Patient-specific 3D pretreatment and potential 3D online dose verification of Monte Carlo-calculated IMRT prostate treatment plans.

Ramesh Boggula1, Lennart Jahnke, Hansjoerg Wertz, Frank Lohr, Frederik Wenz.   

Abstract

PURPOSE: Fast and reliable comprehensive quality assurance tools are required to validate the safety and accuracy of complex intensity-modulated radiotherapy (IMRT) plans for prostate treatment. In this study, we evaluated the performance of the COMPASS system for both off-line and potential online procedures for the verification of IMRT treatment plans. METHODS AND MATERIALS: COMPASS has a dedicated beam model and dose engine, it can reconstruct three-dimensional dose distributions on the patient anatomy based on measured fluences using either the MatriXX two-dimensional (2D) array (offline) or a 2D transmission detector (T2D) (online). For benchmarking the COMPASS dose calculation, various dose-volume indices were compared against Monte Carlo-calculated dose distributions for five prostate patient treatment plans. Gamma index evaluation and absolute point dose measurements were also performed in an inhomogeneous pelvis phantom using extended dose range films and ion chamber for five additional treatment plans.
RESULTS: MatriXX-based dose reconstruction showed excellent agreement with the ion chamber (<0.5%, except for one treatment plan, which showed 1.5%), film (∼100% pixels passing gamma criteria 3%/3 mm) and mean dose-volume indices (<2%). The T2D based dose reconstruction showed good agreement as well with ion chamber (<2%), film (∼99% pixels passing gamma criteria 3%/3 mm), and mean dose-volume indices (<5.5%).
CONCLUSION: The COMPASS system qualifies for routine prostate IMRT pretreatment verification with the MatriXX detector and has the potential for on-line verification of treatment delivery using T2D.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21093168     DOI: 10.1016/j.ijrobp.2010.09.010

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  23 in total

1.  Dose verification of IMRT by use of a COMPASS transmission detector.

Authors:  Yuji Nakaguchi; Fujio Araki; Masato Maruyama; Shunji Saiga
Journal:  Radiol Phys Technol       Date:  2011-10-26

2.  [Image-guided radiation therapy. Paradigm change in radiation therapy].

Authors:  F Wenz; C Belka; M Reiser; S O Schönberg
Journal:  Radiologe       Date:  2012-03       Impact factor: 0.635

3.  Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA.

Authors:  Yuji Nakaguchi; Fujio Araki; Takeshi Ono; Yuki Tomiyama; Masato Maruyama; Nozomu Nagasue; Yoshinobu Shimohigashi; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2014-09-27

4.  Commissioning and validation of fluence-based 3D VMAT dose reconstruction system using new transmission detector.

Authors:  Yuji Nakaguchi; Takeshi Oono; Masato Maruyama; Yoshinobu Shimohigashi; Yudai Kai; Yuya Nakamura
Journal:  Radiol Phys Technol       Date:  2018-03-12

5.  Error Detectability of Isodose Volumes as ROIs in Prostate Intensity-modulated RT QA.

Authors:  Ryuta Nakahara; Masayuki Fujiwara; Haruyuki Takaki; Masao Tanooka; Kentaro Ishii; Ryu Kawamorita; Koichiro Yamakado
Journal:  In Vivo       Date:  2022 Jul-Aug       Impact factor: 2.406

6.  Validation of Dolphin dosimetry in three dimensional patient-specific quality assurance programme.

Authors:  Niyas Puzhakkal; Abdullah Kallikuzhiyil Kochunny; Dinesh Makuny; Arun Krishnan M P; Ranjith C Poyil; Vysakh Raveendran
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-12

7.  Patient specification quality assurance for glioblastoma multiforme brain tumors treated with intensity modulated radiation therapy.

Authors:  H I Al-Mohammed
Journal:  Int J Med Sci       Date:  2011-08-02       Impact factor: 3.738

8.  Comparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma.

Authors:  Hailei Lin; Shaomin Huang; Xiaowu Deng; Jinhan Zhu; Lixin Chen
Journal:  Radiat Oncol       Date:  2014-03-07       Impact factor: 3.481

9.  Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans.

Authors:  Murugesan Kathirvel; Shanmuga Subramanian; Alessandro Clivio; Gandhi Arun; Antonella Fogliata; Giorgia Nicolini; Vellaiyan Subramani; Shanmugam Thirumalai Swamy; Eugenio Vanetti; Luca Cozzi
Journal:  Radiat Oncol       Date:  2013-06-11       Impact factor: 3.481

10.  Evaluation of the efficiency and effectiveness of independent dose calculation followed by machine log file analysis against conventional measurement based IMRT QA.

Authors:  Baozhou Sun; Dharanipathy Rangaraj; Sunita Boddu; Murty Goddu; Deshan Yang; Geethpriya Palaniswaamy; Sridhar Yaddanapudi; Omar Wooten; Sasa Mutic
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.