| Literature DB >> 27455495 |
Paul Kinsella1, Laura Shields, Patrick McCavana, Brendan McClean, Brian Langan.
Abstract
Multileaf collimators (MLCs) need to be characterized accurately in treatment planning systems to facilitate accurate intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT). The aim of this study was to examine the use of MapCHECK 2 and ArcCHECK diode arrays for optimizing MLC parameters in Monaco X-ray voxel Monte Carlo (XVMC) dose calculation algorithm. A series of radiation test beams designed to evaluate MLC model parameters were delivered to MapCHECK 2, ArcCHECK, and EBT3 Gafchromic film for comparison. Initial comparison of the calculated and ArcCHECK-measured dose distributions revealed it was unclear how to change the MLC parameters to gain agreement. This ambiguity arose due to an insufficient sampling of the test field dose distributions and unexpected discrepancies in the open parts of some test fields. Consequently, the XVMC MLC parameters were optimized based on MapCHECK 2 measurements. Gafchromic EBT3 film was used to verify the accuracy of MapCHECK 2 measured dose distributions. It was found that adjustment of the MLC parameters from their default values resulted in improved global gamma analysis pass rates for MapCHECK 2 measurements versus calculated dose. The lowest pass rate of any MLC-modulated test beam improved from 68.5% to 93.5% with 3% and 2 mm gamma criteria. Given the close agreement of the optimized model to both MapCHECK 2 and film, the optimized model was used as a benchmark to highlight the relatively large discrepancies in some of the test field dose distributions found with ArcCHECK. Comparison between the optimized model-calculated dose and ArcCHECK-measured dose resulted in global gamma pass rates which ranged from 70.0%-97.9% for gamma criteria of 3% and 2 mm. The simple square fields yielded high pass rates. The lower gamma pass rates were attributed to the ArcCHECK overestimating the dose in-field for the rectangular test fields whose long axis was parallel to the long axis of the ArcCHECK. Considering ArcCHECK measurement issues and the lower gamma pass rates for the MLC-modulated test beams, it was concluded that MapCHECK 2 was a more suitable detector than ArcCHECK for the optimization process.Entities:
Mesh:
Year: 2016 PMID: 27455495 PMCID: PMC5690063 DOI: 10.1120/jacmp.v17i4.6190
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
TPF MLC parameters in Monaco 5.0
|
|
|
|
|---|---|---|
| Leaf transmission | 0.0120 | 0.007 |
| Leaf groove width (mm) | 1.0 | 1.2 |
| Interleaf leakage | 5.00 | 25 |
| Leaf tip Leakage | 1.10 | 1.10 |
| Leaf offset (mm) | 0 |
|
| MLC corner leakage | 0 | 0.1 |
Manufacturer‐supplied test beams for MLC model evaluation
|
|
| |
|---|---|---|
|
|
|
|
|
|
|
|
| 3ABUT | Three 6 cm wide fields via step‐and‐shoot delivery. The fields were designed to abut and created two junctions. This plan was used to evaluate MLC calibration and the leaf offset parameter. |
|
| 7SEGA | Seven 2 cm wide fields via step‐and‐shoot delivery. The fields were designed to abut and created six junctions. This plan was used to evaluate MLC calibration and the leaf offset parameter. |
|
| FOURL | Four L‐shaped fields via step‐and‐shoot delivery. Each sequential L is smaller producing an overall large L‐shape distribution. The plan was designed to ‘abut’ on each side of the L‐shape. This field was used to examine the leaf position offset, the MLC transmission, and the MLC groove width settings. |
|
| DMLC | A dynamic 10 cm sweep with a 2 cm wide field (defined by both the leaves and backup jaws). The dose distribution was sensitive to changes in the MLC position offset, transmission, and calibration. |
|
| HIMRT | Head and neck step‐and‐shoot IMRT field. The intention of this field was to evaluate the impact the TPF MLC settings in a complex clinical situation. |
|
| HDMLC | Head and neck sliding window IMRT field. The intention of this field was to evaluate the impact the TPF MLC settings in a complex clinical situation. |
|
Figure 1Dose distributions resulting from the T‐shaped test beam: (a) planar greyscale dose distribution; (b) and (d) dose profiles along the dashed line in (a). The profile in (c) corresponds to the highlighted rectangular region of interest in (b). The data labels in (b). 1a.lsDoo apply in (c).
Gamma analysis results for measured vs. calculated test beam doses
|
|
| |||||||
|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
| 100.0 | 100.0 | 100.0 | 100.0 | 99.0 | 98.8 | 99.2 | 99.0 |
|
| 97.0 | 99.8 | 100.0 | 100.0 | 97.7 | 98.3 | 99.1 | 99.1 |
| T‐shape | 55.2 | 82.2 | 60.9 | 84.8 | 35.6 | 31.4 | 38.7 | 33.4 |
| 3ABUT | 89.8 | 98.7 | 97.2 | 99.9 | 70.0 | 58.3 | 86.9 | 80.8 |
| 7SEGA | 63.4 | 93.4 | 86.8 | 97.8 | 62.0 | 43.5 | 80.2 | 70.0 |
| DMLC | 64.5 | 80.3 | 68.5 | 97.1 | 68.5 | 66.8 | 85.8 | 84.7 |
| HIMRT | 98.7 | 99.6 | 99.6 | 100.0 | 94.7 | 93.4 | 97.8 | 97.0 |
| HDMLC | 98.6 | 99.4 | 99.3 | 99.9 | 95.5 | 95.1 | 98.2 | 97.9 |
| FOURL | 86.8 | 91.8 | 89.4 | 93.5 | 84.4 | 87.7 | 92.6 | 95.2 |
Local gamma analysis with 2% dose threshold.
Figure 2Dose distributions resulting from the 3ABUT test beam: (a) planar greyscale dose distribution; (b)–(e): dose profiles along the dashed lines in (a). The data labels in (c) apply to (b), (d), and (e).
Figure 3Dose distributions resulting from the FOURL test beam: (a) planar greyscale dose distribution; (b) dose profiles along the dashed lines in (a). Note: The water phantom calculation with the default and optimized settings overlay and make them indistinguishable.