Improvement of attenuation correction in time-of-flight PET/MR imaging with a positron-emitting source.

Pieter Mollet1, Vincent Keereman, Jason Bini, David Izquierdo-Garcia, Zahi A Fayad, Stefaan Vandenberghe.   

Abstract

UNLABELLED: Quantitative PET imaging relies on accurate attenuation correction. Recently, there has been growing interest in combining state-of-the-art PET systems with MR imaging in a sequential or fully integrated setup. As CT becomes unavailable for these systems, an alternative approach to the CT-based reconstruction of attenuation coefficients (μ values) at 511 keV must be found. Deriving μ values directly from MR images is difficult because MR signals are related to the proton density and relaxation properties of tissue. Therefore, most research groups focus on segmentation or atlas registration techniques. Although studies have shown that these methods provide viable solutions in particular applications, some major drawbacks limit their use in whole-body PET/MR. Previously, we used an annulus-shaped PET transmission source inside the field of view of a PET scanner to measure attenuation coefficients at 511 keV. In this work, we describe the use of this method in studies of patients with the sequential time-of-flight (TOF) PET/MR scanner installed at the Icahn School of Medicine at Mount Sinai, New York, NY.
METHODS: Five human PET/MR and CT datasets were acquired. The transmission-based attenuation correction method was compared with conventional CT-based attenuation correction and the 3-segment, MR-based attenuation correction available on the TOF PET/MR imaging scanner.
RESULTS: The transmission-based method overcame most problems related to the MR-based technique, such as truncation artifacts of the arms, segmentation artifacts in the lungs, and imaging of cortical bone. Additionally, the TOF capabilities of the PET detectors allowed the simultaneous acquisition of transmission and emission data. Compared with the MR-based approach, the transmission-based method provided average improvements in PET quantification of 6.4%, 2.4%, and 18.7% in volumes of interest inside the lung, soft tissue, and bone tissue, respectively.
CONCLUSION: In conclusion, a transmission-based technique with an annulus-shaped transmission source will be more accurate than a conventional MR-based technique for measuring attenuation coefficients at 511 keV in future whole-body PET/MR studies.

Entities:  

Keywords:  MR/PET; TOF; attenuation correction; transmission

Mesh:

Substances:

Year:  2014        PMID: 24434291      PMCID: PMC4010111          DOI: 10.2967/jnumed.113.125989

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  26 in total

1.  Simultaneous reconstruction of activity and attenuation for PET/MR.

Authors:  André Salomon; Andreas Goedicke; Bernd Schweizer; Til Aach; Volkmar Schulz
Journal:  IEEE Trans Med Imaging       Date:  2010-11-29       Impact factor: 10.048

2.  Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data.

Authors:  V Schulz; I Torres-Espallardo; S Renisch; Z Hu; N Ojha; P Börnert; M Perkuhn; T Niendorf; W M Schäfer; H Brockmann; T Krohn; A Buhl; R W Günther; F M Mottaghy; G A Krombach
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-10-05       Impact factor: 9.236

3.  Whole-body 18F-FDG PET/CT in the presence of truncation artifacts.

Authors:  Thomas Beyer; Andreas Bockisch; Hilmar Kühl; Maria-Jose Martinez
Journal:  J Nucl Med       Date:  2006-01       Impact factor: 10.057

Review 4.  Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques.

Authors:  Matthias Hofmann; Bernd Pichler; Bernhard Schölkopf; Thomas Beyer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03       Impact factor: 9.236

5.  MRI-based attenuation correction for PET/MRI: a novel approach combining pattern recognition and atlas registration.

Authors:  Matthias Hofmann; Florian Steinke; Verena Scheel; Guillaume Charpiat; Jason Farquhar; Philip Aschoff; Michael Brady; Bernhard Schölkopf; Bernd J Pichler
Journal:  J Nucl Med       Date:  2008-10-16       Impact factor: 10.057

6.  Three-region MRI-based whole-body attenuation correction for automated PET reconstruction.

Authors:  Jeffrey Steinberg; Guang Jia; Steffen Sammet; Jun Zhang; Nathan Hall; Michael V Knopp
Journal:  Nucl Med Biol       Date:  2009-12-05       Impact factor: 2.408

7.  MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences.

Authors:  Vincent Keereman; Yves Fierens; Tom Broux; Yves De Deene; Max Lonneux; Stefaan Vandenberghe
Journal:  J Nucl Med       Date:  2010-05       Impact factor: 10.057

8.  Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype.

Authors:  Ciprian Catana; Andre van der Kouwe; Thomas Benner; Christian J Michel; Michael Hamm; Matthias Fenchel; Bruce Fischl; Bruce Rosen; Matthias Schmand; A Gregory Sorensen
Journal:  J Nucl Med       Date:  2010-09       Impact factor: 10.057

9.  The effect of limited MR field of view in MR/PET attenuation correction.

Authors:  Gaspar Delso; Axel Martinez-Möller; Ralph A Bundschuh; Stephan G Nekolla; Sibylle I Ziegler
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

10.  Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data.

Authors:  Axel Martinez-Möller; Michael Souvatzoglou; Gaspar Delso; Ralph A Bundschuh; Christophe Chefd'hotel; Sibylle I Ziegler; Nassir Navab; Markus Schwaiger; Stephan G Nekolla
Journal:  J Nucl Med       Date:  2009-03-16       Impact factor: 10.057

View more
  16 in total

1.  Attenuation correction in emission tomography using the emission data--A review.

Authors:  Yannick Berker; Yusheng Li
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

2.  Clinical evaluation of TOF versus non-TOF on PET artifacts in simultaneous PET/MR: a dual centre experience.

Authors:  Edwin E G W Ter Voert; Patrick Veit-Haibach; Sangtae Ahn; Florian Wiesinger; M Mehdi Khalighi; Craig S Levin; Andrei H Iagaru; Greg Zaharchuk; Martin Huellner; Gaspar Delso
Journal:  Eur J Nucl Med Mol Imaging       Date:  2017-01-26       Impact factor: 9.236

Review 3.  From simultaneous to synergistic MR-PET brain imaging: A review of hybrid MR-PET imaging methodologies.

Authors:  Zhaolin Chen; Sharna D Jamadar; Shenpeng Li; Francesco Sforazzini; Jakub Baran; Nicholas Ferris; Nadim Jon Shah; Gary F Egan
Journal:  Hum Brain Mapp       Date:  2018-08-04       Impact factor: 5.038

4.  PET/MRI in the Presence of Metal Implants: Completion of the Attenuation Map from PET Emission Data.

Authors:  Niccolo Fuin; Stefano Pedemonte; Onofrio A Catalano; David Izquierdo-Garcia; Andrea Soricelli; Marco Salvatore; Keith Heberlein; Jacob M Hooker; Koen Van Leemput; Ciprian Catana
Journal:  J Nucl Med       Date:  2017-01-26       Impact factor: 10.057

5.  Performance study of a radio-frequency field-penetrable PET insert for simultaneous PET/MRI.

Authors:  Chen-Ming Chang; Brian J Lee; Alexander M Grant; Andrew N Groll; Craig S Levin
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2018-07-03

Review 6.  Attenuation Correction for Magnetic Resonance Coils in Combined PET/MR Imaging: A Review.

Authors:  Mootaz Eldib; Jason Bini; David D Faul; Niels Oesingmann; Charalampos Tsoumpas; Zahi A Fayad
Journal:  PET Clin       Date:  2015-11-27

Review 7.  MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

Authors:  David Izquierdo-Garcia; Ciprian Catana
Journal:  PET Clin       Date:  2016-01-26

8.  Transmission-less attenuation estimation from time-of-flight PET histo-images using consistency equations.

Authors:  Yusheng Li; Michel Defrise; Scott D Metzler; Samuel Matej
Journal:  Phys Med Biol       Date:  2015-08-12       Impact factor: 3.609

9.  Intrascanner Reproducibility of an SPM-based Head MR-based Attenuation Correction Method.

Authors:  David Izquierdo-Garcia; Mark C Eldaief; Mark G Vangel; Ciprian Catana
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2018-09-06

10.  Transmission imaging for integrated PET-MR systems.

Authors:  Spencer L Bowen; Niccolò Fuin; Michael A Levine; Ciprian Catana
Journal:  Phys Med Biol       Date:  2016-07-06       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.