Continuous glucose monitoring: quality of hypoglycaemia detection.

E Zijlstra1, T Heise, L Nosek, L Heinemann, S Heckermann.   

Abstract

AIMS: To evaluate the accuracy of a (widely used) continuous glucose monitoring (CGM)-system and its ability to detect hypoglycaemic events.
METHODS: A total of 18 patients with type 1 diabetes mellitus used continuous glucose monitoring (Guardian REAL-Time CGMS) during two 9-day in-house periods. A hypoglycaemic threshold alarm alerted patients to sensor readings <70 mg/dl. Continuous glucose monitoring sensor readings were compared to laboratory reference measurements taken every 4 h and in case of a hypoglycaemic alarm.
RESULTS: A total of 2317 paired data points were evaluated. Overall, the mean absolute relative difference (MARD) was 16.7%. The percentage of data points in the clinically accurate or acceptable Clarke Error Grid zones A + B was 94.6%. In the hypoglycaemic range, accuracy worsened (MARD 38.8%) leading to a failure to detect more than half of the true hypoglycaemic events (sensitivity 37.5%). Furthermore, more than half of the alarms that warn patients for hypoglycaemia were false (false alert rate 53.3%). Above the low alert threshold, the sensor confirmed 2077 of 2182 reference values (specificity 95.2%).
CONCLUSIONS: Patients using continuous glucose monitoring should be aware of its limitation to accurately detect hypoglycaemia.
© 2012 Blackwell Publishing Ltd.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22974231     DOI: 10.1111/dom.12001

Source DB:  PubMed          Journal:  Diabetes Obes Metab        ISSN: 1462-8902            Impact factor:   6.577


  30 in total

Review 1.  Sensor-Augmented Insulin Pumps and Hypoglycemia Prevention in Type 1 Diabetes.

Authors:  Isabelle Steineck; Ajenthen Ranjan; Kirsten Nørgaard; Signe Schmidt
Journal:  J Diabetes Sci Technol       Date:  2016-10-06

Review 2.  AP@home: The Artificial Pancreas Is Now at Home.

Authors:  Lutz Heinemann; Carsten Benesch; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2016-06-28

3.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

4.  A toolbox to improve algorithms for insulin-dosing decision support.

Authors:  K Donsa; P Beck; J Plank; L Schaupp; J K Mader; T Truskaller; B Tschapeller; B Höll; S Spat; T R Pieber
Journal:  Appl Clin Inform       Date:  2014-06-11       Impact factor: 2.342

5.  Professional continuous glucose monitoring in subjects with type 1 diabetes: retrospective hypoglycemia detection.

Authors:  Morten Hasselstrøm Jensen; Toke Folke Christensen; Lise Tarnow; Zeinab Mahmoudi; Mette Dencker Johansen; Ole Kristian Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

6.  Glycemic variability during algorithmic titration of insulin among hospitalized patients with type 2 diabetes and heart failure.

Authors:  Kathleen Dungan; Philip Binkley; Kwame Osei
Journal:  J Diabetes Complications       Date:  2015-09-21       Impact factor: 2.852

Review 7.  Current Trends in Continuous Glucose Monitoring.

Authors:  Volker Lodwig; Bernhard Kulzer; Oliver Schnell; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2014-03-13

8.  HypoDE: Research Design and Methods of a Randomized Controlled Study Evaluating the Impact of Real-Time CGM Usage on the Frequency of CGM Glucose Values <55 mg/dl in Patients With Type 1 Diabetes and Problematic Hypoglycemia Treated With Multiple Daily Injections.

Authors:  Lutz Heinemann; Dorothee Deiss; Norbert Hermanns; Claudia Graham; Matthias Kaltheuner; Andreas Liebl; David Price
Journal:  J Diabetes Sci Technol       Date:  2015-03-09

Review 9.  Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices.

Authors:  Guido Freckmann; Stefan Pleus; Mike Grady; Steven Setford; Brian Levy
Journal:  J Diabetes Sci Technol       Date:  2018-11-19

10.  Effect of the approach to insulin therapy on glycaemic fluctuations and autonomic tone in hospitalized patients with diabetes.

Authors:  K M Dungan; K Osei; C Sagrilla; P Binkley
Journal:  Diabetes Obes Metab       Date:  2013-02-17       Impact factor: 6.577

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.