The accuracy of home glucose meters in hypoglycemia.

Alper Sonmez1, Zeynep Yilmaz, Gokhan Uckaya, Selim Kilic, Serkan Tapan, Abdullah Taslipinar, Aydogan Aydogdu, Mahmut Yazici, Mahmut Ilker Yilmaz, Muhittin Serdar, M Kemal Erbil, Mustafa Kutlu.   

Abstract

BACKGROUND: Home glucose meters (HGMs) may not be accurate enough to sense hypoglycemia. We evaluated the accuracy and the capillary and venous comparability of five different HGMs (Optium Xceed [Abbott Diabetes Care, Alameda, CA, USA], Contour TS [Bayer Diabetes Care, Basel, Switzerland], Accu-Chek Go [Roche Ltd., Basel, Switzerland], OneTouch Select [Lifescan, Milpitas, CA, USA], and EZ Smart [Tyson Bioresearch Inc., Chu-Nan, Taiwan]) in an adult population.
METHODS: The insulin hypoglycemia test was performed to 59 subjects (56 males; 23.6 +/- 3.2 years old). Glucose was measured from forearm venous blood and finger capillary samples both before and after regular insulin (0.1 U/kg) was injected. Venous samples were analyzed in the reference laboratory by the hexokinase method. In vitro tests for method comparison and precision analyses were also performed by spiking the glucose-depleted venous blood.
RESULTS: All HGMs failed to sense hypoglycemia to some extend. EZ Smart was significantly inferior in critical error Zone D, and OneTouch Select was significantly inferior in the clinically unimportant error Zone B. Accu-Chek Go, Optium Xceed, and Contour TS had similar performances and were significantly better than the other two HGMs according to error grid analysis or International Organization for Standardization criteria. The in vitro tests were consistent with the above clinical data. The capillary and venous consistencies of Accu-Chek Go and OneTouch Select were better than the other HGMs.
CONCLUSIONS: The present results show that not all the HGMs are accurate enough in low blood glucose levels. The patients and the caregivers should be aware of these restrictions of the HGMs and give more credit to the symptoms of hypoglycemia than the values obtained by the HGMs. Finally, these results indicate that there is a need for the revision of the accuracy standards of HGMs in low blood glucose levels.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20615103     DOI: 10.1089/dia.2009.0183

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  7 in total

1.  Accuracy in blood glucose measurement: what will a tightening of requirements yield?

Authors:  Lutz Heinemann; Volker Lodwig; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 2.  Performance of Cleared Blood Glucose Monitors.

Authors:  David C Klonoff; Priya Prahalad
Journal:  J Diabetes Sci Technol       Date:  2015-06-08

3.  Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting.

Authors:  Christoph Hasslacher; Felix Kulozik; Isabel Platten
Journal:  J Diabetes Sci Technol       Date:  2014-02-18

4.  Accuracy of Human and Veterinary Point-of-Care Glucometers for Use in Rhesus Macaques (Macaca mulatta), Sooty Mangabeys (Cercocebus atys), and Chimpanzees (Pan troglodytes).

Authors:  Elizabeth A Clemmons; Melissa I Stovall; Devon C Owens; Jessica A Scott; Amelia C Jones-Wilkes; Doty J Kempf; Kelly F Ethun
Journal:  J Am Assoc Lab Anim Sci       Date:  2016       Impact factor: 1.232

5.  Comparative Accuracy of 17 Point-of-Care Glucose Meters.

Authors:  Laya Ekhlaspour; Debbie Mondesir; Norman Lautsch; Courtney Balliro; Mallory Hillard; Kendra Magyar; Laura Goergen Radocchia; Aryan Esmaeili; Manasi Sinha; Steven J Russell
Journal:  J Diabetes Sci Technol       Date:  2016-10-03

6.  Analytical Performance Evaluation of Infopia Element™ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose.

Authors:  Hae-Il Park; Seong-Su Lee; Jang-Won Son; Hee-Sun Kwon; Sung Rae Kim; Hyojin Chae; Myungshin Kim; Yonggoo Kim; Soonjib Yoo
Journal:  J Clin Lab Anal       Date:  2016-04-13       Impact factor: 2.352

7.  The association of hyperglycaemia with prevalent tuberculosis: a population-based cross-sectional study.

Authors:  Sarah Lou Bailey; Helen Ayles; Nulda Beyers; Peter Godfrey-Faussett; Monde Muyoyeta; Elizabeth du Toit; John S Yudkin; Sian Floyd
Journal:  BMC Infect Dis       Date:  2016-12-05       Impact factor: 3.090

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.