Noise, sampling, and the number of projections in cone-beam CT with a flat-panel detector.

Z Zhao1, G J Gang2, J H Siewerdsen3.   

Abstract

PURPOSE: To investigate the effect of the number of projection views on image noise in cone-beam CT (CBCT) with a flat-panel detector.
METHODS: This fairly fundamental consideration in CBCT system design and operation was addressed experimentally (using a phantom presenting a uniform medium as well as statistically motivated "clutter") and theoretically (using a cascaded systems model describing CBCT noise) to elucidate the contributing factors of quantum noise (σ(Q)), electronic noise (σ(E)), and view aliasing (σ(view)). Analysis included investigation of the noise, noise-power spectrum, and modulation transfer function as a function of the number of projections (N(proj)), dose (D(tot)), and voxel size (b(vox)).
RESULTS: The results reveal a nonmonotonic relationship between image noise and N(proj) at fixed total dose: for the CBCT system considered, noise decreased with increasing N(proj) due to reduction of view sampling effects in the regime N(proj) <~200, above which noise increased with N(proj) due to increased electronic noise. View sampling effects were shown to depend on the heterogeneity of the object in a direct analytical relationship to power-law anatomical clutter of the form κ/f(β)--and a general model of individual noise components (σ(Q), σ(E), and σ(view)) demonstrated agreement with measurements over a broad range in N(proj), D(tot), and b(vox).
CONCLUSIONS: The work elucidates fairly basic elements of CBCT noise in a manner that demonstrates the role of distinct noise components (viz., quantum, electronic, and view sampling noise). For configurations fairly typical of CBCT with a flat-panel detector (FPD), the analysis reveals a "sweet spot" (i.e., minimum noise) in the range N(proj) ~ 250-350, nearly an order of magnitude lower in N(proj) than typical of multidetector CT, owing to the relatively high electronic noise in FPDs. The analysis explicitly relates view aliasing and quantum noise in a manner that includes aspects of the object ("clutter") and imaging chain (including nonidealities of detector blur and electronic noise) to provide a more rigorous basis for commonly held intuition and heurism in CBCT system design and operation.

Entities:  

Mesh:

Year:  2014        PMID: 24877820      PMCID: PMC4032401          DOI: 10.1118/1.4875688

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  36 in total

1.  A statistically defined anthropomorphic software breast phantom.

Authors:  Beverly A Lau; Ingrid Reiser; Robert M Nishikawa; Predrag R Bakic
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

2.  Cone-beam CT for breast imaging: Radiation dose, breast coverage, and image quality.

Authors:  Avice O'Connell; David L Conover; Yan Zhang; Posy Seifert; Wende Logan-Young; Chuen-Fu Linda Lin; Lawrence Sahler; Ruola Ning
Journal:  AJR Am J Roentgenol       Date:  2010-08       Impact factor: 3.959

3.  Aliasing: a source of streaks in computed tomograms.

Authors:  R D Brooks; G H Glover; A J Talbert; R L Eisner; F A DiBianca
Journal:  J Comput Assist Tomogr       Date:  1979-08       Impact factor: 1.826

4.  The effects of sampling on CT images.

Authors:  P M Joseph; R D Spital; C D Stockham
Journal:  Comput Tomogr       Date:  1980 Jul-Sep

5.  Detectability in computed tomographic images.

Authors:  K M Hanson
Journal:  Med Phys       Date:  1979 Sep-Oct       Impact factor: 4.071

6.  Computed tomography with fan beam geometry.

Authors:  T M Peters; R M Lewitt
Journal:  J Comput Assist Tomogr       Date:  1977-10       Impact factor: 1.826

7.  Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CT.

Authors:  Junguo Bian; Jeffrey H Siewerdsen; Xiao Han; Emil Y Sidky; Jerry L Prince; Charles A Pelizzari; Xiaochuan Pan
Journal:  Phys Med Biol       Date:  2010-10-20       Impact factor: 3.609

8.  Effect of slice thickness on detectability in breast CT using a prewhitened matched filter and simulated mass lesions.

Authors:  Nathan J Packard; Craig K Abbey; Kai Yang; John M Boone
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

9.  An anthropomorphic breast model for breast imaging simulation and optimization.

Authors:  Baiyu Chen; Jamie Shorey; Robert S Saunders; Samuel Richard; John Thompson; Loren W Nolte; Ehsan Samei
Journal:  Acad Radiol       Date:  2011-03-11       Impact factor: 3.173

10.  Association between power law coefficients of the anatomical noise power spectrum and lesion detectability in breast imaging modalities.

Authors:  Lin Chen; Craig K Abbey; John M Boone
Journal:  Phys Med Biol       Date:  2013-02-19       Impact factor: 3.609

View more
  11 in total

1.  Sparse sampling computed tomography (SpSCT) for detection of pulmonary embolism: a feasibility study.

Authors:  Andreas P Sauter; Felix K Kopp; Rolf Bippus; Julia Dangelmaier; Dominik Deniffel; Alexander A Fingerle; Felix Meurer; Daniela Pfeiffer; Roland Proksa; Ernst J Rummeny; Peter B Noël
Journal:  Eur Radiol       Date:  2019-05-09       Impact factor: 5.315

2.  Self-calibration of cone-beam CT geometry using 3D-2D image registration.

Authors:  S Ouadah; J W Stayman; G J Gang; T Ehtiati; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2016-03-10       Impact factor: 3.609

3.  Task-driven image acquisition and reconstruction in cone-beam CT.

Authors:  Grace J Gang; J Webster Stayman; Tina Ehtiati; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-03-24       Impact factor: 3.609

4.  Multiresolution iterative reconstruction in high-resolution extremity cone-beam CT.

Authors:  Qian Cao; Wojciech Zbijewski; Alejandro Sisniega; John Yorkston; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Phys Med Biol       Date:  2016-10-03       Impact factor: 3.609

5.  Standardization of histogram- and GLCM-based radiomics in the presence of blur and noise.

Authors:  Grace Jianan Gang; Radhika Deshpande; Joseph Webster Stayman
Journal:  Phys Med Biol       Date:  2021-03-15       Impact factor: 4.174

6.  C-arm orbits for metal artifact avoidance (MAA) in cone-beam CT.

Authors:  P Wu; N Sheth; A Sisniega; A Uneri; R Han; R Vijayan; P Vagdargi; B Kreher; H Kunze; G Kleinszig; S Vogt; S F Lo; N Theodore; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2020-08-19       Impact factor: 4.174

7.  Computed Tomography Angiography of Carotid Arteries and Vertebrobasilar System: A Simulation Study for Radiation Dose Reduction.

Authors:  Manuel Kramer; Stephan Ellmann; Thomas Allmendinger; Achim Eller; Ferdinand Kammerer; Matthias S May; João F Baigger; Michael Uder; Michael M Lell
Journal:  Medicine (Baltimore)       Date:  2015-07       Impact factor: 1.889

8.  Technical Note: spektr 3.0-A computational tool for x-ray spectrum modeling and analysis.

Authors:  J Punnoose; J Xu; A Sisniega; W Zbijewski; J H Siewerdsen
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

9.  Optimising complementary soft tissue synchrotron X-ray microtomography for reversibly-stained central nervous system samples.

Authors:  Merrick C Strotton; Andrew J Bodey; Kazimir Wanelik; Michele C Darrow; Esau Medina; Carl Hobbs; Christoph Rau; Elizabeth J Bradbury
Journal:  Sci Rep       Date:  2018-08-13       Impact factor: 4.379

10.  Effects of virtual tube current reduction and sparse sampling on MDCT-based femoral BMD measurements.

Authors:  N Sollmann; K Mei; B J Schwaiger; A S Gersing; F K Kopp; R Bippus; C Maegerlein; C Zimmer; E J Rummeny; J S Kirschke; P B Noël; T Baum
Journal:  Osteoporos Int       Date:  2018-08-24       Impact factor: 4.507

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.