Literature DB >> 24989384

Automatic tracking of arbitrarily shaped implanted markers in kilovoltage projection images: a feasibility study.

Rajesh Regmi1, D Michael Lovelock1, Margie Hunt1, Pengpeng Zhang1, Hai Pham1, Jianping Xiong1, Ellen D Yorke1, Karyn A Goodman2, Andreas Rimner2, Hassan Mostafavi3, Gig S Mageras1.   

Abstract

PURPOSE: Certain types of commonly used fiducial markers take on irregular shapes upon implantation in soft tissue. This poses a challenge for methods that assume a predefined shape of markers when automatically tracking such markers in kilovoltage (kV) radiographs. The authors have developed a method of automatically tracking regularly and irregularly shaped markers using kV projection images and assessed its potential for detecting intrafractional target motion during rotational treatment.
METHODS: Template-based matching used a normalized cross-correlation with simplex minimization. Templates were created from computed tomography (CT) images for phantom studies and from end-expiration breath-hold planning CT for patient studies. The kV images were processed using a Sobel filter to enhance marker visibility. To correct for changes in intermarker relative positions between simulation and treatment that can introduce errors in automatic matching, marker offsets in three dimensions were manually determined from an approximately orthogonal pair of kV images. Two studies in anthropomorphic phantom were carried out, one using a gold cylindrical marker representing regular shape, another using a Visicoil marker representing irregular shape. Automatic matching of templates to cone beam CT (CBCT) projection images was performed to known marker positions in phantom. In patient data, automatic matching was compared to manual matching as an approximate ground truth. Positional discrepancy between automatic and manual matching of less than 2 mm was assumed as the criterion for successful tracking. Tracking success rates were examined in kV projection images from 22 CBCT scans of four pancreas, six gastroesophageal junction, and one lung cancer patients. Each patient had at least one irregularly shaped radiopaque marker implanted in or near the tumor. In addition, automatic tracking was tested in intrafraction kV images of three lung cancer patients with irregularly shaped markers during 11 volumetric modulated arc treatments. Purpose-built software developed at our institution was used to create marker templates and track the markers embedded in kV images.
RESULTS: Phantom studies showed mean ± standard deviation measurement uncertainty of automatic registration to be 0.14 ± 0.07 mm and 0.17 ± 0.08 mm for Visicoil and gold cylindrical markers, respectively. The mean success rate of automatic tracking with CBCT projections (11 frames per second, fps) of pancreas, gastroesophageal junction, and lung cancer patients was 100%, 99.1% (range 98%-100%), and 100%, respectively. With intrafraction images (approx. 0.2 fps) of lung cancer patients, the success rate was 98.2% (range 97%-100%), and 94.3% (range 93%-97%) using templates from 1.25 mm and 2.5 mm slice spacing CT scans, respectively. Correction of intermarker relative position was found to improve the success rate in two out of eight patients analyzed.
CONCLUSIONS: The proposed method can track arbitrary marker shapes in kV images using templates generated from a breath-hold CT acquired at simulation. The studies indicate its feasibility for tracking tumor motion during rotational treatment. Investigation of the causes of misregistration suggests that its rate of incidence can be reduced with higher frequency of image acquisition, templates made from smaller CT slice spacing, and correction of changes in intermarker relative positions when they occur.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24989384      PMCID: PMC4187346          DOI: 10.1118/1.4881335

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

1.  Image-guided radiosurgery for the spine and pancreas.

Authors:  M J Murphy; J R Adler; M Bodduluri; J Dooley; K Forster; J Hai; Q Le; G Luxton; D Martin; J Poen
Journal:  Comput Aided Surg       Date:  2000

Review 2.  Errors and margins in radiotherapy.

Authors:  Marcel van Herk
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

3.  Registration accuracy and possible migration of internal fiducial gold marker implanted in prostate and liver treated with real-time tumor-tracking radiation therapy (RTRT).

Authors:  Kei Kitamura; Hiroki Shirato; Shinichi Shimizu; Nobuo Shinohara; Toru Harabayashi; Tadashi Shimizu; Yoshihisa Kodama; Hideho Endo; Rikiya Onimaru; Seiko Nishioka; Hidefumi Aoyama; Kazuhiko Tsuchiya; Kazuo Miyasaka
Journal:  Radiother Oncol       Date:  2002-03       Impact factor: 6.280

Review 4.  Overview of image-guided radiation therapy.

Authors:  Lei Xing; Brian Thorndyke; Eduard Schreibmann; Yong Yang; Tian-Fang Li; Gwe-Ya Kim; Gary Luxton; Albert Koong
Journal:  Med Dosim       Date:  2006       Impact factor: 1.482

5.  Fluoroscopic tracking of multiple implanted fiducial markers using multiple object tracking.

Authors:  Xiaoli Tang; Greg C Sharp; Steve B Jiang
Journal:  Phys Med Biol       Date:  2007-06-11       Impact factor: 3.609

6.  The deep inspiration breath-hold technique in the treatment of inoperable non-small-cell lung cancer.

Authors:  K E Rosenzweig; J Hanley; D Mah; G Mageras; M Hunt; S Toner; C Burman; C C Ling; B Mychalczak; Z Fuks; S A Leibel
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-08-01       Impact factor: 7.038

7.  Fluoroscopic study of tumor motion due to breathing: facilitating precise radiation therapy for lung cancer patients.

Authors:  Q S Chen; M S Weinhous; F C Deibel; J P Ciezki; R M Macklis
Journal:  Med Phys       Date:  2001-09       Impact factor: 4.071

8.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

9.  Marker seed migration in prostate localization.

Authors:  Matthew M Poggi; Dean A Gant; Wilfred Sewchand; William B Warlick
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-08-01       Impact factor: 7.038

10.  Intrafractional tumor motion: lung and liver.

Authors:  Hiroki Shirato; Yvette Seppenwoolde; Kei Kitamura; Rikiya Onimura; Shinichi Shimizu
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

View more
  11 in total

1.  Optimizing fiducial visibility on periodically acquired megavoltage and kilovoltage image pairs during prostate volumetric modulated arc therapy.

Authors:  Pengpeng Zhang; Laura Happersett; Bosky Ravindranath; Michael Zelefsky; Gig Mageras; Margie Hunt
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

2.  Technical Note: Intrafractional changes in time lag relationship between anterior-posterior external and superior-inferior internal motion signals in abdominal tumor sites.

Authors:  Rajesh Regmi; D Michael Lovelock; Pengpeng Zhang; Hai Pham; Jianping Xiong; Ellen D Yorke; Karyn A Goodman; Abraham J Wu; Gig S Mageras
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

3.  Evaluation of respiratory motion-corrected cone-beam CT at end expiration in abdominal radiotherapy sites: a prospective study.

Authors:  Russell E Kincaid; Agung E Hertanto; Yu-Chi Hu; Abraham J Wu; Karyn A Goodman; Hai D Pham; Ellen D Yorke; Qinghui Zhang; Qing Chen; Gig S Mageras
Journal:  Acta Oncol       Date:  2018-01-19       Impact factor: 4.089

4.  Automated target tracking in kilovoltage images using dynamic templates of fiducial marker clusters.

Authors:  Warren G Campbell; Moyed Miften; Bernard L Jones
Journal:  Med Phys       Date:  2017-02       Impact factor: 4.071

5.  Investigation of fiducial marker recognition possibility by water equivalent length in real-time tracking radiotherapy.

Authors:  Kenji Yasue; Hiraku Fuse; Yuto Asano; Miho Kato; Kazuya Shinoda; Hideaki Ikoma; Tatsuya Fujisaki; Yoshio Tamaki
Journal:  Jpn J Radiol       Date:  2021-10-16       Impact factor: 2.374

6.  Couch and multileaf collimator tracking: A clinical feasibility study for pancreas and liver treatment.

Authors:  Lei Zhang; Thomas LoSasso; Pengpeng Zhang; Margie Hunt; Gig Mageras; Grace Tang
Journal:  Med Phys       Date:  2020-09-11       Impact factor: 4.071

7.  Intrafractional 3D localization using kilovoltage digital tomosynthesis for sliding-window intensity modulated radiation therapy.

Authors:  Pengpeng Zhang; Margie Hunt; Hai Pham; Grace Tang; Gig Mageras
Journal:  Phys Med Biol       Date:  2015-08-25       Impact factor: 3.609

8.  Fundamental study on quality assurance (QA) procedures for a real-time tumor tracking radiotherapy (RTRT) system from the viewpoint of imaging devices.

Authors:  Suguru Kimura; Naoki Miyamoto; Kenneth L Sutherland; Ryusuke Suzuki; Hiroki Shirato; Masayori Ishikawa
Journal:  J Appl Clin Med Phys       Date:  2021-06-02       Impact factor: 2.102

9.  Simultaneous MV-kV imaging for intrafractional motion management during volumetric-modulated arc therapy delivery.

Authors:  Margie A Hunt; Mark Sonnick; Hai Pham; Rajesh Regmi; Jian-ping Xiong; Daniel Morf; Gig S Mageras; Michael Zelefsky; Pengpeng Zhang
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

10.  Automatic patient positioning and gating window settings in respiratory-gated stereotactic body radiation therapy for pancreatic cancer using fluoroscopic imaging.

Authors:  Niclas Pettersson; Daniel Simpson; Todd Atwood; Jona Hattangadi-Gluth; James Murphy; Laura Cerviño
Journal:  J Appl Clin Med Phys       Date:  2018-01-27       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.