Arthroprosthetic cobaltism and pseudotumour also occur in patients with small diameter femoral ball head metal-on-metal total hip arthroplasties.

Philip P Roessler1, Florian Witt, Turgay Efe, Jan Schmitt.   

Abstract

Metal-on-metal implants have been widely discredited in public during the past few months. At the same time patients' awareness has risen leading to an increased number of consultations regarding the topic. As hip resurfacings and total hip arthroplasties with large diameter femoral ball heads are already known for excessive metal wear, even small diameter ball heads with low friction get into focus now. We report a case of a 75-year-old man with recurrent hip pain almost 14 years after low friction metal-on-metal total hip arthroplasty. In addition to elevated serum cobalt levels, a type IV sensitivity reaction to cobalt could be noted. Explantation of a highly worn small diameter femoral ball head as well as development of the so-called pseudotumour indicate that all patients should be screened for signs of metal wear regardless of their specific risk factors or the type of metal-on-metal implant.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24682139      PMCID: PMC3975518          DOI: 10.1136/bcr-2013-203362

Source DB:  PubMed          Journal:  BMJ Case Rep        ISSN: 1757-790X


  10 in total

1.  Arthroprosthetic cobaltism associated with metal on metal hip implants.

Authors:  Stephen S Tower
Journal:  BMJ       Date:  2012-01-17

2.  Biomechanical, morphological, and histological analysis of early failures in hip resurfacing arthroplasty.

Authors:  M M Morlock; N Bishop; W Rüther; G Delling; M Hahn
Journal:  Proc Inst Mech Eng H       Date:  2006-02       Impact factor: 1.617

3.  Proposal for a histopathological consensus classification of the periprosthetic interface membrane.

Authors:  L Morawietz; R-A Classen; J H Schröder; C Dynybil; C Perka; A Skwara; J Neidel; T Gehrke; L Frommelt; T Hansen; M Otto; B Barden; T Aigner; P Stiehl; T Schubert; C Meyer-Scholten; A König; P Ströbel; C P Rader; S Kirschner; F Lintner; W Rüther; I Bos; C Hendrich; J Kriegsmann; V Krenn
Journal:  J Clin Pathol       Date:  2006-06       Impact factor: 3.411

Review 4.  Biological effects of clinically relevant wear particles from metal-on-metal hip prostheses.

Authors:  C Brown; J Fisher; E Ingham
Journal:  Proc Inst Mech Eng H       Date:  2006-02       Impact factor: 1.617

Review 5.  Wear and osteolysis in total joint replacements.

Authors:  Y Kadoya; A Kobayashi; H Ohashi
Journal:  Acta Orthop Scand Suppl       Date:  1998-02

6.  What causes unexplained pain in patients with metal-on metal hip devices? A retrieval, histologic, and imaging analysis.

Authors:  Danyal H Nawabi; Nader A Nassif; Huong T Do; Kirsten Stoner; Marcella Elpers; Edwin P Su; Timothy Wright; Hollis G Potter; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

7.  The role of arthroscopy in evaluation of painful hip arthroplasty.

Authors:  Joseph C McCarthy; Stefan R Jibodh; Jo-Ann Lee
Journal:  Clin Orthop Relat Res       Date:  2008-10-02       Impact factor: 4.176

8.  Corrosion at the cone/taper interface leads to failure of large-diameter metal-on-metal total hip arthroplasties.

Authors:  Heiko Meyer; Tina Mueller; Gesine Goldau; Kathrin Chamaon; Marcel Ruetschi; Christoph H Lohmann
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

Review 9.  Metal-on-metal hip resurfacing arthroplasty.

Authors:  Andrew Shimmin; Paul E Beaulé; Pat Campbell
Journal:  J Bone Joint Surg Am       Date:  2008-03       Impact factor: 5.284

Review 10.  Metal ion concentrations in body fluids after implantation of hip replacements with metal-on-metal bearing--systematic review of clinical and epidemiological studies.

Authors:  Albrecht Hartmann; Franziska Hannemann; Jörg Lützner; Andreas Seidler; Hans Drexler; Klaus-Peter Günther; Jochen Schmitt
Journal:  PLoS One       Date:  2013-08-07       Impact factor: 3.240

  10 in total
  5 in total

1.  Adverse Reactions to Metal on Metal Are Not Exclusive to Large Heads in Total Hip Arthroplasty.

Authors:  Adolph V Lombardi; Keith R Berend; Joanne B Adams; Keri L Satterwhite
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

2.  The biological response to nanometre-sized polymer particles.

Authors:  Aiqin Liu; Laura Richards; Catherine L Bladen; Eileen Ingham; John Fisher; Joanne L Tipper
Journal:  Acta Biomater       Date:  2015-05-22       Impact factor: 8.947

3.  Arthroprosthetic cobaltism associated with cardiomyopathy.

Authors:  Ryan S Charette; Alexander L Neuwirth; Charles L Nelson
Journal:  Arthroplast Today       Date:  2016-12-15

4.  A Systematic Review of Systemic Cobaltism After Wear or Corrosion of Chrome-Cobalt Hip Implants.

Authors:  Bradford D Gessner; Thomas Steck; Erik Woelber; Stephen S Tower
Journal:  J Patient Saf       Date:  2019-06       Impact factor: 2.844

5.  Alpha lipoic acid antagonizes cytotoxicity of cobalt nanoparticles by inhibiting ferroptosis-like cell death.

Authors:  Yake Liu; Wenfeng Zhu; Dalong Ni; Zihua Zhou; Jin-Hua Gu; Weinan Zhang; Huanjian Sun; Fan Liu
Journal:  J Nanobiotechnology       Date:  2020-10-02       Impact factor: 10.435

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.