The accuracy and efficacy of real-time continuous glucose monitoring sensor in patients with type 1 diabetes.

John Mastrototaro1, John Shin, Alan Marcus, Giri Sulur.   

Abstract

BACKGROUND: The accuracy and efficacy of the Medtronic Diabetes (Northridge, CA) Real-Time (RT)-Continuous Glucose Monitoring (CGM) sensor were analyzed in 72 subjects with type 1 diabetes.
METHODS: This was a retrospective analysis of 60,050 temporally paired data points (sensor and glucose meter values) obtained during the course of an outpatient ambulatory study evaluating the efficacy of a sensor-augmented pump system in adults and adolescents. Subjects uploaded sensor values and self-monitoring blood glucose data to the CareLink Clinical Application (Medtronic Diabetes) via the Internet, every 2 weeks during the course of the study.
RESULTS: The overall percentage of sensor readings within +/-20% or +/-30% agreement of reference glucose readings was 75.6% and 86.8%, respectively. The highest rate of agreement occurred in the 240-400 mg/dL range, where 79.9% of sensor readings were within +/-20% of meter values and 91.5% of sensor readings were within 30% of meter values. The mean absolute relative difference for all subjects was 15.8%, and the median absolute relative difference was 10.9%. The bias was -2.13 mg/dL. Paired glucose measurements from the RT-CGM and meter demonstrated that 95.9% of paired points in the overall subject population fell in zones A and B of the Clarke Error Grid. Consensus Error Grid Analysis established that 99.2% of paired data points were in zones A and B.
CONCLUSIONS: This study reports the accuracy of a continuous glucose sensor with a large number of paired data points (60,050). RT-CGM is safe and well tolerated and provides readings that are in close agreement with glucose meter values.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18715215     DOI: 10.1089/dia.2007.0291

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  37 in total

1.  Continuous glucose monitoring to assess the ecologic validity of dietary glycemic index and glycemic load.

Authors:  Anthony N Fabricatore; Cara B Ebbeling; Thomas A Wadden; David S Ludwig
Journal:  Am J Clin Nutr       Date:  2011-11-09       Impact factor: 7.045

2.  Continuous glucose monitoring considerations for the development of a closed-loop artificial pancreas system.

Authors:  D Barry Keenan; Benyamin Grosman; Harry W Clark; Anirban Roy; Stuart A Weinzimer; Rajiv V Shah; John J Mastrototaro
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

3.  Use of subcutaneous interstitial fluid glucose to estimate blood glucose: revisiting delay and sensor offset.

Authors:  Kerstin Rebrin; Norman F Sheppard; Garry M Steil
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

4.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

5.  A consensus perceived glycemic variability metric.

Authors:  Cynthia R Marling; Nigel W Struble; Razvan C Bunescu; Jay H Shubrook; Frank L Schwartz
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

6.  Accuracy and reliability of continuous glucose monitoring systems: a head-to-head comparison.

Authors:  Yoeri M Luijf; Julia K Mader; Werner Doll; Thomas Pieber; Anne Farret; Jerome Place; Eric Renard; Daniela Bruttomesso; Alessio Filippi; Angelo Avogaro; Sabine Arnolds; Carsten Benesch; Lutz Heinemann; J Hans DeVries
Journal:  Diabetes Technol Ther       Date:  2013-05-07       Impact factor: 6.118

7.  Characterizing blood glucose variability using new metrics with continuous glucose monitoring data.

Authors:  Cynthia R Marling; Jay H Shubrook; Stanley J Vernier; Matthew T Wiley; Frank L Schwartz
Journal:  J Diabetes Sci Technol       Date:  2011-07-01

8.  Benefits of blinded continuous glucose monitoring during a randomized clinical trial.

Authors:  Douglas Muchmore; Melissa Sharp; Daniel Vaughn
Journal:  J Diabetes Sci Technol       Date:  2011-05-01

Review 9.  New-generation diabetes management: glucose sensor-augmented insulin pump therapy.

Authors:  Eda Cengiz; Jennifer L Sherr; Stuart A Weinzimer; William V Tamborlane
Journal:  Expert Rev Med Devices       Date:  2011-07       Impact factor: 3.166

10.  Closed-loop insulin delivery using a subcutaneous glucose sensor and intraperitoneal insulin delivery: feasibility study testing a new model for the artificial pancreas.

Authors:  Eric Renard; Jerome Place; Martin Cantwell; Hugues Chevassus; Cesar C Palerm
Journal:  Diabetes Care       Date:  2009-10-21       Impact factor: 19.112

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.