Evaluation of the ArcCHECK QA system for IMRT and VMAT verification.

Guangjun Li1, Yingjie Zhang, Xiaoqin Jiang, Sen Bai, Guang Peng, Kui Wu, Qingfeng Jiang.   

Abstract

The purposes of this study were to perform tests for the ArcCHECK QA system, and to evaluate the suitability of this system for IMRT and VMAT verification. The device was tested for short term reproducibility, dose linearity, dose rate dependence, dose per pulse dependence, field size dependence, out of field dependence and directional dependence. Eight simple plans that each used four beams of different field sizes as well as IMRT and VMAT plans for various organs of 10 patients were measured by ArcCHECK. The phantom data was then compared with ion chamber measurements and planned results. The ArcCHECK diodes performed well for all tests except directional dependence, which varies from a minimum of -4.9% (seen only when the beam is incident on the diode at 180°) to a maximum of 9.1% (approximately at 105°). For simple plan verification, the absolute dose pass rates of γ index (3%/3 mm) were almost identical. They had an average pass rate of 94.6% ± 1.3% when the field size was ≤20 cm in the X direction (right to left direction), but the pass rate fell rapidly when the field size was >20 cm in the X direction. For all patient-specific IMRT and VMAT QA, the pass rates exceeded 95% and 93%, respectively, and high reproducibility of these results has been observed from week to week. The comparative measurements show that the ArcCHECK QA system is completely suitable for clinical IMRT and VMAT verification.
Copyright © 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22583979     DOI: 10.1016/j.ejmp.2012.04.005

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  27 in total

1.  Application of TG-218 action limits to SRS and SBRT pre-treatment patient specific QA.

Authors:  Yuqing Xia; Justus Adamson; Yana Zlateva; Will Giles
Journal:  J Radiosurg SBRT       Date:  2020

2.  Patient-related quality assurance with different combinations of treatment planning systems, techniques, and machines : A multi-institutional survey.

Authors:  Beatrice Steiniger; René Berger; Sabine Eilzer; Christine Kornhuber; Kathleen Lorenz; Torsten Peil; Carsten Reiffenstuhl; Johannes Schilz; Dirk Schröder; Michael Schwedas; Stephanie Pensold; Mathias Walke; Kirsten Weibert; Ulrich Wolf; Tilo Wiezorek
Journal:  Strahlenther Onkol       Date:  2016-11-03       Impact factor: 3.621

3.  Volumetric-modulated arc therapy for oropharyngeal carcinoma: a dosimetric and delivery efficiency comparison with static-field IMRT.

Authors:  Xiaofang Dai; Yingchao Zhao; Zhiwen Liang; Meera Dassarath; Lu Wang; Lihui Jin; Lili Chen; James Dong; Robert A Price; C-M Ma
Journal:  Phys Med       Date:  2014-10-03       Impact factor: 2.685

4.  Clinical experience with machine log file software for volumetric-modulated arc therapy techniques.

Authors:  Luis Alberto Vazquez-Quino; Claudia Ivette Huerta-Hernandez; Dharanipathy Rangaraj
Journal:  Proc (Bayl Univ Med Cent)       Date:  2017-07

5.  Dosimetric verification by using the ArcCHECK system and 3DVH software for various target sizes.

Authors:  Jin Ho Song; Hun-Joo Shin; Chul Seung Kay; Seok Hyun Son
Journal:  PLoS One       Date:  2015-03-25       Impact factor: 3.240

6.  Comparisons of volumetric modulated arc therapy (VMAT) quality assurance (QA) systems: sensitivity analysis to machine errors.

Authors:  Bin Liang; Bo Liu; Fugen Zhou; Fang-Fang Yin; Qiuwen Wu
Journal:  Radiat Oncol       Date:  2016-11-07       Impact factor: 3.481

7.  Comparison of patient-specific intensity modulated radiation therapy quality assurance for the prostate across multiple institutions.

Authors:  Kazuki Kubo; Hajime Monzen; Kohei Shimomura; Kenji Matsumoto; Tomoharu Sato; Mikoto Tamura; Kiyoshi Nakamatsu; Kentaro Ishii; Ryu Kawamorita
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-17

8.  Simplification of head and neck volumetric modulated arc therapy patient-specific quality assurance, using a Delta4 PT.

Authors:  Motoharu Sasaki; Wataru Sugimoto; Hitoshi Ikushima
Journal:  Rep Pract Oncol Radiother       Date:  2020-08-08

9.  Comparison of dose calculations between pencil-beam and Monte Carlo algorithms of the iPlan RT in arc therapy using a homogenous phantom with 3DVH software.

Authors:  Jin Ho Song; Hun-Joo Shin; Chul Seung Kay; Soo-Min Chae; Seok Hyun Son
Journal:  Radiat Oncol       Date:  2013-12-05       Impact factor: 3.481

10.  Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance.

Authors:  Hosang Jin; Vance P Keeling; Daniel A Johnson; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.