane or pupil plane, and TL denotes the tube lens
with focal length ftube. The green line
illustrates how rays are refracted in the microscope, and red vectors
show how electric field polarization vectors propagate through the
imaging system. With fobj ≪ ftube, the electric field emerging from the tube
lens has a negligible z component. (B) Coordinate
system for expressing the orientation of a molecule’s transition
dipole moment in terms of a unit vector μ̂ or a pair of polar and azimuthal angles {Θ,Φ}. (C) Schematic
of the azimuthally polarized electric field in back focal plane of
the imaging system. The proposed azimuthal polarization filter has
a spatially varying output polarization axis; it perfectly transmits
electric fields parallel to the illustrated vectors and attenuates
perpendicular ones. (D) A conventional clear-aperture microscope and
(E) a microscope with an azimuthal polarization filter project different
intensity distributions in their back focal planes from a molecular
dipole emitter ({Θ = 45°,Φ = 0°} depicted here)
embedded in matched media. In-focus intensity images of a molecular
dipole emitter for (F) a clear-aperture microscope and (G) a microscope
with the azimuthal polarizer. Azimuthal polarization filtering enforces
a C2 symmetry in both the pupil and the
image plane intensity distributions but reduces the light transmission
through the microscope; the intensities of these images are plotted
relative to those of the clear-aperture microscope. Scale bar = 200
nm.A fully vectorial electromagnetic
model for the image formation
of a single-molecule dipole by a high numerical aperture microscope
(Figure 1A) gives key insight into the source
of localization errors and how to mitigate them. In particular, it
allows us to predict how an oscillating electric dipole’s angular
emission spectrum is projected by the microscope into its back focal
plane, also referred to as the pupil or Fourier plane. The specific
patterns of illumination within this plane give rise to localization
errors in super-resolution microscopy, and the symmetry of the polarized
electric fields here too provides an elegant solution for eliminating
these errors.
Figure 1
Simulation
of image formation of a single dipole emitter by a high
numerical aperture microscope. (A) Schematic of the optical imaging
system (focal lengths not to scale), where OL denotes the objective
lens with focal length fobj, BFP denotes
the back focal plane or pupil plane, and TL denotes the tube lens
with focal length ftube. The green line
illustrates how rays are refracted in the microscope, and red vectors
show how electric field polarization vectors propagate through the
imaging system. With fobj ≪ ftube, the electric field emerging from the tube
lens has a negligible z component. (B) Coordinate
system for expressing the orientation of a molecule’s transition
dipole moment in terms of a unit vector μ̂ or a pair of polar and azimuthal angles {Θ,Φ}. (C) Schematic
of the azimuthally polarized electric field in back focal plane of
the imaging system. The proposed azimuthal polarization filter has
a spatially varying output polarization axis; it perfectly transmits
electric fields parallel to the illustrated vectors and attenuates
perpendicular ones. (D) A conventional clear-aperture microscope and
(E) a microscope with an azimuthal polarization filter project different
intensity distributions in their back focal planes from a molecular
dipole emitter ({Θ = 45°,Φ = 0°} depicted here)
embedded in matched media. In-focus intensity images of a molecular
dipole emitter for (F) a clear-aperture microscope and (G) a microscope
with the azimuthal polarizer. Azimuthal polarization filtering enforces
a C2 symmetry in both the pupil and the
image plane intensity distributions but reduces the light transmission
through the microscope; the intensities of these images are plotted
relative to those of the clear-aperture microscope. Scale bar = 200
nm.
To consider the emission pattern of a single fluorescent
molecule,
first the orientation of the molecule’s transition dipole moment μ̂ can be parametrized in terms of its projection
onto Cartesian unit vectors or in terms of a polar angle Θ and
an azimuthal angle Φ in spherical coordinates (Figure 1B). The two descriptions are related byIn
the far field, that is for distances away from the dipole such that ||
≫ λ, the intensity distribution resembles
a torus given bywhere λ is the emission wavelength and
η is the angle between the vectors r and μ̂. The objective lens of a microscope collects
the radiation from the dipole and collimates it to project its angular
emission spectrum onto the back focal plane. We specify the lens to
have a focal length fobj and a numerical
aperture NA = n1 sin (θmax), where n1 is the refractive index of
the immersion medium (n1 = 1.518 for an
oil-immersion lens) and θmax is the maximum collection
angle for the objective lens. Rays with inclination θ ≤
θmax relative to the optical axis, collected from
a dipole located at the focal point of the lens, are refracted such
that they are parallel to the optical axis. The S-polarized component
with respect to the face of the objective lens remains unchanged,
but the P-polarized component is rotated so that it is perpendicular
to the optical axis (Figure 1A). The field
in the back focal plane is thus succinctly written as[39,40]where ρ is the normalized
position in the back focal plane (with |ρ| = 1
corresponding to a distance of fobj from
the optical axis) and ρ̂ and ϕ̂ are the radial and azimuthal unit vectors, respectively, in cylindrical
coordinates. Therefore, the radius of the circle that illuminates
the back focal plane is given by ρmax = sin(θmax) = NA/n1. Transforming eq 4 into Green’s tensor notation,[25] we obtainandwhere the ρϕz superscript explicitly denotes that the rows of the Green’s
tensor Gbfpρϕ and electric field Ebfpρϕ are expressed in terms of the cylindrical unit vectors ρ̂, ϕ̂, and ẑ; n0 is the refractive index at the back
focal plane (typically air, n0 = 1); we
have assumed that the molecule resides in the ideal immersion media
for the objective lens; the wavevector is given by k = 2π/λ; z is the distance between the
molecule and the focal plane of the objective lens; and A is the amplitude of the molecule’s dipole moment. Note that
these expressions are mathematically equivalent to the more conventional
form of the Green’s tensor in Cartesian unit vectors, given
by[41]and eq 6 with
the cylindrical superscript ρϕz replaced
by the Cartesian superscript xyz. Conversion between
the two unit vector bases can be accomplished via the multiplication
of a spatially varying rotation matrix R(ϕ), such
thatandThe electric field in the back focal plane is then projected into
the microscope image plane by the tube lens with focal length ftube. Since high numerical aperture objectives
typically have small focal lengths such that fobj ≪ ftube, light rays
emerging from the tube lens travel mostly parallel to the optical
axis, and we can use scalar diffraction theory to model light propagation
through the tube lens. Thus, the electric field distribution in the
image plane is the scaled Fourier transform of the electric field
in the back focal plane, written aswhere (ρ′,ϕ′)
are
cylindrical coordinates in the image plane and C is
a complex constant. When displaying intensity images of the back focal
plane or the image plane, we plot the field intensity ∥Ebfp∥2 or ∥Eimg∥2.Single-molecule dipoles that are
oriented perpendicular to the
optical axis (Θ = 90°) exhibit negligible localization
error.[23,26,27] This property
arises from the fact that the intensity distribution in the back focal
plane exhibits C2 symmetry that is preserved
in the molecule’s image when the objective lens is defocused.
When the molecule is tilted out of the sample plane (0° <
Θ < 90°), the intensity distribution in the back focal
plane no longer exhibits this symmetry (Figure 1D), leading to asymmetry in the molecule’s image and localization
errors when the objective is defocused and symmetric estimators are
used. Examining eq 5, we observe the special
property that the z component of any single-molecule
dipole does not couple into the azimuthally polarized electric field
in the back focal plane (note the zero in row 2, column 3). In other
words, the azimuthally polarized light arises from the in-plane projection
of the dipole moment (μx̂ + μŷ), transforming
the effective orientation of the single molecule to be perpendicular
to the optical axis (Θ = 90°). Thus, if an azimuthal polarization
filter (a spatially varying polarizer whose polarization axis at each
position is oriented along ϕ̂, Figure 1C) is inserted into the optical detection path at
the back focal plane, the resulting back focal plane intensity (Figure 1E) and image (Figure 1G)
regain their C2 symmetry regardless of
defocus. If the centroid of this image is used for measuring the transverse
(xy) position of a single emitter, this measurement
will not contain any localization error resulting from the dipole
orientation effect. This filtering scheme is compatible with any optical
excitation geometry since it only needs to be inserted into the fluorescence
detection optics at the pupil plane, and it provides the proper filtering
for all molecules in the sample volume simultaneously.Simulated 3D PSFs and
localization errors of rotationally fixed
single dipole emitters. PSFs of a molecule oriented at {Θ =
45°,Φ = 0°} produced by (A) a microscope with a clear
back focal plane and (B) a microscope with an azimuthal polarization
filter. Cross sections of the 3D PSF in the xz (left)
and xy (right) planes show the apparent shift of
the dipole’s position as a function of defocus in the conventional
microscope (solid green line in A), while the apparent lateral position
of the molecule in the azimuthally polarized microscope remains fixed
for all z positions (dashed green line in B). The
intensity of the azimuthally polarized images is plotted relative
to that of the clear-aperture microscope images. The apparent lateral
position of an oriented molecule (black: Θ = 15°, red:
Θ = 30°, green: Θ = 45°, blue: Θ = 60°,
magenta: Θ = 75°, and gold: Θ = 90°) is computed
as a function of defocus z for (C) the conventional
microscope and (D) the azimuthally polarized microscope by fitting
images to a 2D elliptical Gaussian function. The localization error
Δr can be as large as ±100 nm for a defocus
of 200 nm for conventional imaging. The imaging system with an azimuthal
polarizer exhibits no localization error for any molecular orientation
and any amount of objective lens defocus. Scale/axes arrows = 200
nm.Imaging simulations using the
above equations were implemented
by a custom MATLAB (MathWorks, Natick, MA) script (see Supporting Information for details). We model
a single-molecule dipole emitting at λ = 600 nm embedded in
a medium matching the refractive index of immersion oil (n1 = 1.518). This light is collected by a high numerical
aperture objective lens (NA = 1.4). The 3D PSFs for a dipole oriented
at {Θ = 45°,Φ = 0°} are computed for a conventional
clear-aperture microscope (Figure 2A) and a
microscope with an azimuthal polarizer (Figure 2B). (The azimuthal PSFs for other molecular orientations are provided
in Supporting Figure S1.) The apparent
shift in the PSF’s center is clearly visible for the normal
microscope, while the azimuthally polarized PSF remains centered regardless
of objective defocus. Nonlinear least-squares fitting of xy cross sections of these PSFs to a 2D elliptical Gaussian function
quantifies the localization error Δr for both
imaging systems (Figure 2); similar localization
errors would occur for any other estimator ignoring the exact dipole
radiation pattern. The localization error for unpolarized imaging
increases for both increasing defocus (z) and increasing
molecular tilt away from the sample plane (smaller Θ, Figure 2C); for Θ = 60°, Δr = 13 nm at z = 100 nm, and Δr = 28 nm at z = 200 nm, while for Θ = 45°,
Δr = 46 nm at z = 200 nm.
Remarkably, the imaging system with an azimuthal polarization filter
exhibits zero localization error for any molecular orientation and
any misfocus amount (Figure 2D). Furthermore,
the azimuthal filter restores localization accuracy when any dipole-like
emitter’s position is measured.
Figure 2
Simulated 3D PSFs and
localization errors of rotationally fixed
single dipole emitters. PSFs of a molecule oriented at {Θ =
45°,Φ = 0°} produced by (A) a microscope with a clear
back focal plane and (B) a microscope with an azimuthal polarization
filter. Cross sections of the 3D PSF in the xz (left)
and xy (right) planes show the apparent shift of
the dipole’s position as a function of defocus in the conventional
microscope (solid green line in A), while the apparent lateral position
of the molecule in the azimuthally polarized microscope remains fixed
for all z positions (dashed green line in B). The
intensity of the azimuthally polarized images is plotted relative
to that of the clear-aperture microscope images. The apparent lateral
position of an oriented molecule (black: Θ = 15°, red:
Θ = 30°, green: Θ = 45°, blue: Θ = 60°,
magenta: Θ = 75°, and gold: Θ = 90°) is computed
as a function of defocus z for (C) the conventional
microscope and (D) the azimuthally polarized microscope by fitting
images to a 2D elliptical Gaussian function. The localization error
Δr can be as large as ±100 nm for a defocus
of 200 nm for conventional imaging. The imaging system with an azimuthal
polarizer exhibits no localization error for any molecular orientation
and any amount of objective lens defocus. Scale/axes arrows = 200
nm.
Effect of light collection
efficiency on localization precision.
(A) The clear aperture microscope is able to capture more light from
a molecule oriented in the sample plane (Θ = 90°) than
a dipole oriented along the optical axis (Θ = 0°) (solid
line). With azimuthal polarization filtering, the light collection
efficiency is reduced to 65% for in-plane dipoles (dashed line). Collection
efficiencies are plotted relative to the total light captured by the
clear-aperture microscope for Θ = 90°. (B) The limit of
2D localization precision, as computed by the Cramér–Rao
lower bound (CRLB), as a function of dipole orientation for the clear
aperture microscope (solid lines) and a microscope with an azimuthal
polarizer (dashed lines) along the direction parallel to the in-plane
transition dipole moment (μx̂ + μŷ, black lines) and the perpendicular direction (red lines). The CRLB
for molecules whose polar orientation angle is less than 15°
(<3.4% of all possible molecular orientations) is omitted for readability.
The limit of localization precision is also shown as a function of
lens defocus z for an emitter oriented at (C) Θ
= 45° and (D) Θ = 90°. All calculations assume 1000
photons are captured in the clear aperture microscope images for all
molecular orientations and defocus positions, while the azimuthally
polarized images have correspondingly fewer numbers of detected photons
due to a reduced light collection efficiency (A).Azimuthal polarization filtering naturally leads to reduced
light
collection efficiency for the imaging system. This collection efficiency
is a function of molecular orientation (Figure 3A), since an anisotropic emission pattern will be collected differently
by the imaging system for different orientations of a dipole emitter.
For Θ = 90°, the azimuthally polarized imaging system collects
65% of the light captured by the conventional microscope. For an inclined
molecule (Θ = 45°), the clear-aperture microscope collection
efficiency drops to 81%, while with an azimuthal polarizer, the light
collection efficiency is reduced to 33% compared to the unpolarized
microscope with Θ = 90°. Since the azimuthally polarized
light in the back focal plane only contains contributions from the
in-plane component of the dipole, the collection efficiency for the
azimuthally polarized microscope drops to zero for molecules oriented
along the optical axis (Θ = 0°). While these optical losses
could be regarded as a disadvantage, the lost photons are precisely
those that lead to unwanted localization errors, so the loss in photons
is a small price to pay for greatly improved accuracy.
Figure 3
Effect of light collection
efficiency on localization precision.
(A) The clear aperture microscope is able to capture more light from
a molecule oriented in the sample plane (Θ = 90°) than
a dipole oriented along the optical axis (Θ = 0°) (solid
line). With azimuthal polarization filtering, the light collection
efficiency is reduced to 65% for in-plane dipoles (dashed line). Collection
efficiencies are plotted relative to the total light captured by the
clear-aperture microscope for Θ = 90°. (B) The limit of
2D localization precision, as computed by the Cramér–Rao
lower bound (CRLB), as a function of dipole orientation for the clear
aperture microscope (solid lines) and a microscope with an azimuthal
polarizer (dashed lines) along the direction parallel to the in-plane
transition dipole moment (μx̂ + μŷ, black lines) and the perpendicular direction (red lines). The CRLB
for molecules whose polar orientation angle is less than 15°
(<3.4% of all possible molecular orientations) is omitted for readability.
The limit of localization precision is also shown as a function of
lens defocus z for an emitter oriented at (C) Θ
= 45° and (D) Θ = 90°. All calculations assume 1000
photons are captured in the clear aperture microscope images for all
molecular orientations and defocus positions, while the azimuthally
polarized images have correspondingly fewer numbers of detected photons
due to a reduced light collection efficiency (A).
The effect
of this reduced collection efficiency on localization
precision can be quantified by calculating the Fisher information
matrix , used previously to characterize
the limit of localization precision obtainable via super-resolution
microscopy.[42−44] Fisher information provides a convenient, quantitative
metric for comparing the performance of estimation algorithms in the
presence of statistical noise regardless of their implementation details.
The inverse of the Fisher information matrix, called the Cramér-Rao
lower bound (CRLB), provides a lower limit for the variance var (ξ̂)
of any unbiased estimator ξ̂; that isSince
we are interested in the limit of 2D
localization precision σCRLB, we calculate the square
root of the inverse of the Fisher information matrix to compare the
localization performance of conventional single-molecule microscopy
against azimuthally polarized microscopy. The equations that express
Fisher information as a function of PSF shape, number of detected
photons, camera pixelation, and fluorescence background noise are
described in the Supporting Information.The limit of localization precision for the conventional
and the
azimuthally polarized microscopes are plotted in Figure 3B as a function of molecular polar orientation Θ for
an in-focus molecule, with 1000 photons detected by the conventional
microscope. (Camera pixelation and fluorescence background noise are
not included here. However, they are incorporated into our computations
of the limit of localization precision in Supporting
Figure S2.) Instead of computing σCRLB along
the Cartesian axes, we plot it along axes relative to the molecule’s
orientation: the direction parallel to the in-plane projection of
the dipole μx̂ + μŷ and
the perpendicular direction, thereby exploiting the symmetry of the
image. Due to the reduced number of photons in the azimuthally polarized
image, its limit of localization precision is 4.6 nm versus 2.6 nm
for the unpolarized image along the direction parallel to the dipole
for an in-plane molecule. However, since the azimuthal image is narrower
than the conventional image along the perpendicular direction, its
limit of localization precision compares better against the conventional
image: 2.8 nm versus 2.1 nm. As the molecule becomes more inclined
relative to the sample plane (smaller Θ), the limit of localization
precision for both imaging methods worsens gradually due to reduced
collection efficiency. Therefore, one can expect a localization precision
degradation of 50–75% along the parallel direction and <30%
along the perpendicular direction when using an azimuthal polarizer
across a range of molecular orientations that are pumped reasonably
efficiently by widefield excitation (Θ ≥ 30°). Nevertheless,
this loss in localization precision is more than offset by the substantial
improvement in accuracy of azimuthal imaging.We also compare
the limit of localization precision for the two
methods as a function of defocus (z) for two dipole
orientations, Θ = 45° and Θ = 90°. We observe
that both imaging methods have worse localization precision as defocus
increases as expected from gradual blurring of the molecular PSFs.
For an in-focus molecule inclined at Θ = 45° (Figure 3C), the azimuthally polarized image has modestly
worse localization precision (5.8 nm along the parallel and 3.5 nm
along the perpendicular directions) compared to the conventional image
(3.6 nm along the parallel and 3.8 nm along the perpendicular). Defocus
makes the localization precision penalty worse along the parallel
direction (7.4 nm for azimuthal vs 4.2 nm for conventional at z = 200 nm), but the localization precision along the perpendicular
direction is nearly identical for the two microscopes for all z. For Θ = 90°, the localization precision curves
for the azimuthally polarized and conventional microscopes are nearly
parallel as a function of defocus. Thus, we can expect the localization
precision of an azimuthally polarized microscope to be 1.3–1.75
times larger than that of a conventional microscope, depending on
the molecular orientation and localization axis. (The precision of
using a least-squares algorithm to fit pixelated azimuthally polarized
images with a 2D elliptical Gaussian function is comparable to the
limit predicted by the CRLB. See Supporting Figures
S2 and S3 for details.) Again, this decrease in localization
precision is tolerable for most experimental applications because
of the dramatic improvement in localization accuracy.Effect of refractive
index mismatch on images and localization
error of single-molecule dipole emitters at various depths. Images
of a fixed single molecule (Θ = 45°) immersed in water
(n2 = 1.333) at depths of (A) 200 nm,
(B) 400 nm, (C) 600 nm, and (D) 800 nm from the coverglass–water
interface are shown for the clear aperture microscope with an oil-immersion
objective lens (n1 = 1.518); the objective
lens position is 200 nm above the position of best focus. Images of
emitters at corresponding depths are shown for the azimuthally polarized
microscope in E–H. The azimuthally polarized images are plotted
in intensity units relative to those of the clear-aperture images.
Scale bars = 200 nm. (I) Fitting the images to a 2D elliptical Gaussian
function yields measurements of the localization error in the clear-aperture
microscope as a function of emitter depth (black = 200 nm, red = 400
nm, green = 600 nm, and blue = 800 nm) and objective lens defocus z. The localization error worsens for increasing defocus z and decreasing emitter depth. (J) The microscope with
an azimuthal polarizer exhibits no localization error for a large
range of depths and defocus amounts, demonstrating that it is robust
to index mismatch aberrations common during biological imaging.A key difficulty encountered while
calibrating and removing dipole
position error in localization microscopy is the optical aberrations
that can perturb the single-molecule images. Since detailed modeling
of the dipole emission is needed to predict the exact molecular PSFs
projected by an imaging system, aberrations like astigmatism, spherical
aberration, and coma too need to be modeled in order for these methods[18,35−38] to remain accurate. Here, we demonstrate that azimuthal polarization
filtering is robust against the spherical aberration associated with
imaging dipole-like emitters in mismatched media (e.g., focusing an
oil-immersion objective into water) frequently encountered when imaging
biological specimens (see the Supporting Information for formulas that modify Gbfpρϕ for the
presence of index mismatch).Images of a molecular dipole emitter
(Θ = 45°) immersed
in water at several depths (zd = {200,
400, 600, 800} nm) above the coverglass are shown for the conventional
(Figure 4A–D) and azimuthally polarized
(Figure 4E–H) microscopes when the objective
is defocused z = 200 nm above the position of best
focus. Note the increasing circular symmetry apparent in both microscopes
as the emitter moves further from the coverglass; this circular character
is a result of increasing spherical aberration. Remarkably, spherical
aberration slightly increases the accuracy of the conventional microscope
with increasing depth (Δr = 58 nm for a depth
of 200 nm compared to Δr = 41 nm for an emitter
depth of 800 nm). However, these errors are still orders of magnitude
greater than those of the microscope with the azimuthal polarizer.
Since spherical aberration is itself a circularly symmetric aberration
when modeled in the back focal plane of the microscope, adding spherical
aberration does not change the C2 symmetry
of the azimuthally polarized single-molecule images. The center of
the bright central spot in the azimuthally polarized dipole images
remains an accurate indicator of the true lateral position of the
molecule, regardless of how far the molecule is embedded in mismatched
media or the amount of defocus of the objective lens.
Figure 4
Effect of refractive
index mismatch on images and localization
error of single-molecule dipole emitters at various depths. Images
of a fixed single molecule (Θ = 45°) immersed in water
(n2 = 1.333) at depths of (A) 200 nm,
(B) 400 nm, (C) 600 nm, and (D) 800 nm from the coverglass–water
interface are shown for the clear aperture microscope with an oil-immersion
objective lens (n1 = 1.518); the objective
lens position is 200 nm above the position of best focus. Images of
emitters at corresponding depths are shown for the azimuthally polarized
microscope in E–H. The azimuthally polarized images are plotted
in intensity units relative to those of the clear-aperture images.
Scale bars = 200 nm. (I) Fitting the images to a 2D elliptical Gaussian
function yields measurements of the localization error in the clear-aperture
microscope as a function of emitter depth (black = 200 nm, red = 400
nm, green = 600 nm, and blue = 800 nm) and objective lens defocus z. The localization error worsens for increasing defocus z and decreasing emitter depth. (J) The microscope with
an azimuthal polarizer exhibits no localization error for a large
range of depths and defocus amounts, demonstrating that it is robust
to index mismatch aberrations common during biological imaging.
In summary,
we propose an azimuthal polarization filter that counteracts
the localization error arising from the asymmetry inherent in the
emission pattern of a molecular dipole. In particular, we show that
the filtered image only contains light from the in-plane projection
of the molecule’s transition dipole moment, ensuring that no
localization error is incurred as a function of misfocus. As a consequence
of rejecting the radially polarized light emitted by the dipole, this
filter reduces the collection efficiency of the microscope, thereby
worsening its localization precision by 30–75% depending upon
molecular orientation, a small price to pay for the dramatic increase
in accuracy. In principle, light split into radial and azimuthal polarization
channels (as opposed to linear polarization channels common today)
in a microscope can be used to measure molecular orientation and position
simultaneously within a sample; this strategy recaptures the light
thrown away by the azimuthal polarization filter while providing additional
physical information about the dipole-like emitters. We reserve this
demonstration as the subject of a future study.An important
advantage of the proposed azimuthal polarization filter
is its all-optical nature of canceling localization error and, thus,
its simplicity of implementation. This filter can be inserted into
the back focal plane of a conventional epifluorescence microscope
with minimal modifications to its detection optics. A detailed model
of the imaging system is not necessary in order to implement an accurate
fluorophore localization algorithm; in fact, a simple 2D elliptical
Gaussian model of the molecular PSF is sufficient for subnanometer
localization accuracy. Furthermore, this localization accuracy is
robust against the spherical aberration associated with imaging molecules
embedded in mismatched media relative to microscope coverglass, again
using a simple elliptical Gaussian model for the microscope PSF. While
such polarization optics are not currently commercially available,
we believe that certain geometrical transformation strategies[45,46] or metal grating structures[47] could be
adapted to implement an azimuthal polarization filter for super-resolution
fluorescence microscopy and single-particle tracking. In the future
with these optics readily available, the dipole emission effect will
no longer be a source of systematic errors in single-molecule localization
microscopy.
Authors: Gabriele S Kaminski Schierle; Sebastian van de Linde; Miklos Erdelyi; Elin K Esbjörner; Teresa Klein; Eric Rees; Carlos W Bertoncini; Christopher M Dobson; Markus Sauer; Clemens F Kaminski Journal: J Am Chem Soc Date: 2011-08-03 Impact factor: 15.419