Chenchen Bing1, Michelle Ladouceur-Wodzak1, Clinton R Wanner2, John M Shelton2, James A Richardson3, Rajiv Chopra1. 1. Department of Radiology, University of Texas Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX 75390-9061, USA. 2. Department of Internal Medicine, University of Texas Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX 75390-9061, USA. 3. Department of Pathology, University of Texas Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX 75390-9061, USA ; Department of Molecular Biology, University of Texas Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX 75390-9061, USA.
Abstract
OBJECTIVE: The blood-brain barrier (BBB) protects the brain by preventing the entry of large molecules; this poses a major obstacle for the delivery of drugs to the brain. A novel technique using focused ultrasound (FUS) energy combined with microbubble contrast agents has been widely used for non-invasive trans-cranial BBB opening. Traditionally, FUS research is conducted with magnetic resonance imaging (MRI) guidance, which is expensive and poses physical limitations due to the magnetic field. A system that could allow researchers to test brain therapies without MR intervention could facilitate and accelerate translational research. METHODS: In this study, we present a novel FUS system that uses a custom-built FUS generator mounted on a motorized stereotaxic apparatus with embedded brain atlas to locally open the BBB in rodents. The system was initially characterized using a tissue-mimicking phantom. Rodent studies were also performed to evaluate whether non-invasive, localized BBB opening could be achieved using brain atlas-based targeting. Brains were exposed to pulsed focused ultrasound energy at 1.06 MHz in rats and 3.23 MHz in mice, with the focal pressure estimated to be 0.5-0.6 MPa through the skull. BBB opening was confirmed in gross tissue sections by the presence of Evans blue leakage in the exposed region of the brain and by histological assessment. RESULTS: The targeting accuracy of the stereotaxic system was better than 0.5 mm in the tissue-mimicking phantom. Reproducible localized BBB opening was verified with Evans blue dye leakage in 32/33 rats and had a targeting accuracy of ±0.3 mm. The use of higher frequency exposures in mice enabled a similar precision of localized BBB opening as was observed with the low frequency in the rat model. CONCLUSIONS: With this dedicated small-animal motorized stereotaxic-FUS system, we achieved accurate targeting of focused ultrasound exposures in the brain for non-invasive opening of the BBB. This system can be used as an alternative to MR-guided FUS and offers researchers the ability to perform efficient studies (30 min per experiment including preparation) at a reduced cost in a conventional laboratory environment.
OBJECTIVE: The blood-brain barrier (BBB) protects the brain by preventing the entry of large molecules; this poses a major obstacle for the delivery of drugs to the brain. A novel technique using focused ultrasound (FUS) energy combined with microbubble contrast agents has been widely used for non-invasive trans-cranial BBB opening. Traditionally, FUS research is conducted with magnetic resonance imaging (MRI) guidance, which is expensive and poses physical limitations due to the magnetic field. A system that could allow researchers to test brain therapies without MR intervention could facilitate and accelerate translational research. METHODS: In this study, we present a novel FUS system that uses a custom-built FUS generator mounted on a motorized stereotaxic apparatus with embedded brain atlas to locally open the BBB in rodents. The system was initially characterized using a tissue-mimicking phantom. Rodent studies were also performed to evaluate whether non-invasive, localized BBB opening could be achieved using brain atlas-based targeting. Brains were exposed to pulsed focused ultrasound energy at 1.06 MHz in rats and 3.23 MHz in mice, with the focal pressure estimated to be 0.5-0.6 MPa through the skull. BBB opening was confirmed in gross tissue sections by the presence of Evans blue leakage in the exposed region of the brain and by histological assessment. RESULTS: The targeting accuracy of the stereotaxic system was better than 0.5 mm in the tissue-mimicking phantom. Reproducible localized BBB opening was verified with Evans blue dye leakage in 32/33 rats and had a targeting accuracy of ±0.3 mm. The use of higher frequency exposures in mice enabled a similar precision of localized BBB opening as was observed with the low frequency in the rat model. CONCLUSIONS: With this dedicated small-animal motorized stereotaxic-FUS system, we achieved accurate targeting of focused ultrasound exposures in the brain for non-invasive opening of the BBB. This system can be used as an alternative to MR-guided FUS and offers researchers the ability to perform efficient studies (30 min per experiment including preparation) at a reduced cost in a conventional laboratory environment.
Entities:
Keywords:
BBB opening; Drug delivery; Focused ultrasound; Stereotaxic system
For most therapeutic agents aimed at treating central nervous system (CNS) diseases and
disorders, the blood-brain barrier (BBB) is a primary physiological barrier limiting
drug delivery into the brain parenchyma. The BBB is a separation present along all
capillaries in the CNS that controls the molecular transport and diffusion across these
blood vessels. The barrier is formed by layers of cells that are coupled by tight
junctions [1,2]. Only small-molecule drugs with high lipid solubility and a low molecular
mass under 400–500 Da can cross the BBB in pharmacologically significant
amounts, hence excluding most current therapeutic and imaged agents from being used in
the brain [3,4].Non-invasive localized opening of the BBB has been demonstrated using focused ultrasound
(FUS) exposures combined with circulating intravascular microbubble ultrasound contrast
agents. The exact physical mechanisms governing the interactions between the
microbubbles and endothelial cells are not known, but it is likely that when stimulated
by ultrasound energy, oscillation of microbubbles produces mechanical effects induced by
radiation force and/or shear stress on the blood vessel walls, temporarily opening the
BBB without tissue damage [5]–[7]. This combination of FUS and intravascular microbubbles offers a unique
method for remotely actuating mechanical energy at the site of small vessels throughout
the brain. Meanwhile, this opening occurs at lower acoustic power levels than was
previously used without intravenous microbubbles, which makes this method substantially
easier to apply through the intact skull [8]. As early as 2001, Hynynen et al. demonstrated that focused ultrasound
combined with gas bubbles can open the BBB transiently in rabbits [9]. In 2002, Mesiwala et al. confirmed that high-intensity focused ultrasound is
capable of a selective and non-destructive disruption of the BBB in rats [10]. Subsequent studies confirmed the feasibility of non-invasive localized BBB
opening in rodents as models of human disease [11,1,13]–[16]. However, the small size of rodent brains makes FUS experiments challenging,
necessitating the development of dedicated small-animal exposure systems [17].In the majority of studies to date, confirmation and visualization of BBB opening has
been achieved using magnetic resonance imaging (MRI). T1-weighted imaging is often used
to confirm successful delivery of gadolinium-based MR contrast agents across the BBB,
and T2-weighted imaging can evaluate the presence of tissue damage [18,19]. The major drawbacks of MR-guided FUS are the need for MRI-compatible
systems, lack of availability, limited throughput of experiments, and expense. MR-guided
focused ultrasound systems are not usually accessible to neuroscientists or other
researchers outside of the imaging research field. A method for performing trans-cranial
BBB opening without MR guidance would be desirable in order to lower the barrier to
entry in this field and to achieve broader penetration of FUS technology in
neuroscience.Previous studies have incorporated systems for BBB opening outside the MRI environment.
One method for targeting specific regions in the brain is the use of stereotaxy in
combination with an anatomical atlas. This method is widely used in neuroscience to the
injection of agents or implantation of electrodes in specific regions of the rodent
brain. Liu et al.'s group developed a pinhole-assisted mechanical scanning device using
a stereotaxic apparatus for BBB opening [13], and Konofagou et al. applied a stereotaxic frame for localized BBB opening
in rodents [20]. Another method for imaging BBB opening is multi-photon fluorescence
microscopy. The two-photon microscopy allows in vivo visualization of the
cerebral vasculature and neurons at the subcellular level [21]. Besides, cerebrovascular dynamics and kinetics of dye leakage after FUS
sonication can also be imaging with multi-photon microscopy [22]. However, in order to enable targeting of the ultrasound focus to a specific
functional area in the brain, MR images are still required in these systems.In this study, we present a compact stereotaxic-FUS system to perform trans-cranial
localized BBB opening in rodent models using FUS energy and a stereotaxic system with a
built-in rodent brain atlas. The goal of this system is to provide a tool for
neuroscientists to achieve non-invasive targeted BBB opening in rodents for behavioral
and functional research. Using the rodent brain atlas, researchers can easily target a
specific functional area of the brain. In this paper, we introduce our stereotaxic-FUS
system, followed by initial accuracy characterization in tissue-mimicking phantoms and
rodents. Successful BBB opening was confirmed through gross imaging, histopathology, and
MR imaging. By all forms of measure, repeatable non-invasive localized BBB opening by
our rodent stereotaxic-FUS system was verified.
Methods and materials
Focused ultrasound system and stereotaxic apparatus
Ultrasound was transmitted into the brain using a focused transducer with a 25-mm
diameter and a 20-mm radius of curvature. The fundamental frequency of the transducer
was 1.06 MHz, and the third harmonic frequency was 3.23 MHz, as measured
with an impedance analyzer (Via Bravo, AEA Technology Inc., Carlsbad, CA, USA). The
dimensions of the ultrasound focus and the pressure output of the transducer were
characterized using an acoustic hydrophone tank. A needle hydrophone with a 0.2-mm
active area (HGL-0200, Onda Corporation, Sunnyvale, CA, USA) was scanned in three
dimensions to measure the spatial pressure distribution of the ultrasound beam at the
focus for each transmission frequency.The transducers were connected to a custom-built compact driving system comprised of
an arbitrary waveform generator (LMS271D, Vanuix Corp., Haverhill, MA, USA), an RF
amplifier (NP Technologies, Newbury Park, CA, USA), high-power low-pass filters, and
a dual-frequency matching circuit that enabled efficient transmission of power to the
transducer at each operating frequency. This self-contained driving system was
controlled from a laptop computer via USB. One of the main goals was to produce a
portable system for focused ultrasound that could be easily transported for
experiments.A standard stereotaxic apparatus (51730 M, Stoelting Co., Wood Dale, IL, USA)
was used to achieve targeting of ultrasound energy within the brain with an
attachment capable of registering the rat brain atlas (Neurostar, Tubingen, Germany).
These systems are normally used to achieve precise insertion of electrodes or
injection of materials directly into target regions of the brain. The method of
registering the brain atlas to the rodent involves locating the bregma and lambda on
the rodent skull sutures with a metal pointer. Once the x, y, and
z coordinates of these two locations are identified, the brain atlas
(which is referenced to these points) is immediately registered to the animal.
Tilting correction was performed automatically based on the current animal to take
the difference in brain position and size into consideration.
Tissue-mimicking phantom
In order to test the spatial accuracy of ultrasound delivery using the
stereotaxic-FUS system, a series of exposures were performed in a custom-made
tissue-mimicking phantom. The phantom was comprised of a hydrogel with dissolved
bovineserum albumin (BSA), similar to previous recipes [23], which could be used to capture the location of the ultrasound focus
through coagulation of the BSA. Briefly, gellan gum (1% w/v,
Gelrite, CP Kelco, Atlanta, GA, USA) and salt (0.23% w/v)
were added to and dissolved in deionized, degassed water
(90% v/v). Metamucil (0.18% w/v,
P&G, Cincinnati, OH, USA) was added to the mixture to create ultrasound
scattering. BSA (25% v/v, CF-0020, Boval, Cleburne, TX, USA)
was dissolved in the gel to provide ultrasound absorption. BSA also provides an
optical marker of the ultrasound focus since the protein undergoes coagulation when
heated above 70°C and creates a visible region of opacity within the otherwise
transparent gel. The heating of the gel was achieved using a continuous delivery of
ultrasound at the high third harmonic frequency (3.23 MHz). The recipe for the
tissue-mimicking phantom is shown in Table 1. The speed of
sound and ultrasound attenuation coefficient of the phantom material were
characterized using a hydrophone tank and were measured to be
1,518 ± 2 m/s and 0.3 ± 0.1 dB/cm (at
1 MHz), respectively.
Table 1
Spatial accuracy characterization with the phantom
Axis
Desired (mm)
Measured (mm)
Error (%)
Desired (mm)
Measured (mm)
Error (%)
Target A-B
Target A-C
x
0.5
0.5 ± 0.01
±2
1
1 ± 0.01
±1
y
0.5
0.5 ± 0.01
±2
1
1 ± 0.01
±1
z
0.5
0.5 ± 0.01
±2
1
0.9 ± 0.01
-2 ~ 0
Target A-D
Target A-E
x
1.5
1.5 ± 0.01
±0.6
2
2 ± 0.01
±0.5
y
1.5
1.5 ± 0.01
±0.6
2
2 ± 0.01
±0.5
z
1.5
1.4 ± 0.01
-1.3 ~ -0.6
2
1.9 ± 0.01
-1 ~ -0.5
Spatial accuracy characterization with the phantomThe spatial accuracy of ultrasound delivery was evaluated by exposing targets with
different spacing (0.5, 1, 1.5, and 2 mm) in all three directions (x,
y, z). The tissue-mimicking phantom was placed in the position
where the rodent head would be located, and the ultrasound transducer was coupled to
it using degassed water. After the exposures, the distance between the regions of
coagulation in the gel were measured and compared with the desired separation. Tests
were completed in two planes (x-y and x-z). For
each plane, one reference target (A) and four testing targets (B, C, D, E) were
selected to apply the sonication.As an important next step, ultrasound exposures were delivered to the
tissue-mimicking phantom in a three-dimensional (3D) pattern based on the rat brain
atlas in order to cover a spatial region equivalent to the right motor cortex of the
brain. The geometric fidelity of this ultrasound exposure was evaluated in comparison
to the brain atlas.
Animal experiment
Female 200–270-g Sprague Dawley rats (n = 56) and
female 20–25-g Swiss Webster mice (n = 14) were used in
this study. All procedures were approved by UT Southwestern Institutional Animal
Care and Use Committee and followed guidelines set forth by the Guide. Animals
were anesthetized with a mixture of 2%–3.5% isoflurane and
1–2 l/min of 100% oxygen. A 24G I.V. catheter was placed in the lateral
tail vein. A pulse oximeter was attached to the animal's paw to monitor heart rate
and oxygen saturation, and a rectal temperature probe (attached to a homeothermic
control blanket) was used to record and maintain core body temperature
(PhysioSuite, Kent Scientific Corp., Torrington, CT, USA). Hair over the cranial
surface of the skull was removed using an animal trimmer and depilatory cream
(VEET sensitive formula, Reckitt Benckiser, Parsippany, NJ, USA). After
preparation, the animal was transferred to the stereotaxic apparatus and
stabilized using ear bars and bite bar. A custom-built nose cone was placed over
the animal's nose to deliver inhalant anesthetic.In order to perform brain atlas
registration, a skin incision over the skull was performed to identify cranial
landmarks bregma and lambda. Ultrasound gel was applied on the skull, and a
custom-built water reservoir filled with degassed water was lowered over the skull
for ultrasound gel coupling. The ultrasound probe was mounted on the stereotaxic
system followed by target selection from the brain atlas. The ultrasound probe was
then automatically moved to the targeted area to generate a focus at the specific
location. This setup procedure is described in Figure 1.
Figure 1
Overview of the procedure used to perform trans-cranial ultrasound
exposures using a standard stereotaxic positioning system. Step 1: A
skin incision is made to expose the skull surface, and a metal pointer is
used to locate the bregma and lambda sutures. Step 2: An ultrasound coupling
chamber is placed over the skull with a thin layer of ultrasound gel between
the chamber and skull. The chamber is filled with degassed water. Step 3:
The metal pointer is replaced with a focused ultrasound transducer whose
focus is now aligned with bregma. The rat brain atlas is used to determine
the appropriate distance to move the ultrasound focus to a targeted region
in the brain.
Overview of the procedure used to perform trans-cranial ultrasound
exposures using a standard stereotaxic positioning system. Step 1: A
skin incision is made to expose the skull surface, and a metal pointer is
used to locate the bregma and lambda sutures. Step 2: An ultrasound coupling
chamber is placed over the skull with a thin layer of ultrasound gel between
the chamber and skull. The chamber is filled with degassed water. Step 3:
The metal pointer is replaced with a focused ultrasound transducer whose
focus is now aligned with bregma. The rat brain atlas is used to determine
the appropriate distance to move the ultrasound focus to a targeted region
in the brain.Evans blue (2%; 3 ml/kg rats, 4 ml/kg mice) was injected via the tail
vein catheter and allowed to circulate for a minimum of 1–3 min. A
bolus injection of microbubbles (30 μl/kg rats, 60 μl/kg mice,
Optison, GE Healthcare, Milwaukee, WI, USA) was delivered via the tail vein
catheter, with simultaneous start of ultrasound sonication. When multiple
sonications were performed (in the same animal), 5 min of wait time was
applied to allow microbubbles to clear from circulation.All brain targets were selected in the right hemisphere (center of caudoputamen)
while the left hemisphere was kept as a control. Twenty-three rats were utilized
to optimize FUS parameters (frequency, microbubble dosage, focal pressure, etc.)
on BBB opening results. After optimization, 33 rats were utilized to verify the
reliability of the system and to characterize spatial accuracy. Among those rats,
5 were utilized for histology. Nineteen mice were utilized to optimize ultrasound
frequencies specific to mouse cranial space and parietal bone thickness (7 for low
frequency, 1.06 MHz, and 12 for high frequency, 3.23 MHz).Animals were sacrificed and perfused 5–10 min after sonication with
sterile saline and 10% buffered formalin, followed by immediate harvesting of the
brain. Brain samples were then placed in 10% buffered formalin to allow for
24–48 h of fixation time. Brains were sliced using a rat or mouse brain
matrix (World Precision Instrument, Sarasota, FL, USA) to confirm localization of
Evans blue leakage following BBB opening. Photographic record of Evans blue
leakage was made, and a subset of subgross slices was worked up by histology.Subgross coronal brain-matrix slices from perfusion-fixed animals were paraffin
processed, embedded, and sectioned according to standard procedures [24,25]. Serial paraffin sections were prepared at the epicenter of
FUS-mediated hemorrhage, seen by dark-field microscopy [26]. Resulting sections were stained by hematoxylin and eosin (H&E) and
terminal deoxynucleotidyltransferase-mediated UTP end labeling (TUNEL).
H&E-stained sections were analyzed for histopathologic hallmarks of brain
injury (hemorrhage, edema, nuclear condensation, inflammatory cell infiltrate,
rarefaction). Histologic analysis for possible ultrasound damage to neuronal and
glial cell populations in the FUS target area was performed via TUNEL; nuclei of
apoptotic and necrotic cells were labeled with fluorescein according to methods of
the first report [27]. Sections subjected to TUNEL were counterstained with propidium
iodide.Single-frequency pulsed exposures (10-ms burst duration) were transmitted into the
brain with a repetition frequency of 1 Hz and a total exposure duration of
120 s. In rats, the intracranial peak negative pressure at the focal point
under low frequency (1.06 MHz) was adjusted to be between 0.5 and
0.6 MPa, based on the hydrophone calibration of the transducer and published
measurements of the insertion loss through rat parietal bone [28]. Similar opening is expected with lower pressure. The threshold for the
BBB opening was estimated to be inversely proportional to the square root of
frequency [29]. Prior studies have investigated extensively the effects of different
sonication parameters on the threshold of BBB opening [18]: frequency [30], pulse width [31,32], pulse repetition time [31], total sonication duration [33], microbubble size [34], and contrast agent dose [35,36]. The ultrasound exposure parameters selected for this study have been
shown to achieve consistent opening of the BBB in rodent models [37]. For the mouse exposures, a 3.23-MHz frequency was used instead of
1.06 MHz in order to achieve a relatively localized ablation within the mouse
brain. Skull insertion loss under high frequency is still unknown, so it was
assumed to be approximately 50% based on the skull thickness. The focused
ultrasound energy was adjusted to achieve a focal pressure of
0.5–0.6 MPa accordingly.Evans blue dye has been used frequently as an indicator of BBB opening. Evans blue
binds to albumin which cannot cross the BBB; therefore, neural tissue remains
unstained after intravenous administration of the dye [38]. In these experiments, BBB opening was confirmed on gross tissue
sections based on the leakage of Evans blue dye in the targeted area of the
brain.As a secondary validation, two rats that received ultrasound exposures to open the
BBB in a targeted region were administered an intravenous injection of gadodiamide
(0.02 mmol/kg, Omniscan, GE Healthcare, Milwaukee, WI, USA) and were imaged
in a 9.4 T animal imaging and spectroscopy system (Avance, Varian Medical
Systems, Palo Alto, CA, USA). Contrast-enhanced T1-weighted spin-echo MR images
were obtained to visualize BBB opening through diffusion of the contrast agent
into the brain parenchyma (field of view 128 × 128, TR/TE
151/10 ms, slice thickness 1 mm, gap 0 mm, echo number 1).The spatial targeting accuracy of the system was evaluated through localization of
the region of Evans blue staining on gross sections. Two specific targets were
manually selected using the brain atlas, and the spacing between targets was
measured on the gross sections. Atlas spacing was compared with measured spacing
in three axes: anterior-posterior (AP), medial-lateral (ML), and superior-inferior
(SI). Reference in ML and SI axes is the midline of the brain and top surface of
the brain, respectively. For characterization of the AP axis, a reference point
was selected as coordinates 0.00, 3.32, and 5.55 mm (AP, ML, and SI) relative
to bregma since the cranial landmarks disappear after removal of the skull. Five
rats were utilized to characterize targeting accuracy along each axis.
Results
Focused ultrasound transducer characterization
The spatial pressure distribution produced by the focused ultrasound transducer,
driven at the fundamental and the third harmonic frequency, was characterized in an
acoustic hydrophone tank. The pressure distributions are shown in Figure 2 as a 3D map of the ultrasound beam. The scan was completed in
planes of 6 × 10 mm2 along (x-z,
y-z) and 6 × 6 mm2 transverse
(x-y) to the beam to cover the entire focal area. Two sidelobes
along the x direction were caused by electrode tabs on the transducer. The
ultrasound transducer was excited with a 20-cycle pulse (40 cycles at the third
harmonic frequency) and 100-Hz repetition frequency signal. The pressure at each
location was acquired as an average of 256 pulses. The full width at half maximum
(FWHM) of the beam was measured to be
1.6 × 2 × 10 ± 0.1 mm
(x × y × z) for the
fundamental frequency and
0.4 × 0.6 × 3 ± 0.1 mm
(x × y × z) for the
third harmonic frequency. The beam volume is approximately 0.0336 and
0.00072 cm3, respectively. At higher frequency, the focal area is
about 46 times more compact. The pressure profile along each axis (x,
y, z) is shown in Figure 2. The peak
negative pressure at the focal point ranged from 0.1 to 0.8 MPa at the
fundamental frequency and from 0.05 to 1.2 MPa at the third harmonic
frequency.
Figure 2
The ultrasound beam plots. (A, B) 3D beam plots for the fundamental and
the third harmonic frequency, respectively. At the third harmonic frequency,
the focal area is more compact with higher intensity. (C-E) Intensity
profiles in three directions (x, y, z) for the two
frequencies. The peak amplitude could reach 0.25 V at the fundamental
frequency (red curve) while 0.38 V at the third harmonic
frequency (blue curve).
The ultrasound beam plots. (A, B) 3D beam plots for the fundamental and
the third harmonic frequency, respectively. At the third harmonic frequency,
the focal area is more compact with higher intensity. (C-E) Intensity
profiles in three directions (x, y, z) for the two
frequencies. The peak amplitude could reach 0.25 V at the fundamental
frequency (red curve) while 0.38 V at the third harmonic
frequency (blue curve).
Spatial accuracy characterization—phantom
Figure 3 shows a sample pattern of coagulated volumes
generated in the tissue-mimicking phantom. The desired separation and measured
separation for each target are included in Table 2. In the
x and y axes, the precision of targeting was approximately
0.01 mm for four different separations. However, in the z axis, the
measured spacing was smaller than the desired spacing by up to 2%. This was likely
due to the difference in ultrasound attenuation and sound velocities between the
phantom and water.
Figure 3
Isolated regions of coagulated BSA within the phantom caused by focal
heating with the 3.23-MHz transducer. The test was performed in the
x‒z direction (left) and the
x‒y direction (right) to measure the
separation between individual sonications. The precision of targeting was
approximately 0.01 mm, and the accuracy was less than 2%. The greatest
error occurred in the z direction which is due to the difference in
speed of sound between the coupling water and phantom and the different amounts
of each in the beam path with different focal depths.
Table 2
Spatial accuracy characterization with the rat model
Atlas spacing (mm)
Measured spacing (mm)
Percentage error
Rat A
Rat B
Rat C
Rat D
Rat E
Medial/lateral
3.32
3.5
3.1
3.0
3.1
3.1
±6.02
4.03
4.2
4.4
4.0
4.1
3.9
±4.96
Superior/inferior
5.55
5.1
5.4
5.7
5.7
5.7
±3.78
4.37
4
4.5
4.1
4.5
4.3
±4.44
Anterior/posterior
3
3.2
3.1
2.9
2.9
3.2
±4.67
6
6.1
5.9
5.8
5.8
6.1
±2.33
Isolated regions of coagulated BSA within the phantom caused by focal
heating with the 3.23-MHz transducer. The test was performed in the
x‒z direction (left) and the
x‒y direction (right) to measure the
separation between individual sonications. The precision of targeting was
approximately 0.01 mm, and the accuracy was less than 2%. The greatest
error occurred in the z direction which is due to the difference in
speed of sound between the coupling water and phantom and the different amounts
of each in the beam path with different focal depths.Spatial accuracy characterization with the rat modelFigure 4 shows the results obtained for generating a 3D
ultrasound exposure pattern that overlapped with the rat brain right motor cortex.
The dimension of the motor cortex model was measured to be
8.4 × 2.84 × 5.17 mm
(L × H × W), with
123.34-mm3 volume. The dimension of the phantom coagulation was
measured to be
8 × 3.4 × 4.6 ± 0.1 mm
(L × H × W), with
125.12 (117.315 ~ 133.245)-mm3 volume. The percentage error of
pattern volume was within ±10% (-4.88% ~ 8.03%).
Figure 4
Results obtained for generating a 3D ultrasound exposure pattern. (A) A
3D rendering of the right motor cortex (red) and the brain
(yellow), segmented from the standard rat brain atlas. (B)
Views of the motor cortex from each of the three principal axes. (C) The
corresponding pattern of coagulated BSA in the gel phantom along the same views
and spatial scale. The 3D pattern of coagulation in the gel phantom was
equivalent to the shape of the motor cortex, and the volume was within 10%.
Results obtained for generating a 3D ultrasound exposure pattern. (A) A
3D rendering of the right motor cortex (red) and the brain
(yellow), segmented from the standard rat brain atlas. (B)
Views of the motor cortex from each of the three principal axes. (C) The
corresponding pattern of coagulated BSA in the gel phantom along the same views
and spatial scale. The 3D pattern of coagulation in the gel phantom was
equivalent to the shape of the motor cortex, and the volume was within 10%.
Animal studies
Evidence of successful and confined BBB opening following FUS was observed by both
Evans blue dye leakage and contrast-enhanced MR images in the targeted brain area.
Figure 5(A.1) shows the whole brain and targeted
brain slice for rat, under low-frequency (1.06 MHz) sonication. Evans blue
dye can be observed on the cortical brain surface and in the corresponding
underlying slice, indicating the location of BBB opening. On the other hand, when
low frequency was also applied in mice for comparison, Evans blue dye leakage was
observed through the entire targeted hemisphere, likely due to the thinner
parietal bone. In order to achieve a similar tight focus, high frequency
(3.23 MHz) was applied in the mouse model. Figure 5(A.2) shows the BBB opening results for mouse. In both the rat and
mouse models, the un-targeted hemisphere presents as control. Figure 5(B) renders the T1-weighted MR images acquired as secondary
validation. B.1 refers to a transverse slice while B.2 refers to a coronal slice.
The presence of an increased signal in the brain due to extravasation of the
contrast agent indicates localized BBB opening.
Figure 5
Validation with the animal model. Evans blue dye leakage served as
primary validation. A.1/A.2 shows BBB opening on the
rat/mouse separately. A photo of the entire brain is referred at the corner;
bar is 5 mm. T1‒weighted spin‒echo
multi‒slice MR images (TR/TE, 151/10 ms) were acquired with rat
as secondary validation. B.1 and B.2 show images in
transverse and coronal directions, respectively. Yellow rectangle
highlighted the brain, and contrast enhancement (white arrow)
indicates localized BBB opening.
Validation with the animal model. Evans blue dye leakage served as
primary validation. A.1/A.2 shows BBB opening on the
rat/mouse separately. A photo of the entire brain is referred at the corner;
bar is 5 mm. T1‒weighted spin‒echo
multi‒slice MR images (TR/TE, 151/10 ms) were acquired with rat
as secondary validation. B.1 and B.2 show images in
transverse and coronal directions, respectively. Yellow rectangle
highlighted the brain, and contrast enhancement (white arrow)
indicates localized BBB opening.The system's spatial accuracy relative to the brain atlas was characterized in
three axes: anterior/posterior (AP), medial/lateral (ML), and superior/inferior
(SI). Figure 6 shows a stepwise overview of brain
registry as the calvarium was removed. Photos were taken before and after removing
the calvarium to indicate the location of bregma and the reference point.
Validation results in three axes are included in Table 2. Atlas spacing was acquired from the brain atlas during the targeting
procedure. Measured spacing was acquired from the brain slice after dissection,
relative to references. The location of each exposure was manually selected to be
at the centroid of the focal area. Percentage error was calculated based on
average measured spacing and atlas spacing. Percentage error for targeting is
estimated to be around ±5%.
Figure 6
Overview of the dissection procedure applied to verify the accuracy of
the reference target. The location of the reference target is (0,
3.32, 5.55), measured based on bregma. Brain suture and landmarks are shown
in (A), whole brain with the skull removed is shown in (C),
and (B) serves as an in‒between photo. White dash line
performs as an indicator of bregma.
Overview of the dissection procedure applied to verify the accuracy of
the reference target. The location of the reference target is (0,
3.32, 5.55), measured based on bregma. Brain suture and landmarks are shown
in (A), whole brain with the skull removed is shown in (C),
and (B) serves as an in‒between photo. White dash line
performs as an indicator of bregma.As further validation of the rat model, detailed histologic analysis was performed
on coronal-matrix slices collected immediately following FUS and 72 h
post-FUS exposure. Figure 7 shows ultrasound-induced
BBB opening by routine histopathology. H&Es of acute and 72-h brains both show
evidence of perivascular hemorrhage in the target region. Hemorrhage was limited
to the target area in both lateral-medial and anterior-posterior axes. Mild
perivascular edema and condensation of adjacent neuronal nuclei was sparsely
evident. Low-magnification images illustrate the contralateral hemisphere without
hemorrhage.In the absence of major histopathology on H&E in FUSrat brains,
TUNEL was performed to look for apoptotic and/or necrotic vascular endothelium and
the surrounding brain parenchyma. No cells with nicked DNA were detected in the
brains from either acute or 72-h post-FUSrats (Figure 8). Images of juvenile mouse thymus TUNEL-positive control are included
in the figure for contrast to the autofluorescence of extravascular RBCs present
in FUS brain sections.
Figure 7
H&E-stained sections for acute (A, C) and 72‒h recovery (B, D)
rats. Both rats received the same ultrasound exposures in the same
location in the brain, indicated by an asterisk. Low-magnification
images of the entire brain section (A, B) depict the exposed and control
hemispheres. A cutting artifact is observed in (A) on the right cortical
region and is unrelated to the ultrasound exposure. A × 10
magnified region (inset in A and B) depicts the cellular architecture in the
exposed regions of the brain. Evidence of perivascular hemorrhage in the
target region can be observed in the acute exposure.
Figure 8
Tissue sections stained with propidium iodide ( ) and TUNEL ( ) for acute (A)
and 72-h recovery (B) rats. The sections were taken from the same
regions of the brain as are shown in Figure 7C,D.
The absence of green staining in the acute and recovery rats demonstrates
that no cells with nicked DNA were present in these regions. An image of
juvenile mouse thymus TUNEL-positive control (C) is included at the
same magnification for contrast to the slight green autofluorescence of
extravascular red blood cells present in the brain sections exposed to
ultrasound.
H&E-stained sections for acute (A, C) and 72‒h recovery (B, D)
rats. Both rats received the same ultrasound exposures in the same
location in the brain, indicated by an asterisk. Low-magnification
images of the entire brain section (A, B) depict the exposed and control
hemispheres. A cutting artifact is observed in (A) on the right cortical
region and is unrelated to the ultrasound exposure. A × 10
magnified region (inset in A and B) depicts the cellular architecture in the
exposed regions of the brain. Evidence of perivascular hemorrhage in the
target region can be observed in the acute exposure.Tissue sections stained with propidium iodide ( ) and TUNEL ( ) for acute (A)
and 72-h recovery (B) rats. The sections were taken from the same
regions of the brain as are shown in Figure 7C,D.
The absence of green staining in the acute and recovery rats demonstrates
that no cells with nicked DNA were present in these regions. An image of
juvenile mouse thymus TUNEL-positive control (C) is included at the
same magnification for contrast to the slight green autofluorescence of
extravascular red blood cells present in the brain sections exposed to
ultrasound.
Discussion
This study presented a novel system that interfaces a stereotaxic brain atlas and a FUS
energy transducer to achieve non-invasive localized BBB opening in rodents, without MR
intervention. The spatial accuracy of the system was estimated to be ±2% with the
tissue-mimicking phantom and ±5% with the rodent models. Evans blue dye leakage was
used as a subgross indicator of localized BBB opening, which was subsequently validated
by intact animal MR imaging of gadolinium contrast agent. Applying low frequency
(1.06 MHz) and high frequency (3.23 MHz) on rats and mice, respectively, a
tight three-dimensional focus of ultrasound was achieved. This portable system can be
used for rodents' BBB opening in a traditional laboratory environment.The primary research application we envision for this system is for the targeting of
regions within the rodent brain to facilitate functional neuroscience studies in
rodents. Since the majority of these investigations are conducted in conventional
laboratory environments in close proximity to behavioral testing facilities, a portable
FUS system that does not require MRI targeting and guidance is desirable. Furthermore,
the brain atlas is an important targeting tool for functional regions of the brain in
rodents.To our knowledge, this is the first study to explore the use of high-frequency FUS
exposures in mice for BBB opening. Exposure of the mouse brain at 1.06 MHz resulted
in a very large region of BBB opening in the brain encompassing the entire AP direction
within a hemisphere. This is due to the thinner parietal bone and smaller brain size in
mice. Exposures at 3.23 MHz achieved a localized region of BBB opening similar in
proportion to the brain as was achieved using 1.06 MHz in rats. However, unlike
experiments with the rat model, the skull insertion loss is still unknown at this high
frequency, and the focal pressure inside the mouse brain was estimated for this study
based on prior measurements at a lower frequency in the rat skull [28]. Experimental measurement of the insertion loss through the mouse skull bone
at 3 MHz is necessary to obtain more accurate estimates of the intracranial
pressure amplitudes during these exposures.The system described in this manuscript still suffers from a few limitations. In order
to achieve BBB opening within a larger region of the brain, rapid mechanical [17] or electronic [39] steering of the ultrasound focus is desirable. The stereotaxic system
described in this study is amenable to these technical improvements, and this addition
would enable regional opening of target structures within the brain. The incorporation
of acoustic feedback has been demonstrated to achieve more consistent stable cavitation
during ultrasound exposures for BBB opening [40,41]. Due to the lack of imaging confirmation of BBB opening with the stereotaxic
system, this type of feedback would be an important improvement to the FUS system.
Another limitation of this approach is that an incision is needed to identify bregma and
lambda. Potential strategies to overcome this would be to use the external ear bars as a
reference point or to use a technique like high-frequency ultrasound imaging to
non-invasively locate the appropriate stereotactic reference point on the skull. These
approaches will be the subject of future investigations. Nonetheless, while these
improvements would expand the capabilities of the existing system, the current
configuration is still capable of consistent opening of the BBB in rodents.
Conclusion
A novel system that interfaces a stereotaxic brain atlas and a FUS transducer for
non-invasive targeted BBB opening in rodents is presented in this study. The spatial
targeting accuracy of the system was estimated to be ±2% in tissue-mimicking
phantoms and ±5% in the rat brain. Localized opening of the BBB was achieved at
1.06 MHz in rats and 3.23 MHz in mice. Consistent opening of the BBB was
verified through Evans blue staining in gross brain sections and histopathology in
H&E-stained tissue sections. TUNEL staining confirmed that there was minimal to no
apoptosis nor necrosis in the exposed regions of the brain.
Competing interests
The authors declare that they have no competing interests.
Authors' contributions
CB was the main generator of this manuscript. She was responsible for developing and
characterizing the stereotaxic focused ultrasound system. She also worked as an operator
during the animal experiments for this study. MW was the veterinary technologist and
preclinical coordinator in this study. She led the animal experiments involved in this
project. She also contributed to the writing of the animal study section in this
manuscript. CRW, JMS, and JAR performed and analyzed the histological tissue sections
produced in this study. They also co-wrote the histology section in this manuscript. RC
was the principal investigator of this study. He came up with the scientific concept and
overall design for the study and reviewed the manuscript. He also supported the
scientific work financially from existing grant support. All authors read and approved
the final manuscript.
Authors: Chengzu Long; Leonela Amoasii; Alex A Mireault; John R McAnally; Hui Li; Efrain Sanchez-Ortiz; Samadrita Bhattacharyya; John M Shelton; Rhonda Bassel-Duby; Eric N Olson Journal: Science Date: 2015-12-31 Impact factor: 47.728
Authors: H Petra Kok; Erik N K Cressman; Wim Ceelen; Christopher L Brace; Robert Ivkov; Holger Grüll; Gail Ter Haar; Peter Wust; Johannes Crezee Journal: Int J Hyperthermia Date: 2020 Impact factor: 3.914
Authors: Chenchen Bing; Robert M Staruch; Matti Tillander; Max O Köhler; Charles Mougenot; Mika Ylihautala; Theodore W Laetsch; Rajiv Chopra Journal: Int J Hyperthermia Date: 2016-05-22 Impact factor: 3.914
Authors: Markus A Dahlem; Bernd Schmidt; Ingo Bojak; Sebastian Boie; Frederike Kneer; Nouchine Hadjikhani; Jürgen Kurths Journal: Front Comput Neurosci Date: 2015-03-05 Impact factor: 2.380
Authors: Gerben Schooneveldt; Hana Dobšíček Trefná; Mikael Persson; Theo M de Reijke; Klas Blomgren; H Petra Kok; Hans Crezee Journal: Cancers (Basel) Date: 2019-08-15 Impact factor: 6.639