Literature DB >> 16669399

Biomechanical, morphological, and histological analysis of early failures in hip resurfacing arthroplasty.

M M Morlock1, N Bishop, W Rüther, G Delling, M Hahn.   

Abstract

The present revival of hip resurfacing arthroplasty may be related to an increase in early failures owing to the challenging technique of the procedure. Fifty-five retrieved implants were analysed with respect to wear, cement mantle and cement penetration, fracture and head morphology, as well as standard histology. Femoral neck fractures occurred in median after 102 days. The time to failure was shorter for older women. Major deviations from the suggested cement mantle thickness and cement penetration were found. Indications for high trauma during implantation leading to early failure due to weakening of the femoral neck were also observed. Some failures had signs of pseudarthrosis beneath the implant. Four different fracture patterns with different mean survival times were identified. Observed wear was minor with the exception of that due to alignment mistakes (rim loading). The cups were not damaged by the failures. Histological results indicate that avascular necrosis is not necessarily connected with this kind of endoprosthetic surgery. Most of the failures analysed can probably be attributed to the 'learning curve' effect, which is an unsatisfactory situation.

Entities:  

Mesh:

Year:  2006        PMID: 16669399     DOI: 10.1243/095441105X69015

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  28 in total

1.  Cementation and interface analysis of early failure cases after hip-resurfacing arthroplasty.

Authors:  Matthias Krause; Stefan Breer; Michael Hahn; Wolfgang Rüther; Michael M Morlock; Michael Amling; Jozef Zustin
Journal:  Int Orthop       Date:  2012-01-11       Impact factor: 3.075

2.  Audit-induced change reduces complications of the Ganz trochanteric flip approach.

Authors:  P M Guyver; A Cattell; R P Reddy; C Edwards; D Williams; S M Dixon; M R Norton; E D Fern
Journal:  Ann R Coll Surg Engl       Date:  2010-07-02       Impact factor: 1.891

3.  Arthroprosthetic cobaltism and pseudotumour also occur in patients with small diameter femoral ball head metal-on-metal total hip arthroplasties.

Authors:  Philip P Roessler; Florian Witt; Turgay Efe; Jan Schmitt
Journal:  BMJ Case Rep       Date:  2014-03-28

4.  [Resurfacing arthroplasty of the hip].

Authors:  M Rudert; L Gerdesmeyer; H Rechl; P Juhnke; R Gradinger
Journal:  Orthopade       Date:  2007-04       Impact factor: 1.087

5.  Imageless navigation of hip resurfacing arthroplasty increases the implant accuracy.

Authors:  C Schnurr; J W P Michael; P Eysel; D P König
Journal:  Int Orthop       Date:  2007-12-22       Impact factor: 3.075

6.  Femoral oxygenation during hip resurfacing through the trochanteric flip approach.

Authors:  Robert T Steffen; Darren Fern; Mark Norton; David W Murray; Harinderjit S Gill
Journal:  Clin Orthop Relat Res       Date:  2008-07-16       Impact factor: 4.176

Review 7.  [Surgical approaches in hip resurfacing].

Authors:  L Gerdesmeyer; H Gollwitzer; R Bader; M Rudert
Journal:  Orthopade       Date:  2008-07       Impact factor: 1.087

8.  The minimally invasive anterolateral approach combined with hip onlay resurfacing.

Authors:  Ludger Gerdesmeyer; Hans Gollwitzer; Peter Diehl; Björn Buttgereit; Maximilian Rudert
Journal:  Oper Orthop Traumatol       Date:  2009-03       Impact factor: 1.154

9.  2008 Otto Aufranc Award: component design and technique affect cement penetration in hip resurfacing.

Authors:  Paul E Beaulé; Wadih Y Matar; Philippe Poitras; Kevin Smit; Olivier May
Journal:  Clin Orthop Relat Res       Date:  2008-10-16       Impact factor: 4.176

10.  A new classification for proximal femur bone defects in conservative hip arthroplasty revisions.

Authors:  Filippo Casella; Fabio Favetti; Gabriele Panegrossi; Matteo Papalia; Francesco Falez
Journal:  Int Orthop       Date:  2018-12-11       Impact factor: 3.075

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.