Colm T Morrison1,2, Kirsty L Symons1, Simon J Woodings1,2, Michael J House2.
Abstract
A volumetric modulated arc therapy (VMAT) approach to total body irradiation (TBI) has recently been introduced at our institution. The planning target volume (PTV) is divided into separate sub-volumes, each being treated with 2 arcs with their own isocentre. Pre-treatment quality assurance of beams is performed on a Sun Nuclear ArcCHECK diode array. Measurement of junction regions between VMAT arcs with separate isocentres has previously been performed with point dose ionization chamber measurements, or with films. Translations of the ArcCHECK with respect to a known distance between the adjacent isocentres of two arcs, which are repeated with the ArcCHECK in an inverted position, allows the recording of a junction dose map. A 3%/3 mm global gamma analysis (10% threshold) pass rate for arc junctions were comparable to their component arcs. Dose maps of junction regions between adjacent arcs with different isocentres can be readily measured on a Sun Nuclear ArcCHECK diode array.Entities:
Keywords: zzm321990QAzzm321990; zzm321990VMAT TBIzzm321990; ArcCHECK; junction dose
Mesh:
Year: 2017 PMID: 29082594 PMCID: PMC5689932 DOI: 10.1002/acm2.12208
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1(a) Coronal view beam arrangement for each sub‐volume. The entire PTV (blue) is covered by eight arcs assigned to four isocentres, with each isocentre having a superior (red) and inferior (yellow) arc. (b) Final isodose colourwash after optimization of all arcs to deliver 12 Gy to the PTV, with sparing to the lungs, liver, and kidneys.
Figure 2ArcCHECK inversion procedure for merging two exposures to accommodate extended field sizes (a), and schematic diagram of exposing two arcs from a single PTV sub‐volume (b).23
Figure 3ArcCHECK setup method for measuring junction between two arcs from adjacent sub‐volumes, i.e., two overlapping arcs from separate isocentres.
Figure 4Single fraction colourwash isodose maps measured (a) and TPS planned (b), for a head sub‐volume (upper), a chest sub‐volume (lower), and the junction between the inferior head arc and superior chest arc (middle).
3%/3 mm absolute dose gamma analysis results for all arcs for the representative patient
| PTV | Beam | % Pass |
|---|---|---|
| Head | Head 1 | 99.4 |
| Head 2 | 99.2 | |
| Head 1 & 2 | 99.7 | |
| Head 2 & Chest 1 | 99.9 | |
| Chest | Chest 1 | 100 |
| Chest 2 | 99.8 | |
| Chest 1 & 2 | 98.7 | |
| Chest 2 & Abdo 1 | 98.3 | |
| Abdo | Abdo 1 | 98.4 |
| Abdo 2 | 99.7 | |
| Abdo 1 & 2 | 97.5 | |
| Abdo 2 & Pelv 1 | 100 | |
| Pelvic | Pelv 1 | 99.8 |
| Pelv 2 | 100 | |
| Pelv 1 & 2 | 100 |
Figure 5Cross‐section of the ArcCHECK phantom with attached scatter accessory demonstrating scatter and secondary radiation originating from within the scatter accessory and contributing to exit diode signals.23