An anatomically realistic temperature phantom for radiofrequency heating measurements.

Nadine N Graedel1, Jonathan R Polimeni1,2, Bastien Guerin1, Borjan Gagoski3, Giorgio Bonmassar, Lawrence L Wald1,2,4.   

Abstract

PURPOSE: An anthropomorphic phantom with realistic electrical properties allows for a more accurate reproduction of tissue current patterns during excitation. A temperature map can then probe the worst-case heating expected in the unperfused case. We describe an anatomically realistic human head phantom that allows rapid three-dimensional (3D) temperature mapping at 7T.
METHODS: The phantom was based on hand-labeled anatomical imaging data and consists of four compartments matching the corresponding human tissues in geometry and electrical properties. The increases in temperature resulting from radiofrequency excitation were measured with MR thermometry using a temperature-sensitive contrast agent (TmDOTMA(-)) validated by direct fiber optic temperature measurements.
RESULTS: Acquisition of 3D temperature maps of the full phantom with a temperature accuracy better than 0.1°C was achieved with an isotropic resolution of 5 mm and acquisition times of 2-4 minutes.
CONCLUSION: Our results demonstrate the feasibility of constructing anatomically realistic phantoms with complex geometries incorporating the ability to measure accurate temperature maps in the phantom. The anthropomorphic temperature phantom is expected to provide a useful tool for the evaluation of the heating effects of both conventional and parallel transmit pulses and help validate electromagnetic and temperature simulations.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  MR safety; local SAR; parallel transmit; phantom; temperature estimation; thermometry

Mesh:

Year:  2014        PMID: 24549755      PMCID: PMC4136997          DOI: 10.1002/mrm.25123

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  37 in total

1.  Fast MRI of RF heating via phase difference mapping.

Authors:  Erik M Shapiro; Arijitt Borthakur; Michael J Shapiro; Ravinder Reddy; John S Leigh
Journal:  Magn Reson Med       Date:  2002-03       Impact factor: 4.668

2.  Transmit SENSE.

Authors:  Ulrich Katscher; Peter Börnert; Christoph Leussler; Johan S van den Brink
Journal:  Magn Reson Med       Date:  2003-01       Impact factor: 4.668

3.  Development of a calibration phantom set for MRI temperature imaging system quality assurance.

Authors:  Xuegang Xin; Jijun Han; Di Wang; Yanqiu Feng; Qianjin Feng; Wufan Chen
Journal:  Acad Radiol       Date:  2012-03-28       Impact factor: 3.173

4.  Measurement of SAR-induced temperature increase in a phantom and in vivo with comparison to numerical simulation.

Authors:  Sukhoon Oh; Yeun-Chul Ryu; Giuseppe Carluccio; Christopher T Sica; Christopher M Collins
Journal:  Magn Reson Med       Date:  2013-06-26       Impact factor: 4.668

5.  Noninvasive magnetic resonance thermography of recurrent rectal carcinoma in a 1.5 Tesla hybrid system.

Authors:  Johanna Gellermann; Waldemar Wlodarczyk; Bert Hildebrandt; Hildegard Ganter; Anett Nicolau; Beate Rau; Wolfgang Tilly; Horst Fähling; Jacek Nadobny; Roland Felix; Peter Wust
Journal:  Cancer Res       Date:  2005-07-01       Impact factor: 12.701

6.  An anthropomorphic phantom for quantitative evaluation of breast MRI.

Authors:  Melanie Freed; Jacco A de Zwart; Jennifer T Loud; Riham H El Khouli; Kyle J Myers; Mark H Greene; Jeff H Duyn; Aldo Badano
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

7.  A low dielectric constant customized phantom design to measure RF coil nonuniformity.

Authors:  P S Tofts; G J Barker; T L Dean; H Gallagher; A P Gregory; R N Clarke
Journal:  Magn Reson Imaging       Date:  1997       Impact factor: 2.546

8.  Variation with age in the volumes of grey and white matter in the cerebral hemispheres of man: measurements with an image analyser.

Authors:  A K Miller; R L Alston; J A Corsellis
Journal:  Neuropathol Appl Neurobiol       Date:  1980 Mar-Apr       Impact factor: 8.090

9.  Experimental and numerical assessment of MRI-induced temperature change and SAR distributions in phantoms and in vivo.

Authors:  Sukhoon Oh; Andrew G Webb; Thomas Neuberger; BuSik Park; Christopher M Collins
Journal:  Magn Reson Med       Date:  2010-01       Impact factor: 4.668

10.  Absolute temperature MR imaging with thulium 1,4,7,10-tetraazacyclododecane-1,4,7,10-tetramethyl-1,4,7,10-tetraacetic acid (TmDOTMA-).

Authors:  Judy R James; Yong Gao; Michael A Miller; Andriy Babsky; Navin Bansal
Journal:  Magn Reson Med       Date:  2009-08       Impact factor: 4.668

View more
  13 in total

1.  Six-coordinate Iron(II) and Cobalt(II) paraSHIFT Agents for Measuring Temperature by Magnetic Resonance Spectroscopy.

Authors:  Pavel B Tsitovich; Jordan M Cox; Jason B Benedict; Janet R Morrow
Journal:  Inorg Chem       Date:  2015-12-30       Impact factor: 5.165

2.  How to design and construct a 3D-printed human head phantom.

Authors:  Sossena Wood; Tiago Martins; Tamer S Ibrahim
Journal:  J 3D Print Med       Date:  2019-08-21

3.  Characterization of a dielectric phantom for high-field magnetic resonance imaging applications.

Authors:  Qi Duan; Jeff H Duyn; Natalia Gudino; Jacco A de Zwart; Peter van Gelderen; Daniel K Sodickson; Ryan Brown
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

4.  Reducing sensitivity losses due to respiration and motion in accelerated echo planar imaging by reordering the autocalibration data acquisition.

Authors:  Jonathan R Polimeni; Himanshu Bhat; Thomas Witzel; Thomas Benner; Thorsten Feiweier; Souheil J Inati; Ville Renvall; Keith Heberlein; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2015-03-23       Impact factor: 4.668

5.  Dual-polarity GRAPPA for simultaneous reconstruction and ghost correction of echo planar imaging data.

Authors:  W Scott Hoge; Jonathan R Polimeni
Journal:  Magn Reson Med       Date:  2015-07-24       Impact factor: 4.668

6.  External Dynamic InTerference Estimation and Removal (EDITER) for low field MRI.

Authors:  Sai Abitha Srinivas; Stephen F Cauley; Jason P Stockmann; Charlotte R Sappo; Christopher E Vaughn; Lawrence L Wald; William A Grissom; Clarissa Z Cooley
Journal:  Magn Reson Med       Date:  2021-09-04       Impact factor: 4.668

7.  Low-rank plus sparse compressed sensing for accelerated proton resonance frequency shift MR temperature imaging.

Authors:  Zhipeng Cao; John C Gore; William A Grissom
Journal:  Magn Reson Med       Date:  2019-01-31       Impact factor: 4.668

8.  Synthesized tissue-equivalent dielectric phantoms using salt and polyvinylpyrrolidone solutions.

Authors:  Carlotta Ianniello; Jacco A de Zwart; Qi Duan; Cem M Deniz; Leeor Alon; Jae-Seung Lee; Riccardo Lattanzi; Ryan Brown
Journal:  Magn Reson Med       Date:  2017-11-20       Impact factor: 4.668

9.  Realistic head-shaped phantom with brain-mimicking metabolites for 7 T spectroscopy and spectroscopic imaging.

Authors:  Ghil Jona; Edna Furman-Haran; Rita Schmidt
Journal:  NMR Biomed       Date:  2020-10-04       Impact factor: 4.044

10.  Design and fabrication of a realistic anthropomorphic heterogeneous head phantom for MR purposes.

Authors:  Sossena Wood; Narayanan Krishnamurthy; Tales Santini; Shailesh B Raval; Nadim Farhat; John Andy Holmes; Tamer S Ibrahim
Journal:  PLoS One       Date:  2017-08-14       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.