Literature DB >> 24416546

Effect of anode angle on photon beam spectra and depth dose characteristics for X-RAD320 orthovoltage unit.

Asghar Mesbahi1, Seyed-Salman Zakariaee2.   

Abstract

BACKGROUND: In radiation therapy with orthovoltage units, the tube design has a crucial effect on its dosimetric features. AIM: In this study, the effect of anode angle on photon beam spectra, depth dose and photon fluence per initial electron was studied for a commercial orthovoltage unit of X-RAD320 biological irradiator.
MATERIALS AND METHODS: The MCNPX MC code was used for modeling in the current study. We used the Monte Carlo method to model the X-RAD320 X-ray unit based on the manufacturer provided information. The MC model was validated by comparing the MC calculated photon beam spectra with the results of SpekCalc software. The photon beam spectra were calculated for anode angles from 15 to 35 degrees. We also calculated the percentage depth doses for some angles to verify the impact of anode angle on depth dose. Additionally, the heel effect and its relation with anode angle were studied for X-RAD320 irradiator.
RESULTS: Our results showed that the photon beam spectra and their mean energy are changed significantly with anode angle and the optimum anode angle of 30 degrees was selected based on less heel effect and appropriate depth dose and photon fluence per initial electron.
CONCLUSION: It can be concluded that the anode angle of 30 degrees for X-RAD320 unit used by manufacturer has been selected properly considering the heel effect and dosimetric properties.

Entities:  

Keywords:  Anode angle effect; Monte Carlo modeling; Radiation therapy; X-RAD320 irradiator

Year:  2013        PMID: 24416546      PMCID: PMC3863179          DOI: 10.1016/j.rpor.2012.12.001

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  11 in total

1.  Monte Carlo modelling of radiotherapy kV x-ray units.

Authors:  F Verhaegen; A E Nahum; S Van de Putte; Y Namito
Journal:  Phys Med Biol       Date:  1999-07       Impact factor: 3.609

2.  Characterization of the dose perturbation by stents as a function of X-ray beam energy.

Authors:  M C Schell; D P Rosenzweig; K A Weaver; B M Fenton; P Rubin
Journal:  Cardiovasc Radiat Med       Date:  1999 Apr-Jun

3.  A monte carlo comparison of three different media for contrast enhanced radiotherapy of the prostate.

Authors:  H M Garnica-Garza
Journal:  Technol Cancer Res Treat       Date:  2010-06

4.  Monte Carlo simulator of realistic x-ray beam for diagnostic applications.

Authors:  Marco Bontempi; Lucia Andreani; Pier Luca Rossi; Andrea Visani
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

5.  Determining superficial dosimetry for the internal canthus from the Monte Carlo simulation of kV photon and MeV electron beams.

Authors:  B E Currie
Journal:  Australas Phys Eng Sci Med       Date:  2009-06       Impact factor: 1.430

6.  LiF:Mg,Ti TLD response as a function of photon energy for moderately filtered x-ray spectra in the range of 20-250 kVp relative to 60Co.

Authors:  A A Nunn; S D Davis; J A Micka; L A DeWerd
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

7.  SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes.

Authors:  G Poludniowski; G Landry; F DeBlois; P M Evans; F Verhaegen
Journal:  Phys Med Biol       Date:  2009-09-01       Impact factor: 3.609

8.  Converging stereotactic radiotherapy using kilovoltage X-rays: experimental irradiation of normal rabbit lung and dose-volume analysis with Monte Carlo simulation.

Authors:  Takatsugu Kawase; Etsuo Kunieda; Hossain M Deloar; Takanori Tsunoo; Satoshi Seki; Yohei Oku; Hidetoshi Saitoh; Kimiaki Saito; Eileen N Ogawa; Akitoshi Ishizaka; Kaori Kameyama; Atsushi Kubo
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-10-01       Impact factor: 7.038

Review 9.  Small animal radiotherapy research platforms.

Authors:  Frank Verhaegen; Patrick Granton; Erik Tryggestad
Journal:  Phys Med Biol       Date:  2011-05-26       Impact factor: 3.609

10.  Orthovoltage radiation therapy treatment planning using Monte Carlo simulation: treatment of neuroendocrine carcinoma of the maxillary sinus.

Authors:  W Gao; D E Raeside
Journal:  Phys Med Biol       Date:  1997-12       Impact factor: 3.609

View more
  4 in total

1.  Evaluation of a Small Animal Irradiation System for Animal Experiments Using EBT3 Model GAFCHROMIC™ Film.

Authors:  Yasuyuki Shimizu; Hiroaki Akasaka; Daisuke Miyawaki; Naritoshi Mukumoto; Masao Nakayama; Tianyuan Wang; Saki Osuga; Sachiko Inubushi; Ryuichi Yada; Yasuo Ejima; Kenji Yoshida; Takeaki Ishihara; Ryohei Sasaki
Journal:  Kobe J Med Sci       Date:  2018-01-17

2.  Evaluation of Non-Uniform Image Quality Caused by Anode Heel Effect in Digital Radiography Using Mutual Information.

Authors:  Ming-Chung Chou
Journal:  Entropy (Basel)       Date:  2021-04-25       Impact factor: 2.524

3.  Pinhole X-ray fluorescence imaging of gadolinium and gold nanoparticles using polychromatic X-rays: a Monte Carlo study.

Authors:  Seongmoon Jung; Wonmo Sung; Sung-Joon Ye
Journal:  Int J Nanomedicine       Date:  2017-08-11

4.  Kilovoltage radiotherapy for companion animals: dosimetric comparison of 300 kV, 450 kV, and 6 MV X-ray beams.

Authors:  Jaehyeon Seo; Jaeman Son; Yeona Cho; Nohwon Park; Dong Wook Kim; Jinsung Kim; Myonggeun Yoon
Journal:  J Vet Sci       Date:  2018-07-31       Impact factor: 1.672