Comparison of iSED and Ves-Matic Cube 200 Erythrocyte Sedimentation Rate Measurements With Westergren Method.

Nihal Boğdaycioğlu1, Fatma Meric Yilmaz1,2, Sevilay Sezer1, Esra Oğuz1.   

Abstract

BACKGROUND: The erythrocyte sedimentation rate (ESR) test is performed widely in laboratories. Besides the traditional Westergren method, new methods have been developed for ESR measurements. We aimed to compare the instruments using new methods, iSED (Alcor Scientific) and Ves-Matic Cube 200 (Diesse Diagnostica Senese, Italy) with the Westergren method.
METHODS: Blood samples from 136 patients were taken into EDTA tubes for automated analyzers and citrated tubes for Westergren method. Patients were divided into three groups- low, medium, and high-according to their sedimentation rates. Precision, stability, and interference studies of the methods were performed.
RESULTS: The iSED sedimentation method (n = 136) yielded a slope of (0.61-0.84), with an intercept of (-2.32 to 1.89), which resulted in a mean bias of 13; and the Ves-Matic Cube 200 method (n = 136) yielded a slope of (0.85-1.00), with an intercept of (0.00-3.07), which resulted in a mean bias of 1.4 in Passing-Bablok regression analysis compared to the reference method.
CONCLUSION: iSED sedimentation showed a poor correlation and a high bias (>10%) with the Westergren method, and Ves-Matic Cube 200 method showed a higher correlation and a lower bias than the iSED device when compared with Westergren reference method. These instruments should be carefully monitored.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  Citrate; EDTA; Erythrocyte sedimentation rate; Ves-Matic Cube 200; Westergren; iSED

Mesh:

Year:  2014        PMID: 25131836      PMCID: PMC6807232          DOI: 10.1002/jcla.21786

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  9 in total

1.  Validation of the Diesse Mini-Ves erythrocyte sedimentation rate (ESR) analyzer using the Westergren ESR method in patients with systemic inflammatory conditions.

Authors:  Marc R Happe; Daniel F Battafarano; David P Dooley; Thomas A Rennie; Frederick T Murphy; Thomas J Casey; John A Ward
Journal:  Am J Clin Pathol       Date:  2002-07       Impact factor: 2.493

2.  Evaluation of the Ves-Matic Cube 200 erythrocyte sedimentation method: comparison with Westergren-based methods.

Authors:  Joyce Curvers; Jurgen Kooren; Maartje Laan; Edwin van Lierop; Daan van de Kerkhof; Volkher Scharnhorst; Martien Herruer
Journal:  Am J Clin Pathol       Date:  2010-10       Impact factor: 2.493

3.  Assessment of Diesse Ves-matic automated system for measuring erythrocyte sedimentation rate.

Authors:  M Caswell; J Stuart
Journal:  J Clin Pathol       Date:  1991-11       Impact factor: 3.411

4.  Evaluation of Ves-Matic Cube 200--an automated system for the measurement of the erythrocyte sedimentation rate.

Authors:  E Perovic; L Bakovic; A Valcic
Journal:  Int J Lab Hematol       Date:  2009-01-19       Impact factor: 2.877

5.  Evaluation of Ves-Matic Cube 200 for erythrocyte sedimentation rate determination.

Authors:  Sevilay Sezer; Fatma Meriç Yilmaz; Oğuzhan Kaya; Sema Uysal
Journal:  J Clin Lab Anal       Date:  2013-09       Impact factor: 2.352

6.  Three-way comparison of methods for the measurement of the erythrocyte sedimentation rate.

Authors:  Johnny Ndoni Mahlangu; Melony Davids
Journal:  J Clin Lab Anal       Date:  2008       Impact factor: 2.352

7.  The influence of sialic acid on the ESR.

Authors:  R A Lugton
Journal:  Med Lab Sci       Date:  1989-01

8.  ICSH recommendations for measurement of erythrocyte sedimentation rate. International Council for Standardization in Haematology (Expert Panel on Blood Rheology)

Authors: 
Journal:  J Clin Pathol       Date:  1993-03       Impact factor: 3.411

9.  A Comparison Between the StaRRsed Auto-Compact Erythrocyte Sedimentation Rate Instrument and the Westergren Method.

Authors:  Juha Horsti; Riikka Rontu; Auni Collings
Journal:  J Clin Med Res       Date:  2010-12-11
  9 in total
  2 in total

1.  Differences in erythrocyte sedimentation rates using a modified Westergren method and an alternate method.

Authors:  Elise Schapkaitz; Shilla RabuRabu; Marcel Engelbrecht
Journal:  J Clin Lab Anal       Date:  2018-08-26       Impact factor: 2.352

2.  A rapid and accurate method for estimating the erythrocyte sedimentation rate using a hematocrit-corrected optical aggregation index.

Authors:  Makoto Higuchi; Nobuo Watanabe
Journal:  PLoS One       Date:  2022-07-12       Impact factor: 3.752

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.