Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA.

Yuji Nakaguchi1, Fujio Araki, Takeshi Ono, Yuki Tomiyama, Masato Maruyama, Nozomu Nagasue, Yoshinobu Shimohigashi, Yudai Kai.   

Abstract

In this study, we evaluated the dosimetric performance of the three-dimensional (3D) dose verification system, COMPASS version 3 (IBA Dosimetry, GmbH, Germany). The COMPASS has the function of a dedicated beam modeling and dose calculation. It is able to reconstruct 3D dose distributions on patient CT images, using the incident fluence from a linear accelerator measured with the MatriXX 2D array (IBA Dosimetry). The dose profiles measured with various multi-leaf collimator (MLC) test patterns for the COMPASS were checked by comparison with those of EDR2 (Eastman Kodak, Rochester, NY) films and Monte Carlo (MC) simulations. The COMPASS was also used for dose verification in clinical intensity-modulated radiation therapy (IMRT) plans for head and neck cases. The dose distributions were compared with those measured by 3DVH (Sun Nuclear, Melbourne, FL) and MC. In addition, the quality assurance (QA) times among the COMPASS, 3DVH, and EDR2 were compared. For MLC test patterns, the COMPASS dose profiles agreed within 3 % with those of EDR2 films and MC simulations. The physical resolution of the COMPASS detectors was lower than that of film, but the dose resolution for MLC patterns was comparable to that of film. In clinical plans, the dose-volume-histograms were equal for all systems. The average QA times of the COMPASS, 3DVH, and EDR2 film were 40.1, 59.4, and 121.4 min, respectively. The COMPASS system provides fast and reliable 3D dose verification for clinical IMRT QA. The COMPASS QA process does not require phantom plans. Therefore, it allows a simple QA workflow.

Entities:  

Mesh:

Year:  2014        PMID: 25261343     DOI: 10.1007/s12194-014-0294-x

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  17 in total

1.  Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimator.

Authors:  X R Zhu; P A Jursinic; D F Grimm; F Lopez; J J Rownd; M T Gillin
Journal:  Med Phys       Date:  2002-08       Impact factor: 4.071

2.  Dose verification of IMRT by use of a COMPASS transmission detector.

Authors:  Yuji Nakaguchi; Fujio Araki; Masato Maruyama; Shunji Saiga
Journal:  Radiol Phys Technol       Date:  2011-10-26

3.  Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA.

Authors:  Arthur J Olch
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

4.  Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine.

Authors:  J Herzen; M Todorovic; F Cremers; V Platz; D Albers; A Bartels; R Schmidt
Journal:  Phys Med Biol       Date:  2007-01-30       Impact factor: 3.609

5.  Patient-specific 3D pretreatment and potential 3D online dose verification of Monte Carlo-calculated IMRT prostate treatment plans.

Authors:  Ramesh Boggula; Lennart Jahnke; Hansjoerg Wertz; Frank Lohr; Frederik Wenz
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-11-17       Impact factor: 7.038

6.  3D evaluation of 3DVH program using BANG3 polymer gel dosimeter.

Authors:  Yoichi Watanabe; Yuji Nakaguchi
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

7.  Calibration of a novel four-dimensional diode array.

Authors:  Guanghua Yan; Bo Lu; Jakub Kozelka; Chihray Liu; Jonathan G Li
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

8.  VMAT QA: measurement-guided 4D dose reconstruction on a patient.

Authors:  Benjamin E Nelms; Daniel Opp; Joshua Robinson; Theresa K Wolf; Geoffrey Zhang; Eduardo Moros; Vladimir Feygelman
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

9.  Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT.

Authors:  Erik W Korevaar; David J L Wauben; Peter C van der Hulst; Johannes A Langendijk; Aart A Van't Veld
Journal:  Radiother Oncol       Date:  2011-09-29       Impact factor: 6.280

10.  Evaluation of six TPS algorithms in computing entrance and exit doses.

Authors:  Yun I Tan; Mohamed Metwaly; Martin Glegg; Shaun Baggarley; Alex Elliott
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

View more
  2 in total

1.  Evaluation of Dose Distribution in Intensity Modulated Radiosurgery for Lung Cancer under Condition of Respiratory Motion.

Authors:  Mee Sun Yoon; Jae-Uk Jeong; Taek-Keun Nam; Sung-Ja Ahn; Woong-Ki Chung; Ju-Young Song
Journal:  PLoS One       Date:  2016-09-20       Impact factor: 3.240

2.  Dosimetric evaluation of the compass program for patient dose analysis in IMRT delivery quality assurance.

Authors:  Ju-Young Song; Sung-Ja Ahn
Journal:  PLoS One       Date:  2018-12-20       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.