Literature DB >> 24753012

SAR reduction in 7T C-spine imaging using a "dark modes" transmit array strategy.

Yigitcan Eryaman1, Bastien Guerin, Boris Keil, Azma Mareyam, Joaquin L Herraiz, Robert K Kosior, Adrian Martin, Angel Torrado-Carvajal, Norberto Malpica, Juan A Hernandez-Tamames, Emanuele Schiavi, Elfar Adalsteinsson, Lawrence L Wald.   

Abstract

PURPOSE: Local specific absorption rate (SAR) limits many applications of parallel transmit (pTx) in ultra high-field imaging. In this Note, we introduce the use of an array element, which is intentionally inefficient at generating spin excitation (a "dark mode") to attempt a partial cancellation of the electric field from those elements that do generate excitation. We show that adding dipole elements oriented orthogonal to their conventional orientation to a linear array of conventional loop elements can lower the local SAR hotspot in a C-spine array at 7 T.
METHODS: We model electromagnetic fields in a head/torso model to calculate SAR and excitation B1 (+) patterns generated by conventional loop arrays and loop arrays with added electric dipole elements. We utilize the dark modes that are generated by the intentional and inefficient orientation of dipole elements in order to reduce peak 10g local SAR while maintaining excitation fidelity.
RESULTS: For B1 (+) shimming in the spine, the addition of dipole elements did not significantly alter the B1 (+) spatial pattern but reduced local SAR by 36%.
CONCLUSION: The dipole elements provide a sufficiently complimentary B1 (+) and electric field pattern to the loop array that can be exploited by the radiofrequency shimming algorithm to reduce local SAR.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  excitation fidelity; global SAR; local SAR; loop-dipole arrays; parallel transmit; radiative dipole

Mesh:

Year:  2014        PMID: 24753012      PMCID: PMC4761435          DOI: 10.1002/mrm.25246

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  13 in total

1.  Parallel excitation with an array of transmit coils.

Authors:  Yudong Zhu
Journal:  Magn Reson Med       Date:  2004-04       Impact factor: 4.668

2.  A 32-channel lattice transmission line array for parallel transmit and receive MRI at 7 tesla.

Authors:  Gregor Adriany; Edward J Auerbach; Carl J Snyder; Ark Gözübüyük; Steen Moeller; Johannes Ritter; Pierre-François Van de Moortele; Tommy Vaughan; Kâmil Uğurbil
Journal:  Magn Reson Med       Date:  2010-06       Impact factor: 4.668

3.  Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject-dependent transmit phase measurements.

Authors:  Gregory J Metzger; Carl Snyder; Can Akgun; Tommy Vaughan; Kamil Ugurbil; Pierre-Francois Van de Moortele
Journal:  Magn Reson Med       Date:  2008-02       Impact factor: 4.668

4.  Design of a radiative surface coil array element at 7 T: the single-side adapted dipole antenna.

Authors:  A J E Raaijmakers; O Ipek; D W J Klomp; C Possanzini; P R Harvey; J J W Lagendijk; C A T van den Berg
Journal:  Magn Reson Med       Date:  2011-05-31       Impact factor: 4.668

5.  Design, evaluation and application of an eight channel transmit/receive coil array for cardiac MRI at 7.0 T.

Authors:  Andreas Gräßl; Lukas Winter; Christof Thalhammer; Wolfgang Renz; Peter Kellman; Conrad Martin; Florian von Knobelsdorff-Brenkenhoff; Valeriy Tkachenko; Jeanette Schulz-Menger; Thoralf Niendorf
Journal:  Eur J Radiol       Date:  2011-09-14       Impact factor: 3.528

6.  Local specific absorption rate control for parallel transmission by virtual observation points.

Authors:  Gabriele Eichfelder; Matthias Gebhardt
Journal:  Magn Reson Med       Date:  2011-05-20       Impact factor: 4.668

7.  Local specific absorption rate (SAR), global SAR, transmitter power, and excitation accuracy trade-offs in low flip-angle parallel transmit pulse design.

Authors:  Bastien Guérin; Matthias Gebhardt; Steven Cauley; Elfar Adalsteinsson; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2013-06-14       Impact factor: 4.668

8.  Nineteen-channel receive array and four-channel transmit array coil for cervical spinal cord imaging at 7T.

Authors:  Wei Zhao; Julien Cohen-Adad; Jonathan R Polimeni; Boris Keil; Bastien Guerin; Kawin Setsompop; Peter Serano; Azma Mareyam; Philipp Hoecht; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2013-08-20       Impact factor: 4.668

9.  Slice-selective RF pulses for in vivo B1+ inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16-element coil.

Authors:  Kawin Setsompop; Vijayanand Alagappan; Borjan Gagoski; Thomas Witzel; Jonathan Polimeni; Andreas Potthast; Franz Hebrank; Ulrich Fontius; Franz Schmitt; Lawrence L Wald; Elfar Adalsteinsson
Journal:  Magn Reson Med       Date:  2008-12       Impact factor: 4.668

10.  Design and evaluation of a hybrid radiofrequency applicator for magnetic resonance imaging and RF induced hyperthermia: electromagnetic field simulations up to 14.0 Tesla and proof-of-concept at 7.0 Tesla.

Authors:  Lukas Winter; Celal Özerdem; Werner Hoffmann; Davide Santoro; Alexander Müller; Helmar Waiczies; Reiner Seemann; Andreas Graessl; Peter Wust; Thoralf Niendorf
Journal:  PLoS One       Date:  2013-04-22       Impact factor: 3.240

View more
  9 in total

Review 1.  [Safety of implants in high field and ultrahigh field MRI].

Authors:  O Kraff; H H Quick
Journal:  Radiologe       Date:  2019-10       Impact factor: 0.635

2.  Evaluation of a 16-channel transceiver loop + dipole antenna array for human head imaging at 10.5 tesla.

Authors:  Myung Kyun Woo; Lance DelaBarre; Byeong-Yeul Lee; Matt Waks; Russell Luke Lagore; Jerahmie Radder; Yigitcan Eryaman; Kamil Ugurbil; Gregor Adriany
Journal:  IEEE Access       Date:  2020-11-06       Impact factor: 3.367

3.  Radiofrequency heating studies on anesthetized swine using fractionated dipole antennas at 10.5 T.

Authors:  Yiğitcan Eryaman; Russell L Lagore; M Arcan Ertürk; Lynn Utecht; Patrick Zhang; Angel Torrado-Carvajal; Esra Abaci Türk; Lance DelaBarre; Gregory J Metzger; Gregor Adriany; Kâmil Uğurbil; J Thomas Vaughan
Journal:  Magn Reson Med       Date:  2017-03-31       Impact factor: 4.668

Review 4.  Spinal cord MRI at 7T.

Authors:  Robert L Barry; S Johanna Vannesjo; Samantha By; John C Gore; Seth A Smith
Journal:  Neuroimage       Date:  2017-07-03       Impact factor: 6.556

5.  A 7T spine array based on electric dipole transmitters.

Authors:  Qi Duan; Govind Nair; Natalia Gudino; Jacco A de Zwart; Peter van Gelderen; Joe Murphy-Boesch; Daniel S Reich; Jeff H Duyn; Hellmut Merkle
Journal:  Magn Reson Med       Date:  2015-07-20       Impact factor: 4.668

6.  A 16-channel combined loop-dipole transceiver array for 7 Tesla body MRI.

Authors:  M Arcan Ertürk; Alexander J E Raaijmakers; Gregor Adriany; Kâmil Uğurbil; Gregory J Metzger
Journal:  Magn Reson Med       Date:  2016-02-17       Impact factor: 4.668

Review 7.  Parallel transmission for ultrahigh-field imaging.

Authors:  Francesco Padormo; Arian Beqiri; Joseph V Hajnal; Shaihan J Malik
Journal:  NMR Biomed       Date:  2015-05-19       Impact factor: 4.044

8.  Partially orthogonal resonators for magnetic resonance imaging.

Authors:  Jorge Chacon-Caldera; Matthias Malzacher; Lothar R Schad
Journal:  Sci Rep       Date:  2017-02-10       Impact factor: 4.379

9.  Evaluation of 8-Channel Radiative Antenna Arrays for Human Head Imaging at 10.5 Tesla.

Authors:  Myung Kyun Woo; Lance DelaBarre; Matt Thomas Waks; Young Woo Park; Russell Luke Lagore; Steve Jungst; Yigitcan Eryaman; Se-Hong Oh; Kamil Ugurbil; Gregor Adriany
Journal:  Sensors (Basel)       Date:  2021-09-08       Impact factor: 3.576

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.