Assessment of dosimetrical performance in 11 Varian a-Si-500 electronic portal imaging devices.

Awusi Kavuma1, Martin Glegg, Garry Currie, Alex Elliott.   

Abstract

Dosimetrical characteristics of 11 Varian a-Si-500 electronic portal imaging devices (EPIDs) in clinical use for periods ranging between 10 and 86 months were investigated for consistency of performance and portal dosimetry implications. Properties studied include short-term reproducibility, signal linearity with monitor units, response to reference beam, signal uniformity across the detector panel, signal dependence on field size, dose-rate influence, memory effects and image profiles as a function of monitor units. The EPID measurements were also compared with those of the ionization chambers' to ensure stability of the linear accelerators. Depending on their clinical installation date, the EPIDs were interfaced with one of the two different acquisition control software packages, IAS2/IDU-II or IAS3/IDU-20. Both the EPID age and image acquisition system influenced the dosimetric characteristics with the newer version (IAS3 with IDU-20) giving better data reproducibility and linearity fit than the older version (IAS2 with IDU-II). The relative signal response (uniformity) after 50 MU was better than 95% of the central value and independent of detector. Sensitivity for all EPIDs reduced continuously with increasing dose rates for the newer image acquisition software. In the dose-rate range 100-600 MU min(-1), the maximum variation in sensitivity ranged between 1 and 1.8% for different EPIDs. For memory effects, the increase in the measured signal at the centre of the irradiated field for successive images was within 1.8% and 1.0% for the older and newer acquisition systems, respectively. Image profiles acquired at a lower MU in the radial plane (gun-target) had gradients in measured pixel values of up to 25% for the older system. Detectors with software/hardware versions IAS3/IDU-20 have a high degree of accuracy and are more suitable for routine quantitative IMRT dosimetrical verification.

Entities:  

Mesh:

Year:  2008        PMID: 19001691     DOI: 10.1088/0031-9155/53/23/016

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  8 in total

1.  Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules.

Authors:  Janett Liebich; Jörg Licher; Christian Scherf; Eugen Kara; Nadine Koch; Claus Rödel; Ulla Ramm
Journal:  Strahlenther Onkol       Date:  2011-04-26       Impact factor: 3.621

2.  The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device.

Authors:  Omemh Bawazeer; Sisira Herath; Siva Sarasanandarajah; Tomas Kron; Pradip Deb
Journal:  J Med Phys       Date:  2017 Apr-Jun

3.  A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism.

Authors:  Y I Tan; M Metwaly; M Glegg; S P Baggarley; A Elliott
Journal:  Br J Radiol       Date:  2015-05-13       Impact factor: 3.039

4.  Dose-guided radiotherapy for lung tumors.

Authors:  Angelo Piermattei; Andrea Fidanzio; Savino Cilla; Francesca Greco; Luigi Azario; Domenico Sabatino; Mattia Grusio; Mariella Cozzolino; Vincenzo Fusco
Journal:  Med Biol Eng Comput       Date:  2009-12-10       Impact factor: 2.602

5.  Transit dosimetry in dynamic IMRT with an a-Si EPID.

Authors:  Mahsheed Sabet; Pejman Rowshanfarzad; Fred W Menk; Peter B Greer
Journal:  Med Biol Eng Comput       Date:  2014-05-31       Impact factor: 2.602

6.  Evaluation of application of EPID for rapid QC testing of linear accelerator.

Authors:  Shahla Ebrahimi Moghadam; Shahrokh Nasseri; Seyedeh Somayeh Seyedi; Hamid Gholamhosseinian; Mehdi Momennezhad
Journal:  Rep Pract Oncol Radiother       Date:  2018-08-13

7.  Calculation of exit dose for conformal and dynamically-wedged fields, based on water-equivalent path length measured with an amorphous silicon electronic portal imaging device.

Authors:  Awusi Kavuma; Martin Glegg; Mohamed Metwaly; Garry Currie; Alex Elliott
Journal:  J Appl Clin Med Phys       Date:  2011-03-03       Impact factor: 2.102

8.  Dose calibration of EPIDs for segmented IMRT dosimetry.

Authors:  Shrikant Deshpande; Aitang Xing; Lois Holloway; Peter Metcalfe; Philip Vial
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.