MSCT versus CBCT: evaluation of high-resolution acquisition modes for dento-maxillary and skull-base imaging.

Jean-Philippe Dillenseger1, Jean-François Matern, Catherine-Isabelle Gros, Fabien Bornert, Christian Goetz, Jean-Marie Le Minor, André Constantinesco, Philippe Choquet.   

Abstract

OBJECTIVES: Our aim was to conduct a quantitative and qualitative evaluation of high-resolution skull-bone imaging for dentistry and otolaryngology using different architectures of recent X-ray computed tomography systems.
MATERIAL AND METHODS: Three multi-slice computed tomography (MSCT) systems and one Cone-beam computed tomography (CBCT) system were used in this study. All apparatuses were tested with installed acquisition modes and proprietary reconstruction software enabling high-resolution bone imaging. Quantitative analyses were performed with small fields of view with the preclinical vmCT phantom, which permits to measure spatial resolution, geometrical accuracy, linearity and homogeneity. Ten operators performed visual qualitative analyses on the vmCT phantom images, and on dry human skull images.
RESULTS: Quantitative analysis showed no significant differences between protocols in terms of linearity and geometric accuracy. All MSCT systems present a better homogeneity than the CBCT. Both quantitative and visual analyses demonstrate that CBCT acquisitions are not better than the collimated helical MSCT mode.
CONCLUSION: Our results demonstrate that current high-resolution MSCT protocols could exceed the performance of a previous generation CBCT system for spatial resolution and image homogeneity. KEY POINTS: • Quantitative evaluation is a prerequisite for comparison of imaging equipment. • Bone imaging quality could be objectively assessed with a phantom and dry skull. • The current MSCT shows better image quality than a dental CBCT system. • CBCT remains a work-in-progress technology.

Entities:  

Mesh:

Year:  2014        PMID: 25249315     DOI: 10.1007/s00330-014-3439-8

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  17 in total

Review 1.  System performance of multislice spiral computed tomography.

Authors:  T O Fuchs; M Kachelriess; W A Kalender
Journal:  IEEE Eng Med Biol Mag       Date:  2000 Sep-Oct

2.  Comparison of spatial and contrast resolution for cone-beam computed tomography scanners.

Authors:  Ruben Pauwels; Jilke Beinsberger; Harry Stamatakis; Kostas Tsiklakis; Adrian Walker; Hilde Bosmans; Ria Bogaerts; Reinhilde Jacobs; Keith Horner
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2012-05-15

3.  A quality assurance phantom for the performance evaluation of volumetric micro-CT systems.

Authors:  Louise Y Du; Joseph Umoh; Hristo N Nikolov; Steven I Pollmann; Ting-Yim Lee; David W Holdsworth
Journal:  Phys Med Biol       Date:  2007-11-15       Impact factor: 3.609

Review 4.  Conebeam CT of the head and neck, part 1: physical principles.

Authors:  A C Miracle; S K Mukherji
Journal:  AJNR Am J Neuroradiol       Date:  2009-05-13       Impact factor: 3.825

5.  Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications.

Authors:  M Loubele; R Bogaerts; E Van Dijck; R Pauwels; S Vanheusden; P Suetens; G Marchal; G Sanderink; R Jacobs
Journal:  Eur J Radiol       Date:  2008-07-18       Impact factor: 3.528

6.  A comparative study for spatial resolution and subjective image characteristics of a multi-slice CT and a cone-beam CT for dental use.

Authors:  Hiroshi Watanabe; Eiichi Honda; Akemi Tetsumura; Tohru Kurabayashi
Journal:  Eur J Radiol       Date:  2009-10-09       Impact factor: 3.528

Review 7.  Physics of cardiac imaging with multiple-row detector CT.

Authors:  Mahadevappa Mahesh; Dianna D Cody
Journal:  Radiographics       Date:  2007 Sep-Oct       Impact factor: 5.333

8.  A high-resolution XRII-based quantitative volume CT scanner.

Authors:  D W Holdsworth; M Drangova; A Fenster
Journal:  Med Phys       Date:  1993 Mar-Apr       Impact factor: 4.071

9.  A practical method to measure the MTF of CT scanners.

Authors:  R T Droege; R L Morin
Journal:  Med Phys       Date:  1982 Sep-Oct       Impact factor: 4.071

10.  Low-dose and scatter-free cone-beam CT imaging using a stationary beam blocker in a single scan: phantom studies.

Authors:  Xue Dong; Michael Petrongolo; Tianye Niu; Lei Zhu
Journal:  Comput Math Methods Med       Date:  2013-11-20       Impact factor: 2.238

View more
  9 in total

Review 1.  Quality assurance phantoms for cone beam computed tomography: a systematic literature review.

Authors:  Marcus V L de Oliveira; Ann Wenzel; Paulo S F Campos; Rubens Spin-Neto
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

2.  Image quality evaluation of small FOV and large FOV CBCT devices for oral and maxillofacial radiology.

Authors:  Jean-Philippe Dillenseger; Catherine-Isabelle Gros; Amira Sayeh; Johary Rasamimanana; Fabrice Lawniczak; Jean-Marie Leminor; Jean-François Matern; André Constantinesco; Fabien Bornert; Philippe Choquet
Journal:  Dentomaxillofac Radiol       Date:  2016-09-21       Impact factor: 2.419

3.  Effectiveness of CT assists for intraarterial chemotherapy: therapeutic outcome of chemoradiation for advanced head and neck cancer extending across the anatomical midline.

Authors:  Daisuke Yunaiyama; Kazuhiro Saito; Tomokazu Funatsu; Hidetsugu Nakayama; Akira Shimizu; Hiroyuki Ito; Mamoru Suzuki; Soichi Akata; Koichi Tokuuye
Journal:  Jpn J Radiol       Date:  2014-11-08       Impact factor: 2.374

4.  Quantitative performance characterization of image quality and radiation dose for a CS 9300 dental cone beam computed tomography machine.

Authors:  Elham Abouei; Sierra Lee; Nancy L Ford
Journal:  J Med Imaging (Bellingham)       Date:  2015-11-18

5.  Factors affecting modulation transfer function measurements in cone-beam computed tomographic images.

Authors:  Jin-Woo Choi
Journal:  Imaging Sci Dent       Date:  2019-06-24

6.  Intrapetrous Anastomosis between the Internal Jugular Vein and the Superior Petrosal Sinus: Cone Beam Computed Tomography Incidental Finding.

Authors:  Ilson Sepúlveda; Thomas Schmidt; Enrique Platín
Journal:  J Clin Imaging Sci       Date:  2015-08-31

7.  Effectiveness of 2D radiographs in detecting CBCT-based incidental findings in orthodontic patients.

Authors:  Jin-Young Choi; Song Hee Oh; Seong-Hun Kim; Hyo-Won Ahn; Yoon-Goo Kang; Yong-Suk Choi; Yoon-Ah Kook; Gerald Nelson
Journal:  Sci Rep       Date:  2021-04-29       Impact factor: 4.379

8.  A dose-neutral image quality comparison of different CBCT and CT systems using paranasal sinus imaging protocols and phantoms.

Authors:  Ari-Petteri Ronkainen; Ali Al-Gburi; Timo Liimatainen; Hanna Matikka
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-01-27       Impact factor: 3.236

Review 9.  The clinical application of high-resolution peripheral computed tomography (HR-pQCT) in adults: state of the art and future directions.

Authors:  J P van den Bergh; P Szulc; A M Cheung; M Bouxsein; K Engelke; R Chapurlat
Journal:  Osteoporos Int       Date:  2021-05-22       Impact factor: 4.507

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.