| Literature DB >> 25207411 |
Vibha Chaswal1, Michael Weldon, Nilendu Gupta, Arnab Chakravarti, Yi Rong.
Abstract
We present commissioning and comprehensive evaluation for ArcCHECK as a QA equipment for volumetric-modulated arc therapy (VMAT), using the 6 MV photon beam with and without the flattening filter, and the SNC patient software (version 6.2). In addition to commissioning involving absolute dose calibration, array calibration, and PMMA density verification, ArcCHECK was evaluated for its response dependency on linac dose rate, instantaneous dose rate, radiation field size, beam angle, and couch insertion. Scatter dose characterization, consistency and symmetry of response, and dosimetry accuracy evaluation for fixed aperture arcs and clinical VMAT patient plans were also investigated. All the evaluation tests were performed with the central plug inserted and the homogeneous PMMA density value. Results of gamma analysis demonstrated an overall agreement between ArcCHECK-measured and TPS-calculated reference doses. The diode based field size dependency was found to be within 0.5% of the reference. The dose rate-based dependency was well within 1% of the TPS reference, and the angular dependency was found to be ± 3% of the reference, as tested for BEV angles, for both beams. Dosimetry of fixed arcs, using both narrow and wide field widths, resulted in clinically acceptable global gamma passing rates on the 3%/3mm level and 10% threshold. Dosimetry of narrow arcs showed an improvement over published literature. The clinical VMAT cases demonstrated high level of dosimetry accuracy in gamma passing rates.Entities:
Mesh:
Year: 2014 PMID: 25207411 PMCID: PMC5875521 DOI: 10.1120/jacmp.v15i4.4832
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Location of six diodes within 5 mm of the central axes of phantom used for analysis of dose‐rate dependency, dose‐per‐pulse response, and field‐size dependence test. Diodes located inside the red rectangle around origin are the analysis diodes; these lie at 5 mm from CAX.
Figure 2Measurement analysis geometry of the angular dependence test. Upper part of the figure depicts the measurement geometry of a field irradiation on a central axial cut of ArcCHECK. Lower part depicts analysis geometry: 2D projections of the radiation field edge and ~ 1 cm inside are shown on the 2D plane containing all the diodes. Average measured dose from 2 “analysis diodes” is used as one data point for angular dependence.
Figure 3Dose‐per‐pulse (instantaneous dose rate) dependence of 6X and 6F beam as measured by the average dose over six central diodes at 5 mm from the isocenter in axial and transverse directions. Dose is normalized to the dose at 100 SAD.
Dose‐per‐pulse (instantaneous dose rate) dependence of central ArcCHECK diodes
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|
|
| ||||||
| 90 | 155.59 | 1.23 | 1.28 | 1.03 | 4.1 | 0.41 |
| 100 | 122.02 | 1.00 | 1.00 | 1.00 | 0.0 | 0.00 |
| 110 | 99.07 | 0.83 | 0.81 | 0.98 |
|
|
| 120 | 81.04 | 0.69 | 0.66 | 0.96 |
|
|
|
| ||||||
| 90 | 152.16 | 1.23 | 1.26 | 1.02 | 2.1 | 0.21 |
| 100 | 121.23 | 1.00 | 1.00 | 1.00 | 0.0 | 0.00 |
| 110 | 97.52 | 0.83 | 0.80 | 0.97 |
|
|
| 120 | 80.49 | 0.69 | 0.66 | 0.96 |
|
|
Average gamma passing rates for different square field sizes varying from to
|
|
|
| ||
|---|---|---|---|---|
|
|
|
|
|
|
|
| ||||
|
| 100 | 89.4 | 99.7 | 81.8 |
|
| 100 | 97.4 | 100 | 94.4 |
|
| 99.9 | 95.0 | 98.9 | 85.7 |
|
| 99.6 | 86.7 | 93.9 | 74.7 |
|
| ||||
|
| 100 | 98.5 | 100 | 90.9 |
|
| 100 | 94.3 | 100 | 90.0 |
|
| 100 | 85.7 | 98.8 | 99.3 |
|
| 95.6 | 76.9 | 90.9 | 62.4 |
Figure 4Angular dependence factors of ArcCHECK diodes normalized to the diode reading with the field size.
Global and local gamma passing rates of the measured arcs (two exposures with phantom rotated on its long axis) as compared with the arcs calculated in the TPS
|
|
|
| ||
|---|---|---|---|---|
|
|
|
|
| |
|
| 99.9% | 99.9% | 98.3% | 98.1% |
|
| 99.8% | 99.8% | 97.6% | 97.5% |
|
| 99.8% | 99.6% | 95.6% | 95.0% |
|
| 99.6% | 99.4% | 94.5% | 94.0% |
| Average | 99.8% | 99.7% | 96.5% | 96.2% |
| Std Dev | 0.13 | 0.22 | 1.76 | 1.96 |
Dosimetry of narrow beam aperture arcs
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
| |
|
| ||||||||
|
| 75.0% | ‐ | 59.2% | ‐ | 89.9% | ‐ | 69.9% | ‐ |
|
| 87.9% | ‐ | 52.5% | ‐ | 96.1% | ‐ | 68.1% | ‐ |
|
| 96.1% | 80.3% | 75.0% | 66.4% | 99.9% | 90.0% | 88.2% | 80.0% |
Dosimetry of full arcs with wide open apertures
|
|
|
|
| |||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
| |
|
| ||||||||
|
| 99.9% | 77.2% | 76.5% | 63.9% | 99.3% | 84.3% | 77.8% | 74.8% |
|
| 100% | 81.7% | 95.5% | 75.5% | 100.0% | 99.9% | 85.1% | 84.0% |
|
| 99.9% | 99.9% | 98.3% | 98.1% | 100.0% | 99.9% | 98.1% | 97.0% |
|
| 99.9% | 99.9% | 94.1% | 93.6% | 91.6% | 88.5% | 67.0% | 63.6% |
|
| 99.8% | 99.6% | 95.6% | 95.0% | 98.6% | 95.9% | 80.7% | 74.0% |
| Average | 99.9% | 95.3% | 95.9% | 90.6% | 99.5% | 98.6% | 88% | 85.5% |
| SD | 0.00 | 0.09 | 0.02 | 0.10 | 0.01 | 0.03 | 0.09 | 0.16 |
Gamma passing rates (global) of delivered VMAT plans versus planned (calculated) doses in the TPS
|
|
| |||
|---|---|---|---|---|
| γ | γ | γ | γ | |
|
|
|
|
|
|
| Brain retreatment (4 Arcs) | 96.1% | 91.0% | 98.3% | 93.3% |
| RTOG0933 (2 Arcs) | 99.9% | 97.4% | 99.9% | 98.3% |
| Scalp (3 Arcs) | 98.5% | 92.0% | 99.4% | 97.2% |
| RPC‐HN (2 Arcs) | 94.7% | 86.3% | 99.0% | 95.7% |
| RPC‐Spine (4 partial Arcs) | 91.1% | 76.6% | 97.9% | 91.6% |
| Average | 96.1% | 88.7% | 98.9% | 95.2% |
| SD | 3.4 | 7.8 | 0.8 | 2.8 |