Clinical prototype of a plastic water-equivalent scintillating fiber dosimeter array for QA applications.

Fréderic Lacroix1, Louis Archambault, Luc Gingras, Mathieu Guillot, A Sam Beddar, Luc Beaulieu.   

Abstract

A clinical prototype of a scintillating fiber dosimeter array for quality assurance applications is presented. The array consists of a linear array of 29 plastic scintillation detectors embedded in a water-equivalent plastic sheet coupled to optical fibers used to guide optical photons to a charge coupled device (CCD) camera. The CCD is packaged in a light-tight, radiation-shielded housing designed for convenient transport. A custom designed connector is used to ensure reproducible mechanical positioning of the optical fibers relative to the CCD. Profile and depth dose characterization measurements are presented and show that the prototype provides excellent dose measurement reproducibility (+/-0.8%) in-field and good accuracy (+/-1.6% maximum deviation) relative to the dose measured with an IC10 ionization chamber.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18777928     DOI: 10.1118/1.2953564

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector.

Authors:  Frédéric Lacroix; Mathieu Guillot; Malcolm McEwen; Claudiu Cojocaru; Luc Gingras; A Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

2.  Transient noise characterization and filtration in CCD cameras exposed to stray radiation from a medical linear accelerator.

Authors:  Louis Archambault; Tina Marie Briere; Sam Beddar
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

3.  Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations.

Authors:  Mathieu Guillot; Luc Gingras; Louis Archambault; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

4.  Simulation of the precision limits of plastic scintillation detectors using optimal component selection.

Authors:  Frédéric Lacroix; Luc Beaulieu; Louis Archambault; A Sam Beddar
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

5.  A design methodology using signal-to-noise ratio for plastic scintillation detectors design and performance optimization.

Authors:  Frédéric Lacroix; A Sam Beddar; Mathieu Guillot; Luc Beaulieu; Luc Gingras
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

6.  Toward a real-time in vivo dosimetry system using plastic scintillation detectors.

Authors:  Louis Archambault; Tina M Briere; Falk Pönisch; Luc Beaulieu; Deborah A Kuban; Andrew Lee; Sam Beddar
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-16       Impact factor: 7.038

7.  Development of a novel multi-point plastic scintillation detector with a single optical transmission line for radiation dose measurement.

Authors:  François Therriault-Proulx; Louis Archambault; Luc Beaulieu; Sam Beddar
Journal:  Phys Med Biol       Date:  2012-10-12       Impact factor: 3.609

8.  Preliminary evaluation of the dosimetric accuracy of the in vivo plastic scintillation detector OARtrac system for prostate cancer treatments.

Authors:  Slade J Klawikowski; Clint Zeringue; Landon S Wootton; Geoffrey S Ibbott; Sam Beddar
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

9.  Two-dimensional real-time quality assurance dosimetry system using μ-Al2O3:C,Mg radioluminescence films.

Authors:  Luana F Nascimento; Dirk Verellen; Jo Goossens; Lara Struelens; Filip Vanhavere; Paul Leblans; Mark Akselrod
Journal:  Phys Imaging Radiat Oncol       Date:  2020-10-05

10.  Effects of Temperature and X-rays on Plastic Scintillating Fiber and Infrared Optical Fiber.

Authors:  Bongsoo Lee; Sang Hun Shin; Kyoung Won Jang; Wook Jae Yoo
Journal:  Sensors (Basel)       Date:  2015-05-11       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.