Literature DB >> 24038221

Evaluation of Ves-Matic Cube 200 for erythrocyte sedimentation rate determination.

Sevilay Sezer1, Fatma Meriç Yilmaz, Oğuzhan Kaya, Sema Uysal.   

Abstract

BACKGROUND: International Council for Standardization in Haematology suggested Westergren method as the reference method to analyze erythrocyte sedimentation rate (ESR). However, in recent years closed automated systems that measure ESR directly from a capped EDTA blood sample tube have been developed. We evaluated the analytic performance of one of these new methods, the Ves-Matic Cube 200.
METHODS: K(2) EDTA and citrated blood samples were taken from 101 randomly selected outpatients in Ankara Numune Education and Research Hospital. The ESR results using Ves-Matic Cube 200 and Westergren as reference method from 101 patients were compared and interference studies were performed.
RESULTS: We found the mean difference between the two methods as 0.19 ± 15.85 mm/hr (-3.317 to 2.940 mm/hr, 95% confidence interval). Regression analysis yielded the equation "y = -2.59 + 1.15x" between the two methods (r = 0.82).
CONCLUSIONS: Ves-Matic Cube 200 should be monitored carefully for good quality control. Temperature correction should be applied to study control material as recommended by the manufacturer. Ves-Matic Cube 200 device should be monitored carefully, performance studies should be performed, and the results should be checked in order to eliminate the random errors during the routine studies.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  ESR; Ves-Matic Cube 200; Westergren; comparison; evaluation

Mesh:

Substances:

Year:  2013        PMID: 24038221      PMCID: PMC6807416          DOI: 10.1002/jcla.21612

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  10 in total

1.  Evaluation of the Ves-Matic Cube 200 erythrocyte sedimentation method: comparison with Westergren-based methods.

Authors:  Joyce Curvers; Jurgen Kooren; Maartje Laan; Edwin van Lierop; Daan van de Kerkhof; Volkher Scharnhorst; Martien Herruer
Journal:  Am J Clin Pathol       Date:  2010-10       Impact factor: 2.493

2.  Comparison of erythrocyte sedimentation rate measurement by the automated SEDIsystem and conventional Westergren method using the Bland and Altman statistical method.

Authors:  Suad M AlFadhli; Anwar M Al-Awadhi
Journal:  Med Princ Pract       Date:  2005 Jul-Aug       Impact factor: 1.927

3.  Evaluation of Ves-Matic Cube 200--an automated system for the measurement of the erythrocyte sedimentation rate.

Authors:  E Perovic; L Bakovic; A Valcic
Journal:  Int J Lab Hematol       Date:  2009-01-19       Impact factor: 2.877

Review 4.  Laboratory evaluation in neck pain.

Authors:  Susan J Dreyer; Scott D Boden
Journal:  Phys Med Rehabil Clin N Am       Date:  2003-08       Impact factor: 1.784

5.  Three-way comparison of methods for the measurement of the erythrocyte sedimentation rate.

Authors:  Johnny Ndoni Mahlangu; Melony Davids
Journal:  J Clin Lab Anal       Date:  2008       Impact factor: 2.352

6.  Erythrocyte sedimentation rate measurement by VES Matic Cube 80 in relation to inflammation plasma proteins.

Authors:  Helena Cerutti; Chiara Muzzi; Roberto Leoncini; Carlo Scapellato; Alessio Cortelazzo; Emiliano Furlani; Roberto Guerranti
Journal:  J Clin Lab Anal       Date:  2011       Impact factor: 2.352

7.  Automated measurement of erythrocyte sedimentation rate and its relation to red blood cell concentration and plasma proteins.

Authors:  Y Imafuku; H Yoshida; S Greenfield; A Rabinovitch
Journal:  Hematol Cell Ther       Date:  1998-02

8.  Recommendation of measurement of erythrocyte sedimentation rate of human blood.

Authors: 
Journal:  Am J Clin Pathol       Date:  1977-10       Impact factor: 2.493

9.  A Comparison Between the StaRRsed Auto-Compact Erythrocyte Sedimentation Rate Instrument and the Westergren Method.

Authors:  Juha Horsti; Riikka Rontu; Auni Collings
Journal:  J Clin Med Res       Date:  2010-12-11

10.  Sigma ESR: an erythrocyte sedimentation rate adjusted for the hematocrit and hemoglobin concentration.

Authors:  Yves Pawlotsky; Jean Goasguen; Pascal Guggenbuhl; Eric Veillard; Christine Jard; Michel Pouchard; Aleth Perdriger; Jean Meadeb; Gérard Chalès
Journal:  Am J Clin Pathol       Date:  2004-11       Impact factor: 2.493

  10 in total
  2 in total

1.  Comparison of iSED and Ves-Matic Cube 200 Erythrocyte Sedimentation Rate Measurements With Westergren Method.

Authors:  Nihal Boğdaycioğlu; Fatma Meric Yilmaz; Sevilay Sezer; Esra Oğuz
Journal:  J Clin Lab Anal       Date:  2014-08-17       Impact factor: 2.352

2.  Test-1 analyzer and conventional Westergren method for erythrocyte sedimentation rate: A comparative study between two laboratories.

Authors:  Cigdem Sonmez; Ozlem Ceylan Dogan; Aysegul Ozturk Kaymak; Nedim Akkaya; Kadir Okhan Akin; Gulcan Guntas
Journal:  J Clin Lab Anal       Date:  2018-01-17       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.