Migration characteristics and early clinical results of a novel-finned press-fit acetabular cup.

Martin Kaipel1, Anton Prenner, Sebastian Bachl, Sebastian Farr, Günter Sinz.   

Abstract

BACKGROUND: Ana Nova® is a novel-finned press-fit acetabular cup which showed superior biomechanical characteristics in an experimental set-up. Using Einzel Bild Röntgen Analyse (EBRA) measurements should offer the opportunity to predict implant survival at an early stage. The purpose of this study was to assess migration and clinical outcome 2 years after total hip replacement by a novel-finned press-fit acetabular cup.
METHODS: In this study, migration and clinical results of the implant were prospectively assessed in 67 patients. Clinical outcome was assessed using the Harris hip score (HHS). Migration analyses were performed using the computer assisted EBRA system. Data were analyzed for normal distribution using the Kolmogorov-Smirnov test. Group comparisons were performed using the analysis of variance (ANOVA) test. P-values less than 0.05 were considered statistically significant.
RESULTS: At 2 years after surgery, none of the implants needed revision and HHS increased from 39.7 up to 92.2. In contrast to the beneficial clinical outcome, 17 of 44 patients showed increased total migration ( 1 mm/2a).
CONCLUSIONS: Adverse migration data in this study might predict aseptic loosening and decreased survival of the implant. According to previous studies, it is possible that this effect occurred because of limited accuracy of the EBRA system. In our opinion, migration analyses may not be recommended as a screening tool in a 2 year follow-up.

Entities:  

Mesh:

Year:  2014        PMID: 24442859     DOI: 10.1007/s00508-013-0488-y

Source DB:  PubMed          Journal:  Wien Klin Wochenschr        ISSN: 0043-5325            Impact factor:   1.704


  11 in total

1.  Migration of the Duraloc cup at two years.

Authors:  B Stöckl; M Sandow; M Krismer; R Biedermann; C Wimmer; B Frischhut
Journal:  J Bone Joint Surg Br       Date:  1999-01

2.  Radiographic evaluation of the Duraloc cup after 4 years.

Authors:  B Stoeckl; E Brabec; S Wanner; M Krismer; R Biedermann
Journal:  Int Orthop       Date:  2004-10-14       Impact factor: 3.075

3.  Modes of failure of a threaded acetabular cup : a radiographic study with EBRA of 42 revised cups.

Authors:  Thomas Ilchmann; Severin Neher; Franz Maurer; Kuno Weise
Journal:  Int Orthop       Date:  2006-06-08       Impact factor: 3.075

Review 4.  Current status of acetabular fixation in primary total hip arthroplasty.

Authors:  E W Morscher
Journal:  Clin Orthop Relat Res       Date:  1992-01       Impact factor: 4.176

5.  Migration of the Duraloc cup after 5 years.

Authors:  Christoph Stihsen; Christof Pabinger; Roman Radl; Peter Rehak; Reinhard Windhager
Journal:  Int Orthop       Date:  2007-07-03       Impact factor: 3.075

6.  Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins.

Authors:  Wolfgang O Huber; Philip C Noble
Journal:  Int Orthop       Date:  2013-11-22       Impact factor: 3.075

7.  Early migration predicts late aseptic failure of hip sockets.

Authors:  M Krismer; B Stöckl; M Fischer; R Bauer; P Mayrhofer; M Ogon
Journal:  J Bone Joint Surg Br       Date:  1996-05

8.  Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation.

Authors:  W H Harris
Journal:  J Bone Joint Surg Am       Date:  1969-06       Impact factor: 5.284

9.  EBRA: a method to measure migration of acetabular components.

Authors:  M Krismer; R Bauer; J Tschupik; P Mayrhofer
Journal:  J Biomech       Date:  1995-10       Impact factor: 2.712

10.  No differences in outcomes between cemented and uncemented acetabular components after 12-14 years: results from a randomized controlled trial comparing Duraloc with Charnley cups.

Authors:  Kristian Bjørgul; Wendy M Novicoff; S T Andersen; K Brevig; F Thu; M Wiig; O Ahlund
Journal:  J Orthop Traumatol       Date:  2010-03-03
View more
  1 in total

1.  Migration characteristics and early clinical results of the NANOS® short-stem hip arthroplasty.

Authors:  Martin Kaipel; Phillip Grabowiecki; Katrina Sinz; Sebastian Farr; Günter Sinz
Journal:  Wien Klin Wochenschr       Date:  2015-03-05       Impact factor: 1.704

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.