Order of magnitude sensitivity increase in X-ray Fluorescence Computed Tomography (XFCT) imaging with an optimized spectro-spatial detector configuration: theory and simulation.

Moiz Ahmad, Magdalena Bazalova, Liangzhong Xiang, Lei Xing.   

Abstract

The purpose of this study was to increase the sensitivity of XFCT imaging by optimizing the data acquisition geometry for reduced scatter X-rays. The placement of detectors and detector energy window were chosen to minimize scatter X-rays. We performed both theoretical calculations and Monte Carlo simulations of this optimized detector configuration on a mouse-sized phantom containing various gold concentrations. The sensitivity limits were determined for three different X-ray spectra: a monoenergetic source, a Gaussian source, and a conventional X-ray tube source. Scatter X-rays were minimized using a backscatter detector orientation (scatter direction > 110(°) to the primary X-ray beam). The optimized configuration simultaneously reduced the number of detectors and improved the image signal-to-noise ratio. The sensitivity of the optimized configuration was 10 μg/mL (10 pM) at 2 mGy dose with the mono-energetic source, which is an order of magnitude improvement over the unoptimized configuration (102 pM without the optimization). Similar improvements were seen with the Gaussian spectrum source and conventional X-ray tube source. The optimization improvements were predicted in the theoretical model and also demonstrated in simulations. The sensitivity of XFCT imaging can be enhanced by an order of magnitude with the data acquisition optimization, greatly enhancing the potential of this modality for future use in clinical molecular imaging.

Entities:  

Mesh:

Year:  2014        PMID: 24770916      PMCID: PMC4114314          DOI: 10.1109/TMI.2014.2305101

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  13 in total

Review 1.  The effect of nanoparticle size, shape, and surface chemistry on biological systems.

Authors:  Alexandre Albanese; Peter S Tang; Warren C W Chan
Journal:  Annu Rev Biomed Eng       Date:  2012-04-18       Impact factor: 9.590

2.  Investigation of X-ray fluorescence computed tomography (XFCT) and K-edge imaging.

Authors:  Magdalena Bazalova; Yu Kuang; Guillem Pratx; Lei Xing
Journal:  IEEE Trans Med Imaging       Date:  2012-06-05       Impact factor: 10.048

3.  Uptake and retention of platinum in patients undergoing cisplatin therapy.

Authors:  R Jonson; J Börjesson; S Mattsson; B Unsgaard; A Wallgren
Journal:  Acta Oncol       Date:  1991       Impact factor: 4.089

Review 4.  Biodistribution and toxicity of engineered gold nanoparticles: a review of in vitro and in vivo studies.

Authors:  Nikolai Khlebtsov; Lev Dykman
Journal:  Chem Soc Rev       Date:  2010-11-16       Impact factor: 54.564

5.  Maximum likelihood reconstruction for emission tomography.

Authors:  L A Shepp; Y Vardi
Journal:  IEEE Trans Med Imaging       Date:  1982       Impact factor: 10.048

6.  The feasibility of polychromatic cone-beam x-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects: a Monte Carlo study.

Authors:  Bernard L Jones; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2011-05-31       Impact factor: 3.609

7.  First demonstration of multiplexed X-ray fluorescence computed tomography (XFCT) imaging.

Authors:  Yu Kuang; Guillem Pratx; Magdalena Bazalova; Bowen Meng; Jianguo Qian; Lei Xing
Journal:  IEEE Trans Med Imaging       Date:  2012-10-11       Impact factor: 10.048

8.  X-ray fluorescent CT imaging of cerebral uptake of stable-iodine perfusion agent iodoamphetamine analog IMP in mice.

Authors:  Tohoru Takeda; Jin Wu; Qingkai Huo; Tetsuya Yuasa; Kazuyuki Hyodo; F Avraham Dilmanian; Takao Akatsuka
Journal:  J Synchrotron Radiat       Date:  2008-11-21       Impact factor: 2.616

9.  X-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects using 110 kVp x-rays.

Authors:  Seong-Kyun Cheong; Bernard L Jones; Arsalan K Siddiqi; Fang Liu; Nivedh Manohar; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2010-01-13       Impact factor: 3.609

10.  Single-protein nanomechanical mass spectrometry in real time.

Authors:  M S Hanay; S Kelber; A K Naik; D Chi; S Hentz; E C Bullard; E Colinet; L Duraffourg; M L Roukes
Journal:  Nat Nanotechnol       Date:  2012-08-26       Impact factor: 39.213

View more
  6 in total

1.  Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study.

Authors:  H-M Cho; H Ding; B P Ziemer; S Molloi
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  Experimental validation of L-shell x-ray fluorescence computed tomography imaging: phantom study.

Authors:  Magdalena Bazalova-Carter; Moiz Ahmad; Lei Xing; Rebecca Fahrig
Journal:  J Med Imaging (Bellingham)       Date:  2015-10-08

3.  Proton-induced x-ray fluorescence CT imaging.

Authors:  Magdalena Bazalova-Carter; Moiz Ahmad; Taeko Matsuura; Seishin Takao; Yuto Matsuo; Rebecca Fahrig; Hiroki Shirato; Kikuo Umegaki; Lei Xing
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

4.  Feasibility study of Compton cameras for x-ray fluorescence computed tomography with humans.

Authors:  Don Vernekohl; Moiz Ahmad; Garry Chinn; Lei Xing
Journal:  Phys Med Biol       Date:  2016-11-15       Impact factor: 3.609

5.  Technical Note: A benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system with a high-power x-ray source and transmission CT imaging capability.

Authors:  Nivedh Manohar; Francisco J Reynoso; Sang Hyun Cho
Journal:  Med Phys       Date:  2018-09-14       Impact factor: 4.071

6.  Localising functionalised gold-nanoparticles in murine spinal cords by X-ray fluorescence imaging and background-reduction through spatial filtering for human-sized objects.

Authors:  Florian Grüner; Florian Blumendorf; Oliver Schmutzler; Theresa Staufer; Michelle Bradbury; Ulrich Wiesner; Tanja Rosentreter; Gabriele Loers; David Lutz; Bernadette Richter; Markus Fischer; Florian Schulz; Swantje Steiner; Martin Warmer; Anja Burkhardt; Alke Meents; Matthew Kupinski; Christoph Hoeschen
Journal:  Sci Rep       Date:  2018-11-08       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.