Development of a calibration phantom set for MRI temperature imaging system quality assurance.

Xuegang Xin1, Jijun Han, Di Wang, Yanqiu Feng, Qianjin Feng, Wufan Chen.   

Abstract

RATIONALE AND
OBJECTIVES: Magnetic resonance imaging (MRI) temperature imaging systems need to be routinely calibrated to guarantee accurate temperature results and qualified MRI. No independent physical temperature calibration phantom (TCP) set is currently available. An economical TCP set was developed to routinely ensure the quality of MRI temperature imaging system.
MATERIALS AND METHODS: The novel TCP was constructed using a heating unit, temperature sensor, and MRI phantom liquid. A specialized heating unit was developed using carbon fibers. The TCP set design was an integration of the TCP, temperature measurement unit, display unit, and control unit. The proposed MRI calibration kit, which is a combination of the TCP set and standard MRI phantom, was used in the MRI thermometry calibration and MRI quality calibration.
RESULTS: The TCP set provided an efficient, accurate, and homogeneous temperature map as the reference standard temperature for calibration. Accuracy and heating efficiency of the TCP set were 1°C and 1°C/minute, respectively. Calibration of the MRI thermometry and MRI quality were implemented successfully.
CONCLUSION: The proposed TCP set is completely compatible with the MRI system and can be used to calibrate MRI thermometry and MRI quality to ensure the quality performance of the MRI temperature imaging system.
Copyright © 2012 AUR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22459644     DOI: 10.1016/j.acra.2012.02.001

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  1 in total

1.  An anatomically realistic temperature phantom for radiofrequency heating measurements.

Authors:  Nadine N Graedel; Jonathan R Polimeni; Bastien Guerin; Borjan Gagoski; Giorgio Bonmassar; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2014-02-18       Impact factor: 4.668

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.