A calibration method for patient specific IMRT QA using a single therapy verification film.

Arvind Kumar Shukla1, Arun S Oinam1, Sanjeev Kumar2, I S Sandhu2, S C Sharma1.   

Abstract

AIM: The aim of the present study is to develop and verify the single film calibration procedure used in intensity-modulated radiation therapy (IMRT) quality assurance.
BACKGROUND: Radiographic films have been regularly used in routine commissioning of treatment modalities and verification of treatment planning system (TPS). The radiation dosimetery based on radiographic films has ability to give absolute two-dimension dose distribution and prefer for the IMRT quality assurance. However, the single therapy verification film gives a quick and significant reliable method for IMRT verification.
MATERIALS AND METHODS: A single extended dose rate (EDR 2) film was used to generate the sensitometric curve of film optical density and radiation dose. EDR 2 film was exposed with nine 6 cm × 6 cm fields of 6 MV photon beam obtained from a medical linear accelerator at 5-cm depth in solid water phantom. The nine regions of single film were exposed with radiation doses raging from 10 to 362 cGy. The actual dose measurements inside the field regions were performed using 0.6 cm(3) ionization chamber. The exposed film was processed after irradiation using a VIDAR film scanner and the value of optical density was noted for each region. Ten IMRT plans of head and neck carcinoma were used for verification using a dynamic IMRT technique, and evaluated using the gamma index method with TPS calculated dose distribution.
RESULTS: Sensitometric curve has been generated using a single film exposed at nine field region to check quantitative dose verifications of IMRT treatments. The radiation scattered factor was observed to decrease exponentially with the increase in the distance from the centre of each field region. The IMRT plans based on calibration curve were verified using the gamma index method and found to be within acceptable criteria.
CONCLUSION: The single film method proved to be superior to the traditional calibration method and produce fast daily film calibration for highly accurate IMRT verification.

Entities:  

Keywords:  EDR 2 film; IMRT; Patient specific quality assurance; Sensitometric curve

Year:  2013        PMID: 24416558      PMCID: PMC3863263          DOI: 10.1016/j.rpor.2013.04.033

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  16 in total

1.  Film dosimetry for intensity modulated radiation therapy: dosimetric evaluation.

Authors:  Sang Gyu Ju; Yong Chan Ahn; Seung Jae Huh; Inhwan Jason Yeo
Journal:  Med Phys       Date:  2002-03       Impact factor: 4.071

2.  The value of the PinPoint ion chamber for characterization of small field segments used in intensity-modulated radiotherapy.

Authors:  C Martens; C De Wagter; W De Neve
Journal:  Phys Med Biol       Date:  2000-09       Impact factor: 3.609

3.  Rapid radiographic film calibration for IMRT verification using automated MLC fields.

Authors:  Nathan L Childress; Lei Dong; Isaac I Rosen
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

4.  Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance.

Authors:  Arthur J Olch
Journal:  Med Phys       Date:  2002-09       Impact factor: 4.071

5.  Predicting energy response of radiographic film in a 6 MV x-ray beam using Monte Carlo calculated fluence spectra and absorbed dose.

Authors:  Asa Palm; Assen S Kirov; Thomas LoSasso
Journal:  Med Phys       Date:  2004-12       Impact factor: 4.071

6.  Variation of sensitometric curves of radiographic films in high energy photon beams.

Authors:  C Danciu; B S Proimos; J C Rosenwald; B J Mijnheer
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

7.  Anatomy-corresponding method of IMRT verification.

Authors:  Janusz Winiecki; Zbigniew Zurawski; Barbara Drzewiecka; Krzysztof Slosarek
Journal:  Rep Pract Oncol Radiother       Date:  2010-12-04

8.  A new approach to film dosimetry for high energy photon beams: lateral scatter filtering.

Authors:  S E Burch; K J Kearfott; J H Trueblood; W C Sheils; J I Yeo; C K Wang
Journal:  Med Phys       Date:  1997-05       Impact factor: 4.071

9.  Patient-specific point dose measurement for IMRT monitor unit verification.

Authors:  Lei Dong; John Antolak; Mohammad Salehpour; Kenneth Forster; Laura O'Neill; Robin Kendall; Isaac Rosen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-07-01       Impact factor: 7.038

10.  Analysis of the sources of uncertainty for EDR2 film-based IMRT quality assurance.

Authors:  Chengyu Shi; Nikos Papanikolaou; Yulong Yan; Xuejun Weng; Hongyu Jiang
Journal:  J Appl Clin Med Phys       Date:  2006-05-25       Impact factor: 2.102

View more
  1 in total

1.  Adjustment of multi-leaf collimator parameters in 4-MV and 6-MV IMRT: A study of veterinary clinical cases.

Authors:  Toshie Iseri; Yoshinori Tanabe; Hiro Horikirizono; Hiroshi Sunahara; Harumichi Itoh; Yuki Nemoto; Kazuhito Itamoto; Kenji Tani; Hidekazu Tanaka; Munekazu Nakaichi
Journal:  Open Vet J       Date:  2022-06-21
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.