Corrosion at the cone/taper interface leads to failure of large-diameter metal-on-metal total hip arthroplasties.

Heiko Meyer1, Tina Mueller, Gesine Goldau, Kathrin Chamaon, Marcel Ruetschi, Christoph H Lohmann.   

Abstract

BACKGROUND: Metal-on-metal (MoM) THAs have reduced wear rates compared with metal-on-polyethylene. However, elevated serum metal ion levels and pseudotumors have been reported in large MoM articulations. QUESTIONS/PURPOSES: We therefore determined (1) if corrosion occurred at the cone/taper interface leading to instability in patients with large-diameter THAs; (2) how patients presented clinically and radiographically; (3) if adverse periprosthetic tissue reactions occurred; (4) whether metal was released from the implants into the periprosthetic tissues; and (5) if head size correlated with metal release.
METHODS: We reviewed 114 patients who had revisions of large-diameter head MoM articulations. Mean time of implantation was 46 months. To identify adverse reactions and particle load, tissues were stained by hematoxylin and eosin and CD3/CD20/CD68 antibodies. Periprosthetic tissues were analyzed for metal content and distribution in different regions. Electrochemical reactions between the stem and adapter were investigated by a minicell electrode.
RESULTS: Electrochemical studies on the stem and the head adapter showed a risk for galvanic corrosion. Ninety-four percent of patients had instability at the cone/taper interface. All patients presented with early clinical symptoms; 59 patients had radiographic signs of loosening. One hundred four patients had foreign body reactions and necrosis. The largest amounts of metal released were titanium or iron. We found no correlation between head size and metal ion release.
CONCLUSIONS: These findings suggest that in modular cone/taper connections, friction of the MoM articulations may cause failure of the cone/taper interface leading to galvanic corrosion and loosening. It is unclear whether the design of this MoM system provides sufficient stability at the taper.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22864616      PMCID: PMC3462871          DOI: 10.1007/s11999-012-2502-5

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  44 in total

1.  In vitro corrosion testing of modular hip tapers.

Authors:  Jay R Goldberg; Jeremy L Gilbert
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2003-02-15       Impact factor: 3.368

2.  Reactions of the articular capsule to wear products of artificial joint prostheses.

Authors:  H G Willert; M Semlitsch
Journal:  J Biomed Mater Res       Date:  1977-03

3.  Ultra-low wear rates for rigid-on-rigid bearings in total hip replacements.

Authors:  I C Clarke; V Good; P Williams; D Schroeder; L Anissian; A Stark; H Oonishi; J Schuldies; G Gustafson
Journal:  Proc Inst Mech Eng H       Date:  2000       Impact factor: 1.617

4.  Large diameter head metal-on-metal bearings total hip arthroplasty: preliminary results.

Authors:  P Mertl; O Boughebri; E Havet; P Triclot; J-F Lardanchet; A Gabrion
Journal:  Orthop Traumatol Surg Res       Date:  2010-02       Impact factor: 2.256

5.  Concentration of wear products in hair, blood, and urine after total hip replacement.

Authors:  R F Coleman; J Herrington; J T Scales
Journal:  Br Med J       Date:  1973-03-03

6.  Tissue reactions to joint prostheses and the products of wear and corrosion.

Authors:  P A Revell
Journal:  Curr Top Pathol       Date:  1982

7.  Biomaterial-induced sarcoma: A novel model to study preneoplastic change.

Authors:  C J Kirkpatrick; A Alves; H Köhler; J Kriegsmann; F Bittinger; M Otto; D F Williams; R Eloy
Journal:  Am J Pathol       Date:  2000-04       Impact factor: 4.307

8.  Changes seen in lymph nodes draining the sites of large joint prostheses.

Authors:  M H Gray; M L Talbert; W M Talbert; M Bansal; A Hsu
Journal:  Am J Surg Pathol       Date:  1989-12       Impact factor: 6.394

9.  Ion release in patients with metal-on-metal hip bearings in total joint replacement: a comparison with metal-on-polyethylene bearings.

Authors:  L Savarino; D Granchi; G Ciapetti; E Cenni; A Nardi Pantoli; R Rotini; C A Veronesi; N Baldini; A Giunti
Journal:  J Biomed Mater Res       Date:  2002

Review 10.  Cobalt, chromium, and nickel concentrations in body fluids of patients with porous-coated knee or hip prostheses.

Authors:  F W Sunderman; S M Hopfer; T Swift; W N Rezuke; L Ziebka; P Highman; B Edwards; M Folcik; H R Gossling
Journal:  J Orthop Res       Date:  1989       Impact factor: 3.494

View more
  38 in total

1.  Arthroprosthetic cobaltism and pseudotumour also occur in patients with small diameter femoral ball head metal-on-metal total hip arthroplasties.

Authors:  Philip P Roessler; Florian Witt; Turgay Efe; Jan Schmitt
Journal:  BMJ Case Rep       Date:  2014-03-28

2.  Is increased modularity associated with increased fretting and corrosion damage in metal-on-metal total hip arthroplasty devices?: a retrieval study.

Authors:  Genymphas B Higgs; Josa A Hanzlik; Daniel W MacDonald; Jeremy L Gilbert; Clare M Rimnac; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2013-08-01       Impact factor: 4.757

3.  Taper design affects failure of large-head metal-on-metal total hip replacements.

Authors:  Nader A Nassif; Danyal H Nawabi; Kirsten Stoner; Marcella Elpers; Timothy Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

4.  Cemented metal-on-metal total hip replacement with 28-mm head: prospective, long-term, clinical, radiological and metal ions data.

Authors:  Ibrahim A Malek; Sheethal Prasad Patange Subba Rao; Narendra Kumar Rath; U N Mallya
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-12-19

Review 5.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

6.  [Spontaneous reposition of a hip prosthesis after atraumatic dislocation].

Authors:  M S Ostapczuk; T Ritte; M Dicks; M Jonas
Journal:  Orthopade       Date:  2017-10       Impact factor: 1.087

7.  [Failure analysis as basis for quality assurance strategies in implant technology].

Authors:  D Behrend; M Warkentin; D Klüß; R Bader; S Kopp; M Frank; W Mittelmeier
Journal:  Orthopade       Date:  2014-06       Impact factor: 1.087

8.  [Complications of metal-on-metal tribological pairing].

Authors:  M Stiehler; F Zobel; F Hannemann; J Schmitt; J Lützner; S Kirschner; K-P Günther; A Hartmann
Journal:  Orthopade       Date:  2014-01       Impact factor: 1.087

9.  Ion release in ceramic bearings for total hip replacement: Results from an in vitro and an in vivo study.

Authors:  Jan Philippe Kretzer; Ulrike Mueller; Marcus R Streit; Hartmuth Kiefer; Robert Sonntag; Robert M Streicher; Joern Reinders
Journal:  Int Orthop       Date:  2017-07-20       Impact factor: 3.075

10.  Mechanically assisted taper corrosion in modular TKA.

Authors:  Christina M Arnholt; Daniel W MacDonald; Mariya Tohfafarosh; Jeremy L Gilbert; Clare M Rimnac; Steven M Kurtz; Gregg Klein; Michael A Mont; Javad Parvizi; Harold E Cates; Gwo-Chin Lee; Arthur Malkani; Mattheuw Kraay
Journal:  J Arthroplasty       Date:  2014-05-28       Impact factor: 4.757

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.