Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins.

Wolfgang O Huber1, Philip C Noble.   

Abstract

PURPOSE: This experimental study was undertaken to examine the fixation characteristics of a six-finned acetabular cup in both primary and revision arthroplasty in comparison with two other commonly used cup designs without fins.
METHODS: All three cup designs (Ananova® [Intraplant], Plasmacup® NSC [Aesculap]; Exceed ABT™ [Biomet]) were implanted into validated models of normal and revision acetabula. The defect models were designed to simulate a dorso-cranial rim defect of 90° width and 10 mm in depth (moderate rim defect) and a dorso-cranial rim defect of 130° width and 15 mm in depth (severe rim defect). The fixation strength of the three cup designs was tested by cyclically edge-loading the implanted cups using a mechanical testing machine.
RESULTS: The six-finned Ananova implant exhibited greater resistance to foam-cup interface motion than both the Plasmacup and Exceed ABT implants. The largest average differences were observed in the resistance to ultimate spin-out, with Ananova outperforming Exceed ABT and Plasmacup by 26% and 17% in the moderate and by 36% and 38% in the severe defect models, respectively.
CONCLUSIONS: The six-finned Ananova cup was significantly more resistant to edge loading both in the normal acetabulum and in acetabula with moderate to severe dorso-cranial rim defects than cup designs without fins, indicating that it may cover a wider range of clinical indications than conventional press-fit cups and provide clinicians with the confidence that, in primary and simple revision arthroplasty, adequate fixation strength can be obtained.

Entities:  

Mesh:

Year:  2013        PMID: 24263376      PMCID: PMC3971254          DOI: 10.1007/s00264-013-2187-8

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  30 in total

1.  Photoelastic analysis of stresses produced by different acetabular cups.

Authors:  M D Ries; A Salehi; J Shea
Journal:  Clin Orthop Relat Res       Date:  1999-12       Impact factor: 4.176

2.  Initial stability of a cementless acetabular cup design: experimental investigation on the effect of adding fins to the rim of the cup.

Authors:  M Baleani; R Fognani; A Toni
Journal:  Artif Organs       Date:  2001-08       Impact factor: 3.094

3.  Load transfer and fixation mode of press-fit acetabular sockets.

Authors:  K-H Widmer; B Zurfluh; E W Morscher
Journal:  J Arthroplasty       Date:  2002-10       Impact factor: 4.757

4.  Acetabular revision in total hip replacement with a press-fit cup.

Authors:  R Elke; B Berli; A Wagner; E W Morscher
Journal:  J Bone Joint Surg Br       Date:  2003-11

5.  Radiographic evaluation of screw position in revision total hip arthroplasty.

Authors:  Daniel D Galat; Jasper A Petrucci; Ray C Wasielewski
Journal:  Clin Orthop Relat Res       Date:  2004-02       Impact factor: 4.176

6.  Structures at risk from medially placed acetabular screws.

Authors:  E M Keating; M A Ritter; P M Faris
Journal:  J Bone Joint Surg Am       Date:  1990-04       Impact factor: 5.284

7.  Observations on the effect of movement on bone ingrowth into porous-surfaced implants.

Authors:  R M Pilliar; J M Lee; C Maniatopoulos
Journal:  Clin Orthop Relat Res       Date:  1986-07       Impact factor: 4.176

8.  Revision of a cemented acetabular component to a cementless acetabular component. A ten to fourteen-year follow-up study.

Authors:  J E Templeton; J J Callaghan; D D Goetz; P M Sullivan; R C Johnston
Journal:  J Bone Joint Surg Am       Date:  2001-11       Impact factor: 5.284

9.  Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up.

Authors:  Pacharapol Udomkiat; Lawrence D Dorr; Zhinian Wan
Journal:  J Bone Joint Surg Am       Date:  2002-07       Impact factor: 5.284

10.  Fixation of trochanteric hip fractures. A cadaver study of static and dynamic loading.

Authors:  S Larsson; M Elloy; L I Hansson
Journal:  Acta Orthop Scand       Date:  1987-08
View more
  3 in total

1.  Migration characteristics and early clinical results of a novel-finned press-fit acetabular cup.

Authors:  Martin Kaipel; Anton Prenner; Sebastian Bachl; Sebastian Farr; Günter Sinz
Journal:  Wien Klin Wochenschr       Date:  2014-01-18       Impact factor: 1.704

2.  Segmental acetabular rim defects, bone loss, oversizing, and press fit cup in total hip arthroplasty evaluated with a probabilistic finite element analysis.

Authors:  Farid Amirouche; Giovanni F Solitro; Amit Walia; Mark Gonzalez; Aimee Bobko
Journal:  Int Orthop       Date:  2016-12-23       Impact factor: 3.075

3.  The effect of dorsal rim loss on the initial stability of the BioMedtrix cementless acetabular cup.

Authors:  Meredith L Montgomery; Stanley E Kim; Jonathan Dyce; Antonio Pozzi
Journal:  BMC Vet Res       Date:  2015-03-18       Impact factor: 2.741

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.